Impedance pneumography—A survey of instrumentation techniques

  • Published:
Medical and biological engineering Aims and scope Submit manuscript

Abstract

The impedance pneumograph is a bioimpedance recorder for indirect measurement of respiration. Using superficial thoracic electrodes, the impedance pneumograph measures respiratory volume and rate through the relationship between respiratory depth and thoracic impedance change. This paper reviews the development of impedance pneumography, and pneumographic instrumentation techniques. The advantages and disadvantages of the bridge, modulated oscillator, constant voltage, voltage divider, and constant current designs are examined and both the bridge and constant current designs are found to be useful. The relationship between impedance change component and instrumention technique is examined and a method for reducing the subjectrground artifact is presented.

The instrumentation techniques discussed also apply to other bioimpedance change measurements.

Sommaire

Un pneumographe d'impédance est un enregistreur de bio-impédance utilisé pour des mesures respiratoires indirectes. L'emploi d'électrodes thoraciques superficielles permet de mesurer au moyen d'un pneumographe d'impédance le volume et le rythme respiratoires en fonction du rapport entre la profondeur de la respiration et les variations de la résistance thoracique. Les différents stades de développement des techniques de la pneumographie d'impédance et de l'instrumentation pneumographique sont revues. Il est procédé à un examen des avantages et désavantages du pont, de l'oscillateur à modulation de fréquence, de l'intensité constante et des système à courant continu. L'utilité du pont et des systèmes à courant continu est démontrée. Les variations d'impédance en fonction des techniques instrumentales sont également ètudiées et une méthode réduisant l'artéfact sujet-terre est proposée.

Les techniques instrumentales étudiées peuvent ègalement être appliquées à d'autres mesures concernant les variations de bio-impédance.

Zusammenfussung

Der Impedanzpneumograph misst indirekt die Atmung durch Registrierung der Bioimpedanz. Mit Hilfe von oberflächlichen Thoraxelektroden wird Atmungsvolumen und-geschwindigkeit gemessen; dabei wird die Beziehung der Atemtiefe zu Änderungen der Thoraximpedanz ausgenutzt. Die vorliegende Arbeit beschäftigt sich mit der Entwicklung der Impedanzpneumographie und der pneumographischen instrumentellen Verfahren. Die Vor-und Nachteile von Geräten mit Brückenschaltung, moduliertem Oszillator, konstanter Spannung, Spannungsteiler und mit konstanter Stromstärke werden erörtert; die Verfahren mit Brückenschaltung und mit konstanter Stromstärke werden als besonders brauchbar bewertet. Der Einfluss der jeweiligen apparativen Technik auf die Impedanzänderungen wird untersucht, und es wird eine Methode zur Verringerung der durch die Versuchsanordnung bedingten Störungen gezeigt.

Die beschriebenen instrumentellen Verfahren sind auch zu anderen Messungen von Bioimpedanzänderungen geeignet.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Abbreviations

Z :

electrical impedance (ω)

Z s :

thoracic impedance of subject (at carrier frequency), alsoZ 0 orZ 1 (ω)

Q:

quality factor, ratio of reactance to resistance (dimensionless)

R e (Z) :

real component ofZ (ω)

I m (Z) :

imaginary component ofZ (ω)

ϱ:

phase angle (°)

R :

resistance (ω)

X c :

capacitive reactance (ω)

ΔZ :

impedance change (ω)

ΔZ :

magnitude of impedance change (ω)

ΔZ :

change of magnitude of impedance (ω)

Δ ϱ:

phase angle change (°)

ΔR :

resistance change (ω)

ΔX c :

capacitive reactance change (ω)

e :

equivalent a.c. source voltage (v)

I 1 andI 2 :

designating current electrodes

E 1 andE 2 :

designating voltage electrodes

References

  • Allison, R. D. (1962)Volumetric Dynamics of Respiration as Measured by Electrical Impedance Plethysmography. Ph.D. Thesis, Wayne State University, Detroit, Mich.

    Google Scholar 

  • Allison, R., Holmes, E. andNyboer, J. (1964) Volumetric dynamics of respiration as measured by electrical impedance plethysmography.J. appl. Physiol. 19, 166–173.

