Abstract
Purpose
The aim of this study was to determine the accuracy of orthopantomograms (OPGs) when assessing post-operative temporomandibular joint (TMJ) implant position, compared with cone beam computerized tomography (CBCT).
Methods
A retrospective analysis was undertaken on six adult patients who were implanted with a custom TMJ prosthesis due to end-stage TMJ disease. Post-operative CBCT was compared with post-operative OPGs. Overall magnification of each OPG was calculated and used to linearly rescale each image. Implant position was assessed by measuring the gonion angle and the distance between each surgical screw and the mandibular gonion (SG length).
Results
Mean magnification for OPGs was 24.2%. There were no significant differences (p > 0.05) in the gonion angle on OPGs compared with CBCT images. There was a mean decrease in SG lengths of 0.02 mm on OPGs, corresponding to error level of 5.31%. The 95% limits of agreement between OPGs and CBCT images for SG lengths were 1.65 mm and − 1.73 mm.
Conclusion
This study presents a clinically applicable and accurate first-line radiographic screening tool to assess TMJ implant position. When combined with clinical assessment, OPGs can help reduce the need for further imaging and radiation exposure post-operatively.
Similar content being viewed by others
References
Dimitroulis G (2013) A new surgical classification for temporomandibular joint disorders. Int J Oral Maxillofac Surg 42(2):218–222
Dimitroulis G, Austin S, Sin Lee PV, Ackland D (2018) A new three-dimensional, print-on-demand temporomandibular prosthetic total joint replacement system: preliminary outcomes. J Craniomaxillofac Surg 46(8):1192–1198
Gupta S, Patil N, Solanki J, Singh R, Laller S (2015) Oral implant imaging: a review. Malays J Med Sci 22(3):7–17
Ko EW, Huang CS, Chen YR, Figueroa AA (2005) Cephalometric craniofacial characteristics in patients with temporomandibular joint ankylosis. Chang Gung Med J 28(7):456–466
Devlin H, Yuan J (2013) Object position and image magnification in dental panoramic radiography: a theoretical analysis. Dentomaxillofac Radiol 42(1):29951683
Kim YK, Park JY, Kim SG, Kim JS, Kim JD (2011) Magnification rate of digital panoramic radiographs and its effectiveness for pre-operative assessment of dental implants. Dentomaxillofac Radiol 40(2):76–83
Schlicher W, Nielsen I, Huang JC, Maki K, Hatcher DC, Miller AJ (2012) Consistency and precision of landmark identification in three-dimensional cone beam computed tomography scans. Eur J Orthod 34(3):263–275
Rejebian GP (1979) A statistical correlation of individual tooth size distortions on the orthopantomographic radiograph. Am J Orthod 75(5):525–534
Liang H, Frederiksen NL (2004) Focal trough and patient positioning. Dentomaxillofac Radiol 33(2):128–129
Sadat-Khonsari R, Fenske C, Behfar L, Bauss O (2011) Panoramic radiography: effects of head alignment on the vertical dimension of the mandibular ramus and condyle region. Eur J Orthod 34(2):164–169
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The senior author (G.D.) holds shares in OMX Solutions Pty Ltd. Co-authors (S.F. and N.W.) are employed by OMX Solutions Pty Ltd.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Mian, M., Fink, S., Ackland, D. et al. Accuracy of orthopantomograms in the assessment of implant position following alloplastic temporomandibular joint replacement. Oral Maxillofac Surg 24, 203–209 (2020). https://doi.org/10.1007/s10006-020-00842-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10006-020-00842-x