Introduction

Dental clinicians are reliant on conventional drilling systems for tooth preparation and cutting diseased dental hard tissues for many decades. Although efficient, they are rather painful, irritating, and noisy and thus may discourage patients to accept treatment. Lasers as a potential painless comfortable alternative to mechanical cutting and drilling systems [1] have stimulated many researchers to investigate their various applications in clinical dentistry.

Since the development of the ruby laser, several types of lasers have been developed and were subjected to numerous investigations. The potential efficiency of the ruby laser in removing carious dental hard tissues was investigated early on [2]. However, because the ruby, argon, carbon dioxide (CO2), and neodymium-doped yttrium–aluminum garnet (Nd:YAG) laser systems require relatively high-energy densities to ablate dental hard tissues, they usually produce subsequent major thermal side effects, such as melting, cracking, glazing, carbonisation, or charring of the enamel and dentin, and produce excessive intrapulpal temperature [1, 35]. These laser systems, despite their common application in soft tissue management procedures, were not therefore strongly recommended for common use in hard tissue treatments.

In order to minimize thermal side effects [68] and increase the ablation efficiency of dental hard tissues, the erbium-doped yttrium aluminum garnet (Er:YAG) and erbium, chromium-doped yttrium scandium gallium garnet (Er,Cr:YSGG) laser systems [912] were developed with expected expanded clinical applications.

The Er:YAG laser was found to have the ability to remove enamel and dentin efficiently and cleanly, and the tissue depth, amount, and volume removal were comparable with that achieved with the conventional dental drill [5, 9, 10]. The amount of tissue removal was directly proportional to the power setting [5, 13] and irradiation exposure time [1]. Moreover, minimal increase in pulpal temperature was produced when the Er:YAG laser was used with continuous water spray [11, 14]. Laser irradiation with either Er:YAG or Er,Cr:YSGG did not also affect the hardness of dentin at a point 25 μm below the cavity floor [6].

The Er,Cr:YSGG laser is a kind of erbium laser of 2,780 nm wavelength, using a pulsed-beam system, fiber delivery, and a sapphire tip bathed in a mixture of air and water vapor. It has been reported to be effective for cutting enamel, dentin, and bone [12, 15] and produced similar morphological changes on dentin like the Er:YAG laser [16]. It has also produced a similar calcium-to-phosphorous ratio and hardness number of the cavity floor compared to bur-cutting method [6, 17, 18]. However, in another study, dentin hardness decreased significantly after irradiation with Er,Cr:YSGG laser [13].

Water spray is used not only to suppress temperature, but also to increase cutting efficiency of Er,Cr:YSGG laser [7, 15]. Furthermore, no pulpal inflammatory response was associated with laser preparations [10]. As an etching tool, the microtensile bond strengths of porcelain veneers bonded to tooth surfaces were reported to have similar values either treated with Er,Cr:YSGG laser or 37 % orthophosphoric acid [19]. However, bonding orthodontic brackets to enamel etched with Er,Cr:YSGG laser was inferior to that etched with 37 % orthophosphoric acid [20].

The micromechanical properties such as microhardness play an important role in stress distribution when the mastication forces are applied on filling material placed in a restored tooth [21] and may also affect the ability of the adhesive material to interact with the tooth surface [22]. A positive correlation between the hardness and fracture resistance of dentin was also assumed [23].

Therefore, before application of laser in the dental clinic, micromechanical properties need to be evaluated, especially as there are only few reports on the microhardness of dentinal surface prepared by Er,Cr:YSGG laser irradiation. The purpose of the current study was to investigate the effect of irradiation with Er,Cr:YSGG or Nd:YAG laser on the microhardness of prepared dentin. The null hypothesis was that neither Er,Cr:YSGG nor Nd:YAG laser affects the microhardness of irradiated dentin.