    Google Scholar 

  • Atzler, E. (1935)Dielektrographie. Handbuck der Biologischen Arbeitsmethoden. Vol. 5, pt. 8, pp. 1073–1084, Urban und Schwarzenberg, Berlin-Wein.

    Google Scholar 

  • Atzler, E. andLehman, G. (1932) Über ein neues Verfahren zur Darstellung der Herztätigkeit.Arbeitsphysiologie 5, 636–680.

    Article  Google Scholar 

  • Bagno, S. andLiebman, F. (1959) Impedance measurements of living tissue.Electronics 32, 278–279.

    Google Scholar 

  • Castillo, H. (1964) Transducers for physiological measurements during bioastronautical research and operations.Med. Electron. biol. Engng 2, 109–121.

    Google Scholar 

  • Cremer, M. (1907) Uber die Registrierung mechanischer Vorgänge auf elektrischen wege, speziell mit Hilfe des Saitengalvanometers und Saiten elektrometers.Münch. med. Wschr. 33, 1629–1630.

    Google Scholar 

  • Dubois, D. andNims, L. (1945)An Improved Impedance Plethysmograph. National Research Council, Committee on Aviation Medicine, Report 439.

  • Fenning, C. (1937) A new method of recording physiologic activities—I: Recording respiration in small animals.J. Lab. clin. Med. 22, 1279–1280.

    Google Scholar 

  • Fenning, C. andBonar, B. E. (1937) A new method of recording physiologic activities—II: Simultaneous recording of maternal respiration, intrauterine fetal respiration, and uterine contractions.J. Lab. clin. Med. 22, 1280–1284.

    Google Scholar 

  • Geddes, L. A. andHoff, H. E. (1964) The measurement of physiologic events by electrical impedance.Am. J. med. Electron. 3, 16–17.

    Google Scholar 

  • Geddes, L. A., Hoff, H. E., Hickman, D. M. andMoore, A. G. (1962a) The impedance pneumograph.Aerospace Med. 33, 28–33.

    Google Scholar 

  • Geddes, L. A., Hoff, H. E., Hickman, D. M., Hinds, M. andBaker, L. (1962b) Recording respiration and the electrocardiogram with commor electrodes.Aerospace Med. 33, 791–793.

    Google Scholar 

  • Goldensohn, E. andZablow, L. (1959) An electrical impedance spirometerJ. appl. Physiol. 14, 463–464.

    Google Scholar 

  • Kinnen, E. andKubicek, W. (1963a)Thoracic Cage Impedance Measurements: Impedance Product System. U.S.A.F. School of Aerospace Medicine, Brooks A.F.B., Texas, SAM-TDR-63-69.

    Google Scholar 

  • Kinnen, E. andKubicek, W. (1963b)Thoracic Cage Impedance Measurements: Cancellation of Respiration Effects on Transthoracic Impedance. U.S.A.F. School of Aerospace Medicine, Brooks A.F.B., Texas, SAM-TDR-63-99.

    Google Scholar 

  • Kinnen, E., Kubicek, W., Hill, P. andTurton, G. (1964a)Thoracic Cage Impedance Measurements: Tissue Resistivity in vivo and transthoracic Impedance at 100 Kc/s. U.S.A.F. School of Aerospace Medicine, Brooks A.F.B., Texas, SAM-TDR-64-5.

    Google Scholar 

  • Kinnen, E., Kubicek, W. andPatterson, R. (1964b)Thoracic Cage Impedance Measurements: Impedance Plethysmographic Determination of Cardiac Output (A Comparative Study). U.S.A.F. School of Aerospace Medicine, Brooks A.F.B., Texas, SAM-TDR-64-15.

    Google Scholar 

  • Kinnen, E., Kubicek, W. andTurton, G. (1963)Thoracic Cage Impedance Measurements: Dynamic Characteristics of an Impedance Pneumograph. U.S.A.F. School of Aerospace Medicine, Brooks, A.F.B., Texas, SAM-TDR-63-100.

    Google Scholar 

  • Kinnen, E., Kubicek, K. andWitsoe, D. (1964c)Thoracic Cage Impedance Measurements: Impedance Plethysmographic Determination of Cardiac Output (An Interpretive Study). U.S.A.F. School of Aerospace Medicine, Brooks, A.F.B., Texas, SAM-TDR-64-23.

