Abstract
We have investigated radiation effects on a Ge:Ga photoconductive infrared detector at low photon background level of 4×108 ph cm−2s−1, using Cobalt 60 as a gamma ray source. The irradiation immediately induced spike noises which degrded NEP (short term effect), while it gradually increased responsivity (long term effect). After the removal of the gamma ray source, the spikes disappeared while the responsivity still stayed in a higher level and gradually decreased with a time scale of several hours. The responsivity-change-rate before and after the irradiation is smaller for a higher bias voltage. Finally we have made the first trial to cure the long term effect, using a flashing procedure and have found out its effectiveness.
Similar content being viewed by others
References
H. Aumann, B. Brown, F. Gillett, W. Irace, D. Langford, P. Mason, and R. Salazar, IRAS In-Orbit Performance Assessment (JPL-871), p 19 (1983).
J. Wolf and D. Lemke, Proc. of the Second ESO Infrared Workshop, eds. A.F.M. Moorwood and K. Kjär, p191 (1982).
M.D. Petroff, J.C. Pickel and M.P. Curry, IEEE Transactions on Nuclear Science,NS-26, 4840 (1979).
D. Koch, G.G. Fazio, W.A. Traub, G.H. Rieke, T.N. Gautier, W.F. Hoffmann, F.J. Low, W. Poteet, E.T. Young, E.W. Urban, and L. Katz, Opt. Eng.21, 141 (1982).
D. Lemke, M. Grewing, D. Offermann, S. Drapatz, and G. Klipping, Adv. Space Res.2, 123 (1983).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Oda, N., Lemke, D. & Wolf, J. Radiation effects on a GE:GA photoconductive detector. Int J Infrared Milli Waves 5, 1499–1505 (1984). https://doi.org/10.1007/BF01009707
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF01009707