Abstract
The results of studies on mercury capillary coulometers containing acidified electrolyte solutions (fluoroborate or perchlorate) or neutral solutions are presented. The effects of the solution composition on the coulometer properties such as internal resistance, current yield, and electrochemical stability are discussed. It has been shown that the solutions of mercury salts which contain fluoroboric or perchloric acid are more advantageous for use as mercury coulometer electrolytes than the iodide solutions used hitherto.
The mercury capillary coulometers which are discussed here are widely used as electrochemical integrators and as the working time meters of electrical instruments in laboratories and in industry.
Similar content being viewed by others
References
C. C. Beusman,J. E. E. E. Int. Conv. Rec. 8 (1964) 171.
E. V. Kazaryan and L. L. Gvaramadze,Elektrokhimya 9 (1973) 413.
A. P. Shorygin and G. L. Danelyan,ibid 8 (1972) 777.
W. Kiryluk, British Patent No 1 007 403 and 1 070 913.
W. Kiryluk,Process, Control and Automation 11 (1964) 19.
L. Corrosin, USA Patent No 3 045 178.
J. Kotowski, Doctor's Thesis, Warsaw University, (1975).
Z. Koczorowski and J. Kotowski, PRL Patent No P 169.039.
B. E. Conway,Electrochemical Data, Elsevier, Amsterdam, London (1952).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Kotowski, J., Koczorowski, Z. Effect of electrolyte composition on the properties of mercury electrochemical coulometers. J Appl Electrochem 7, 511–515 (1977). https://doi.org/10.1007/BF00616763