Abstract
Laser Doppler perfusion monitoring (LDPM) can be used for monitoring myocardial perfusion in the non-beating heart. However, the movement of the beating heart generates large artifacts. Therefore the aim of the study was to develop an LDPM system capable of correlating the laser Doppler signals to the cardiac cycle and to process the signals to reduce the movement artifacts. Measurements were performed on three calves, both on the normal beating heart and during occlusion of the left anterior descending coronary artery (LAD). The recorded LDPM signals were digitally processed and correlated to the sampled ECG. Large variations in the output (perfusion) and DC signals during the cardiac cycle were found, with average coefficients of variation of 0.36 and 0.14 (n-14), respectively. However, sections with a relatively low, stable output signal were found in late diastole, where the movement of the heart is at a minimum. Occlusion of the LAD showed the importance of recording the laser Doppler signals at an appropriate point in the cardiac cycle, in this case late systole, to minimise movement artifacts. It is possible to further reduce movement artifacts by increasing the lower cutoff frequency when calculating the output signal.
Similar content being viewed by others
References
Ahn, H. C., Ekroth, R., Hedenmark, J., Nilsson, G. E. andSvedjeholm, R. (1988): ‘Assessment of myocardial perfusion in the empty beating porcine heart with laser Doppler flowmetry’,Cardiovasc. Res.,22, pp. 719–725
Ashikawa, K., Kanatsuka, H., Suzuki, T., andTakishima, T. (1986): ‘Phasic blood flow velocity pattern in epimyocardial microvessels in the beating canine left ventricle’,Circ. Res.,59, pp. 704–711
Belboul, A., andAl-Khaja, N. (1997): ‘The effect of protamine on the epicardial microflow and the graft flow in open-heart surgery’,Perfusion,12, pp. 99–106
Borst, C. (2000): ‘Operating on a beating heart’,Sci. Am.,283, pp. 58–63
Calafiore, A. M., Di Giammarco, G., Teodori, G., Mazzei, V., andVitolla, G. (1998): ‘Recent advances in multivessel coronary grafting without cardiopulmonary bypass’,Heart Surg. Forum.,1, pp. 20–25
Derumeaux, G., Ovize, M., Loufoua, J., Andre-Fouet, X., Minaire, Y., Cribier, A., andLetac, B. (1998): ‘Doppler tissue imaging quantitates regional wall motion during myocardial ischemia and reperfusion’,Circulation,97, pp. 1970–1977
Essex, T. J., andByrne, P. O. (1991): ‘A laser Doppler scanner for imaging blood flow in skin’,J. Biomed. Eng.,13, pp. 189–194
Galiuto, L., Ignone, G., andDeMaria, A. N. (1998): ‘Contraction and relaxation velocities of the normal left ventricle using pulsedwave tissue Doppler echocardiography’,Am. J. Cardiol.,81, pp. 609–614
Garcia, M. J., Rodriguez, L., Ares, M., Griffin, B. P., Klein, A. L., Stewart, W. J., andThomas, J. D. (1996): ‘Myocardial wall velocity assessment by pulsed Doppler tissue imaging: characteristic findings in normal subjects’,Am. Heart J.,132, pp. 648–656
Gharagozloo, F., andCohn, L. H. (1987): ‘Measurement of coronary blood flow: a critical review’,J. Surg. Res.,42, pp. 314–327
Gregg, D. E., andFisher, L. C. (1984): ‘Blood supply to the heart’ in ‘Handbook of physiology’, Vol. 4 (American Physiological Society, Bethesda, Maryland, 1984) pp. 1517–1584
Guyton, A. C. (1991): ‘The coronary circulation’ in ‘Textbook of medical physiology, 8th edn’ (W.B. Saunders Company, Philadelphia, 1991), pp. 237–244
Kajiya, E., Yada, T., Matsumoto, T., Goto, M., andOgasawara, Y. (2000): ‘Intramyocardial influences on blood flow distributions in the myocardial wall’,Ann. Biomed. Eng.,28, pp. 897–902
