Abstract
In this work, a magnetorheological (MR) damper valve is designed with the primary objective of controlling swing-phase damping in an above-knee prosthesis. Initially, a swing phase model of the desired single axis knee incorporating MR damper was modelled. The control parameters that govern damping force and displacement of the damper were identified and optimized to enable the prosthetic knee to produce near normal swing phase trajectory for ground walking as obtained from experimental data. Then, the MR damper valve is optimally designed by selecting typical performance indices of the damper for the intended application. A multi-objective optimization problem is formulated where the MR damper valve is constrained in a desired cylindrical volume defined by its radius and height. Effects of the geometrical design variables of the valve are analytically investigated by mapping finite element analysis (FEA) numerical responses with response surface method (RSM). The results show that the MR damper with designed damper valve enables the prosthetic knee to achieve near to normal swing phase trajectory, and compare to the existed MR damper, up to 71 % reduction by weight has been achieved.
Similar content being viewed by others
References
S. N. Whittlesey, R. E. A. Van Emmerik and J. Hamill, The swing phase of human walking is not a passive movement, Motor Control, 4 (3) (2000) 273–292.
J. Z. Chen and W. H. Liao, Design, testing and control of a magnetorheological actuator for assistive knee braces, Smart Materials and Structures, 19 (3) (2010) 035029–38.
K. E. Akdogan, A. Yilmaz, A. Sadeghimorad and I. Sahin, Design of semi active knee joint with magnetorheological (MR) damper, Proceedings of 20th IEEE Signal Processing and Communications Application Conference (SIU), Mugla, Turkey (2012).
S. Solomon, S. Sujatha and C. Sujatha, Design and evaluation of swing phase controllers for single-axis knee, International Journal Bioautomation, 20 (3) (2016) 373–388.
D. Datta, B. Heller and J. Howitt, A comparative evaluation of oxygen consumption and gait pattern in amputees using intelligent prostheses and conventionally damped knee swing-phase control, Clinical Rehabilitation, 19 (4) 2005 398–403.
X. C. Zhu, X. J. Jing and L. Cheng, Magnetorheological fluid dampers: A review on structure design and analysis, Journal of Intelligent Material Systems and Structures, 23 (8) (2012) 839–873.
J. D. Carlson, W. Matthis and J. R. Toscano, Smart prosthetics based on magnetorheological fluids, Proceedings of SPIE 8th Annual Symposium on Smart Structures and Materials, Newport Beach, CA (2001) 308–316.
H. Xie, S. Wang and F. Li, Knee joint optimization design of intelligent bionic leg based on genetic algorithm, International Journal Bioautomation, 18 (3) (2014) 195–206.
A. Poznic, D. Miloradovic and A. Juhas, A new magnetorheological brake`s combined materials design approach, Journal of Mechanical Science and Technology, 31 (3) (2017) 1119–1125.
Q. H. Nguyen and S. B. Choi, Optimal design of a vehicle magnetorheological damper considering the damping force and dynamic range, Smart Materials and Structures, 18 (1) (2009) 015013–22.
B. F. Spencer, S. J. Dyke, M. K. Sain and J. D. Carlson, Phenomenological model for magnetorheological dampers, Journal of Engineering Mechanics, 123 (3) (1997) 230–238.
D. A. Winter, The biomechanics and motor control of human gait: normal, elderly and pathological, 4th Ed., University of Waterloo Press, Waterloo (2005).
M. R. Jolly, J. W. Bender and J. D. Carlson, Properties and applications of commercial magnetorheological fluids, Thomas Lord Research Center, Lord Corp. (1998) 262–275.
D. C. Montgomery, Design and analysis of experiments, 8th Ed., John Wiley & Sons Inc. (2013).
Author information
Authors and Affiliations
Corresponding author
Additional information
Solomon S. received his Ph.D. in Mechanical Engineering from IIT Madras, Chennai, India, in 2017. His research interests include biomechanics, multibody dynamics, and mechanical design.
Rights and permissions
About this article
Cite this article
Seid, S., Chandramohan, S. & Sujatha, S. Optimal design of an MR damper valve for prosthetic knee application. J Mech Sci Technol 32, 2959–2965 (2018). https://doi.org/10.1007/s12206-018-0552-7
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12206-018-0552-7