Abstract
A four-channel impedance plethysmograph has been designed. Impedance signals are obtained at two frequencies by measuring both real and imaginary parts. Particular attention has been paid to the sine wave generation circuits that provide system versatility. The required phase-sensitive demodulation is achieved by means of analogue multiplexers. Results show that there are significant variations in the thoracic equivalent capacitance related to respiration and that there is an increased sensitivity to cardiac activity at low frequencies.
Similar content being viewed by others
References
Arenson, J. S., Cobbold, R. S. C. andJohnston, K. W. (1981) Dual-channel self-balancing impedance plethysmograph for vascular studies.Med. & Biol. Eng. & Comput.,19, 157–164.
Evans, W. A. andTowers, M. S. (1980) Hybrid technique in waveform generation and synthesis.IEE Proc., Part G,127, 119–128.
Griffiths, H. andAhmed, A. (1987) A dual-frequency applied potential tomography technique computer simulations.Clin. Phys. Physiol. Meas,8, Suppl. A, 103–107.
Pething, R. (1987) Dielectric properties of body tissues. —Ibid.,8, 5–12.
Rosell, J., Murphy, D., Pallas, R. andRolfe, P. (1988) Analysis and assessment of errors in a parallel data acquisition system for biomedical impedance tomography, —Ibid.,9, Suppl. A, 93–99.
Shankar, T. M. R. andWebster, J. G. (1984) Automatically balancing electrical impedance plethysmography.J. Clin. Eng. 9, 129–134.
Swanson, D. K. andWebster, J. G. (1982) Simple design for an impedance plethysmograph.Med. & Biol. Eng. & Comput.,20, 461–465.
Swanson, D. K. andWebster, J. G. (1983) Errors in four-electrode impedance plethysmography. —Ibid.,21, 674–680.
thomasset, A., Lenoir, J., Jenin, P. andGharib, C. (1987) Estimation of body fluid shift by electrical impedance in various postures. Proc. 3rd European Symp. on Life Sci. Res. in Space, Graz, Austria, Sept. 1987, 323–325.
Towers, M. S. (1982) Programmable waveform generator using linear interpolation with multiplying D/A convertors.IEE Proc., Part G,129, 19–25.