Klassen, G. A., Barclay, K. D., Wong, R., Paton, B., andWong, A. Y. (1997): ‘Red cell flux during the cardiac cycle in the rabbit myocardial microcirculation’,Cardiovasc. Res.,34, pp. 504–514
Marcus, M. L., Wilson, R. F., andWhite, C. W. (1987): ‘Methods of measurement of myocardial blood flow in patients: a critical review’,Circulation,76, pp. 245–253
Mellander, S., andJohansson, B. (1968): ‘Control of resistance, exchange, and capacitance functions in the peripheral circulation’,Pharmacol. Rev.,20, pp. 117–196
Mizutani, T., Takao, M., Onoda, K., Katayama, Y., Yada, I., Yuasa, H., andKusagawa, M. (1992): ‘Laser measurement of myocardial blood flow’,Mie Medical Journal,42, pp. 1–5
Nagel, E., Al-Saadi, N., andFleck, E. (2000): ‘Cardiovascular magnetic resonance: myocardial perfusion’,Herz,25, pp. 409–416
Nilsson, G. E., Tenland, T., andÖberg, P. A. (1980a): ‘A new instrument for continuous measurement of tissue blood flow by light beating spectroscopy’,IEEE Trans. Biomed. Eng.,27, pp. 12–19
Nilsson, G. E., Tenland, T., andÖberg, P. A. (1980b): ‘Evaluation of a laser Doppler flowmeter for measurement of tissue blood flow’,IEEE Trans. Biomed. Eng.,27, pp. 597–604
Nilsson, G. E. (1984): ‘Signal processor for laser Doppler tissue flowmeters’,Med. Biol. Eng. Comput.,22, pp. 343–348
Palka, P., Lange, A., Fleming, A. D., Sutherland, G. R., Fenn, L. N., andMcDicken, W. N. (1995): ‘Doppler tissue imaging: myocardial wall motion velocities in normal subjects’,J. Am. Soc. Echocardiogr.,8, pp. 659–668
Price, D. J., Wallbridge, D. R., andStewart, M. J. (2000): ‘Tissue Doppler imaging: current and potential clinical applications’,Heart,84, pp. ii11–18
Root, C. R., andTashjian, R. J. (1971): ‘Thoracic and abdominal arteriography in calves’,Am. J. Vet. Res.,32, pp. 1193–1205
Wei, K., Jayaweera, A. R., Firoozan, S., Linka, A., Skyba, D. M., andKaul, S. (1998): ‘Quantification of myocardial blood flow with ultrasound-induced destruction of microbubbles administered as a constant venous infusion’,Circulation,97, pp. 473–483
Welch, P. D. (1967): ‘The use of fast Fourier transform for the estimation of power spectra: a method based on time averaging over short, modified periodograms’,IEEE Trans. Audio Electroacoust.,15, pp. 70–74
Von Ahn H. C., Ekroth, R., Nilsson, G. E., andSvedjeholm, R. (1988a): ‘Assessment of myocardial perfusion with laser Doppler flowmetry. An experimental study on porcine heart’,Scand. J. Thorac. Cardiovasc. Surg.,22, pp. 145–148
Von Ahn, H. C., Ekroth, R., Nilsson, G. E., Svedjeholm, R., andThelin, S. (1988b): ‘Laser Doppler flowmetry estimating myocardial perfusion after internal mammary artery grafting’,Scand. J. Thorac. Cardiovasc. Surg.,22, pp. 281–284
Wårdell, K., Jakobsson, A., andNilsson, G. E. (1993): ‘Laser Doppler perfusion imaging by dynamic light scattering’,IEEE Trans. Biomed. Eng.,40, pp. 309–316
Wårdell, K., Hermansson, U., Nilsson, G. E., andCasimir-Ahn, H. (2000): ‘Laser Doppler imaging of myocardial, perfusion during coronary bypass surgery’, in. ‘Optical diagnostics of biological fluids V’, Proceedings of SPIE Vol. 3923, Photonics West, San José, California, pp. 10–17
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Karlsson, M.G.D., Casimir-Ahn, H., Lönn, U. et al. Analysis and processing of laser Doppler perfusion monitoring signals recorded from the beating heart. Med. Biol. Eng. Comput. 41, 255–262 (2003). https://doi.org/10.1007/BF02348429
Received:
Accepted:
Issue Date: