Introduction

In cochlear implant patients, a large proportion of the success or failure depends on the transfer of stimulating signals from the electrode toward the auditory nerve fibers. An important aspect of the electrode design is electrical impedance, which depends on electrode surface area, morphological processes, and electrochemical processes initiated by electrical stimulation. The development of a new generation of cochlear implant devices as well as modern surgical techniques has, to a great extent, been aimed at improving stimulus-transferring mechanisms. In line with this trend, in the past years, the development of electrode arrays has been focused on the ‘modiolus-hugging’ type of electrodes. The advantages of these so-called perimodiolar designs include lower stimulation levels, a larger dynamic range, better channel separation, and improvement of speech understanding [13]. Despite the advantages of perimodiolar designs, concern about post-implant meningitis has prompted studies reassessing the electrode position after implantation [4]. Thus, there are multiple incentives to strive for an exact documentation of the position of the individual electrode in relation to cochlear structures and the insertion depth of the electrode array [5].

The aim of this study is first to meticulously assess the electrode position in cochlear implant recipients. Formerly, the plain X-ray Stenvers projection was used, but for obvious reasons this method has been abandoned [6, 7]. The study considers the benefits of high-resolution CT (HRCT) scanning as a more exact method to visualize the structures of the temporal bone. The advantages of HRCT include the elimination of respiratory misregistration, decrease of other motion artifacts, and an obvious improvement in patient comfort by reducing the examination time [8]. The main disadvantage of HRCT in the postoperative assessment of a cochlear implant is image degradation, due to partial voluming and the metallic artifacts that may interfere with the visibility of individual electrodes [7, 9, 10]. We recently demonstrated the feasibility of imaging the electrode position within the intracochlear spaces on both an implanted temporal bone and a fresh human cadaver head [11]. So we assume that locating the electrode will be possible with HRCT and that this visualization will allow us to measure the distance to important landmarks in the cochlea, such as the modiolus. Therefore, this study was expected to achieve a standardized HRCT-based assessment of the distance between the electrode and the modiolus.

The second aim of this study is to investigate the possible relations between this morphological parameter and electrophysiological characteristics, that is the electrical impedance (EI) and the electrically evoked compound action potential (ECAP), as measured with neural response telemetry (NRT). The NRT system has become an easy-to-use tool for measuring the ECAP generated by the auditory nerve following electrical stimulation of the cochlea via an electrode of the cochlear implant [12]. The data it yields are currently applied in various methods of fitting speech processors [1316]. When NRT data are collected directly after implantation, we are much better informed about the ability to stimulate spiral ganglion cells. This information might help us to verify the position and function of the implanted electrodes.

The main question that is addressed in this study is to what extent the electrode position is determinative in the electrophysiological functioning of the cochlear implant system.

Materials and methods

Subjects

Five adult patients were included in this study. All of them were subsequently implanted with a Nucleus Contour cochlear implant (Cochlear Corp., Lane Cove, Australia) between October 2005 and November 2005 at the University Medical Center Utrecht. In all patients, the electrodes were inserted as planned, that is according to the insertion depth that had been set out. There were no surgical problems or complications. The patient characteristics are shown in Table 1. The causes of deafness in these patients were meningitis, a congenital and hereditary etiology, and trauma. In one patient, no etiology could be identified.

Table 1 Patients’ characteristics

Electrode impedance measurement

The device was activated using the Windows-based Diagnostic and Programming System software (Win-DPS, release version 126) provided by the manufacturer (Cochlear Corp., Lane Cove, Australia). The stimuli consisted of biphasic current pulses presented at a level of 100 clinical current units, which is approximately 76 μA, and an impulse duration of 25 μs/phase. In the present study, the EI was measured in common ground (CG) mode intra-operatively. In this mode, the impedance is measured between an intracochlear electrode and all other intracochlear electrodes coupled in parallel. Shortcut or open-circuit electrodes were not considered for data analysis. According to the specifications of the manufacturer, this implied the exclusion of data with impedances below 0.7 and above 20 kOhm, respectively.

ECAP measurement

The registration procedure of the ECAP responses was identical to the one we used in previous experiments [17]. The initial examination took place intra-operatively immediately after implantation. The ECAP responses were recorded using a computer equipped with NRT software, version 3.0, distributed by the Cochlear Corporation. We used a modified version of the protocol described by Abbas et al. [18]. Our test parameters are presented in Table 2. The stimulation mode was monopolar (MP1 mode), using the extra cochlear ball reference electrode. Masker advance, which is the masker-probe interval, was fixed at 500 μs. As a rule, the sampling delay, i.e., the interval between stimulation and initiation of sampling, was set at 100 μs. In the event of amplifier saturation, the delay was adjusted until a satisfactory response was obtained. The amplifier gain was set at 60 dB but was decreased to 40 dB in cases of amplifier saturation. We set the number of sweeps at 100, whereas Abbas set it at 50. In conformance with the Abbas protocol, we set the pulse duration at 25 μs per phase. The stimulation levels are described in terms of current level (CL), a quantity defined by the Cochlear Company. The CL ranges from 1 to 255 current units (CU), which corresponds to electrical currents from 10 μA to 1.75 mA. Our aim was to test 22 electrodes (electrodes 1–22) intra-operatively. The electrodes selected for recording were two positions above the stimulation electrode. Thus, for each patient, we selected the second electrode, N + 2, from the stimulation electrode, N, in the apical direction. An exception was made for electrodes 21 and 22, for which the recording electrodes were 19 and 20, respectively.