    Google Scholar 

  • Kubicek, W., Kinnen, E. andEdin, A. (1963)Thoracic Cage Impedance Measurements: Calibration of an Impedance Pneumograph. U.S.A.F. School of Aerospace Medicine, Brooks A.F.B., Texas, SAM-TDR-63-41.

    Google Scholar 

  • Kubicek, W., Kinnen, E. andEdin, A. (1964) Calibration of an impedance pneumograph.J. Appl. Physiol. 19, 557–560.

    Google Scholar 

  • McCally, M., Barnard, G., Robins, K. andMarko, A. (1963)An Impedance Respirometer. 6570th Aerospace Medical Research Laboratories, Aerospace Medical Division, Wright-Patterson A.F.B. Ohio.AMRL-TDR-63-45. (See alsoAm. J. med. Electron. 2, 322–327, 1963).

    Google Scholar 

  • Moskulenko, Yu. Ye., Bayevskiy, R. M. andGazenko, O. G. (1962) Technique for investigating blood circulation in the brain under conditions of an altered gravitational field. InProblems of Space Biology. Vol. I. Moskva, Izd-vo An SSSR, pp. 400–404. (Available in translation as Library of Congress, Aerospace Information Division Report, AID-P-63-42).

  • Nadareyshviei, K. Sh. (1961) A method for electronic pneumography; plethysmography, and blood pressure registration.Fiziol. Zh. SSSR (Journal of Physiology USSSR)47, 1432–1434. (Available in translation as O.T.S. 62-24389).

    Google Scholar 

  • Nyboer, J. (1944) InMedical Physics. Vol. I, Plethysmography, Impedance, pp. 340–341. Ed.O. Glasser. Yr. Book. Pub., Chicago.

    Google Scholar 

  • Nyboer, J. (1950) InMedical Physics, Vol. II, Plethysmography, Impedance, pp. 736–743. Ed.O. Glasser. Yr. Book Pub., Chicago.

    Google Scholar 

  • Nyboer, J. (1959)Electrical Impedance Plethysmography. Charles C. Thomas, Springfield, Ill.

    Google Scholar 

  • Nyboer, J. (1960) InMedical Physics. Vol. III, Plethysmography, Impedance, pp. 459–471. Ed.O. Glasser, Yr. Book Pub.., Chicago.

    Google Scholar 

  • Nyboer, J., Bagno, S., Barnett, A. andHalsey, R. H. (1940) Radiocardiograms: Electrical Impedance Changes of the Heart in Relation to Electrocardiograms and Heart Sounds.J. clin. Invest. 19, 773.

    Google Scholar 

  • Nyboer, J., Bagno, S. andNims, L. (1943)The Impedance Plethysmography: An Electrical Volume Recorder. National Research Council, Committee on Aviation Medicine, Report No. 149.

  • Pacela, A. (1965) InSpace Electronics Symposium Impedance Pneumography. Ed.C. M. Wong, Vol. 6, AAS Science and Technology Series, Sec. III, pp. 25–32. Western Per., N. Hollywood, Calif.

    Google Scholar 

  • Robins, K. E. andMarko, A. R. (1962)An Improved Method of Registering Respiration Rate. Paper 5, Session V, Digest of the 15th Annual Conference on Engineering in Medicine and Biology, Chicago, Ill. p. 18.

  • Rosenblatt, A. (1963) Medical field seeks better skin— Impedance data.Electron. Des. 11, 8.

    MathSciNet  Google Scholar 

  • Schaefer, H. (1944) Ein einfaches Gerät zur Registrierung der Atmung und anderer mechanischer Vorgänge.Pflügers Arch. ges Physiol. 247, 606–610.

    Article  Google Scholar 

  • Schaefer, H., Bleicher, E. andEckervogt, F. (1949) Weitere Beiträge zur elektrischen Reizung und zur Registrierung von elecktrischen Vorgangen und der Atmung.Arch. fid. ges. Physiol. 251, 491–503.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Pacela, A.F. Impedance pneumography—A survey of instrumentation techniques. Med. & biol. Engng. 4, 1–15 (1966). https://doi.org/10.1007/BF02474783

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02474783

Keywords

Navigation