A calibration system for a holter pump for hyperalimentation infusion

  • Published:
Annals of Biomedical Engineering Aims and scope Submit manuscript

Abstract

Pumps are being used for precise, constant fluid delivery. Many pumps of reasonable cost require tedious calibration by serial changing of dial-settings, observing with a stopwatch the time for delivery of a certain volume of fluid, and then calculating the rate for each setting.

A simpler method has been developed whereby, from a graph calculated from observed rate at only two dial-settings, the dial is set to give anapproximate flow-rate. Then, from a table previously calculated from a mathematical formula, the number of observed drops at the bedside is adjusted by the dial to give anexact flow-rate.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Deitel, M. Intravenous hyperalimentation.Canadian Nurse, 1972,69, 38–43.

    Google Scholar 

  • Deitel, M. Sanderson, I., and Petsoulas, T. A system of intravenous hyperalimentation.International Surgery, 1973,58, 670–675.

    PubMed  CAS  Google Scholar 

  • Deitel, M., and Syed, A. K. Elemental diets in the management of clinical nutritional problems.Modern Medicine of Canada, 1974,29, 471–475.

    Google Scholar 

  • Monohan, J. J., and Webb, J. W. Intravenous infusion pumps—An added dimension to parenteral therapy.American Journal of Hospital Pharmacy, 1972,29, 54–59.

    Google Scholar 

  • Sanderson, I., and Deitel, M. Intravenous hyperalimentation without sepsis.Surgery, Gynaecology & Obstetrics, 1973,136, 577–585.

    CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Additional information

This work was aided by a grant from St. Joseph's Hospital (Toronto) Research Foundation and Abbott Laboratories, Ltd.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Deitel, M., Petsoulas, A. & Brenner, D. A calibration system for a holter pump for hyperalimentation infusion. Ann Biomed Eng 3, 111–116 (1975). https://doi.org/10.1007/BF02584492

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02584492

Keywords

Navigation