Table 2 ECAP test parameters

Imaging

Multislice CT was performed 6 weeks after implantation but before the first fitting of the speech processor, using the Philips Brilliance 64-slice CT (Philips Medical Systems, Cleveland, OH, USA). The patients were scanned with the following parameters, according to the imaging protocol we used in previous experiments [11]: 140 kV; effective tube current – time product of 260 mA/s; rotation time 0.5 s. In order to maximize spatial resolution, the collimation was reduced to 2 × 0.5 mm. To reduce aliasing artifacts a low pitch factor of 0.35 was used. The corresponding measures of local exposure dose (CT dose indices, CTDI) was 136 mGy. An ultrahigh resolution mode was used for image acquisition. A corresponding ultrahigh resolution filter was applied for image reconstruction. In order to further optimize resolution, a field of view of 51 mm was chosen, resulting in a pixel size of 0.1 mm. Datasets were transferred to a workstation (Easyvision; Philips, Best, The Netherlands) where 3-D reconstructions were created using multiplanar reformation (MPR), i.e., calculating slices along arbitrary sections. In our study, an MPR was made parallel to the basal turn of the cochlea and perpendicular to the modiolus and thus in the plane of the electrode array. Window width and window level were adjusted until both the cochlear tissues and the individual electrodes could be visualized.

Statistical analysis

All analyses were carried out with Wessa statistical software [19]. The Kendall tau rank correlation was used analyzing the correlation between the ECAP, electrode impedance versus electrode distance.

Results

In all five patients the electrode arrays were fully inserted, and this was confirmed by HRCT (Fig. 1). After initial data analysis the distance of the apical electrodes to the modiolus could not be visualized. The distance of the more basally located electrodes could be reliably assessed, however. In general, about half of all the electrodes inserted could be identified separately (range 8–14). With the limitation of not being able to analyze the more apical part of the array, it was decided to restrict our investigation to the data of electrodes one, four, and seven, as these electrodes could be clearly distinguished in all five patients. Moreover, possible differences between electrophysiological parameters presumably are more apparent with a larger inter-electrode distance.

Fig. 1
figure 1

HRCT image with an MPR in the plane of the electrode array

The distance from the three examined electrodes toward the modiolus is illustrated in Fig. 2. Note that in all five patients a decrease in electrode distance is clearly visible.

Fig. 2
figure 2

Distance electrode–modiolus

Second, we investigated the relationship between the electrode position, the EI, and the ECAP. No shortcut or open-circuit electrodes were measured. In patient 3, no ECAP recordings could be obtained on the study electrodes since the surgical procedure had to be shortened for anesthesiological reasons. No correlation was found between the EI and electrode position. A correlation was found between the electrode distance and the ECAP with a 96% probability using Kendall’s rank correlation. In Fig. 3 the electrode–modiolus distance is plotted against the ECAP levels.

Fig. 3
figure 3

Electrode–modiolus distance versus ECAP

Discussion

In this study, we examined the value of HRCT in assessing the electrode position of cochlear implant recipients. Using an MPR in the plane of the electrode array enabled us to count the number of implanted electrodes in all patients. HRCT seems to have a clear advantage over the previously used plain film radiography with the standard Stenvers projection. With modern multislice CT scanning, more detailed information can be gathered on the intracochlear electrode position. Thus it may be helpful in the assessment of intracochlear damage after implantation, such as degradation of the modiolar wall [20]. When determining the exact position of the electrode in the cochlea, it was difficult to identify the separate electrodes in relation to the modiolus. This was mainly due to image degradation caused by partial volume effects, but metallic artifacts of the electrode array also posed a problem [7, 9, 10]. The window depth in our software configuration limited us in visualizing the three extreme contrasts we wanted to investigate: the fluid compartment of the cochlea; the bony structures of the cochlea and modiolus; and the radiopaque electrode array. In the future, technical improvements in both soft- and hardware will further improve the spatial resolution in HRCT under such extreme contrast conditions.

The second part of this study addressed the relation between electrode distance and electrophysiological parameters. First of all we evaluated the possible relationship between the EI and the electrode–modiolus distance. We could not demonstrate such a relationship. We presume that this is due to the small number of electrodes that we measured. In an earlier study, however, we analyzed the electrical impedance in 52 Nucleus Contour electrode arrays. This showed us a small decrease in electrical impedance in the apical direction. This decrease was especially notable in the first five electrodes [21]. One explanation of this observation is that the more basally located electrodes are less ‘modiolus-hugging’ than the apical ones. The higher electrical impedance can be explained by the larger fluid compartment, which leads to more growth of less conductive fibrous tissue.

A significant correlation was found between the electrode–modiolus distance and the ECAP. Our data suggest that a higher electrode–modiolus distance leads to a higher ECAP. Again the small number of electrodes that could be investigated was a limitation in our analysis. Electrode stimulation led to selective triggering of auditory nerve groups. In this process, many factors are important; the electrode–modiolus distance is just one part of this complex process.

In conclusion, the results of this study suggest that electrode–modiolus distance is of importance in the stimulation of the auditory nerve fibers. In the development of new cochlear implant systems, optimizing the electrode positioning in relation to the modiolus may influence the functional outcome. Further development in HRCT will help us to analyze the electrode positioning postoperatively. This should further increase and refine our insight in the importance of electrode positioning in relation to aspects of electrophysiological functioning. This may eventually lead to an improvement in the functioning of cochlear implant systems in hearing impaired persons.