Materials and methods

Sample preparation

Thirty intact third molars were used in the current study. Informed consent was obtained under a protocol reviewed and approved by the Ethics in Human Research Committee, University of Melbourne. The teeth were cleaned up of soft tissues and stored in 1 % chloramine T solution at 4 °C until use. The root was sectioned 2 mm apical to the cemento-enamel junction using a slow-speed diamond saw with water coolant (Struers, Ballerup, Denmark). The pulp was removed with tweezers; then, the tooth was immersed in 1 % sodium hypochlorite solution (Milton's solution, Procter & Gamble, Paramatta, NSW, Australia) and sonicated for 5 min for complete removal of the remaining pulp tissue.

Dentin disks were prepared by horizontal sectioning of one third of the occlusal surface by means of a slow-speed diamond saw with water coolant (Struers, Ballerup, Denmark) resulting in a crown segment containing a portion of the pulp chamber inferiorly and flat dentin surface superiorly. The surface was visually inspected under light microscopy to ensure complete removal of the enamel. Each dentin surface was finished using a 600-grit SiC paper under running water. The disk thickness was 2 mm.

Laser irradiation

A square plastic block (4 cm square, 5 mm thick) was used to mount each tooth which was attached by means of cyano-acrylate cement (M-Bond 200, Micro-Measurement Group Inc., Raleigh, NC, USA). Each tooth was clamped on a stationary stand. The laser handpiece was held on a movable stand connected to a step-motor at a distance of 1 mm from the dentin surface. The samples were assigned into three experimental groups of ten as the following:

Group 1

The laser used in this group was Er,Cr:YSGG laser (Biolase, Waterlase Millennium, San Clement, CA, USA). This system emits photons at a wavelength of 2.78 μm with pulse duration of 140 μs. The repetition rate was 20 Hz and the power output range was 0 to 6 W. The pulse energy varies from 0 to 300 mJ. A G6 sapphire tip of 600 μm diameter was used. The handpiece was moved horizontally along the mesio-distal plane of the tooth. At the end of each laser line path, the handpiece was moved vertically in a bucco-lingual plane at an interval of 200 μm and then moved at the disto-mesial direction. The repetition line was made automatically at a back and forth direction until one half of the exposed surface is irradiated. The total laser time was about 2 min. The teeth were treated with a 3.5-W setting (70 % water, 60 % air).

Group 2

The same as group 1, except that the teeth were treated with a 4.5-W power output (70 % water, 60 % air).

Group 3

The laser used for this group was a Nd:YAG laser (American Dental Laser dLase 300, Sunrise Technologies Inc., CA, USA) containing a 320-μm-diameter optical fiber. The repetition rate was set to 20 Hz with a 2-W power output. To enhance energy absorption, black ink was applied to the tooth surface before each lasing [24]. The procedure was followed as in groups 1 and 2, except that the laser lines were at an interval distance of 300 μm.

Microhardness measurement

Surface microhardness measurements were performed using Vickers indentor attached to a microhardness tester (MTH-10 Microhardness Tester, Anton Paar GmbH, Graz, Austria). Twenty indentations were made on the sound untreated dentin disk half which served as control and the other 20 indentations on the laser-treated half of each sample. The indentation load was 100 g with 10 s dwell time with at least 300 μm between successive indentations, as a distance of at least two times the impression diagonal for dentin should be kept between the indentations to minimize interaction between neighboring indentations [25].

Subsurface hardness was measured by increasing the indentation load to penetrate the laser-treated layer into the sound-unaffected subsurface layer of dentin. Loads of 200 and 300 g were found not large enough to penetrate the entire thickness of the laser-affected layer. The indentation was inspected under magnification under light microscopy, and if the lased dentin appeared rolled over on the margins and sound dentinal tubules could be visualized within the pyramidal-indentation area, it was concluded that the load has been readily applied onto the sound subsurface area. A load of 400 g with 10 s dwell time was selected.

Ten indentations per laser-treated surface at a load of 400 g and dwell time of 10 s were made. The depth of the laser-affected layer of dentin was calculated from the following equation for Vickers indenter [26]:

$$ A = {4}{h_{\rm{p}}}^{{2}}{\hbox{ta}}{{\hbox{n}}^{{2}}}\theta $$
$$ A = {\left( {d{1} + d{2}/{2}} \right)^{{2}}}/{2} $$

where A is the projected area, d1 and d2 are the length of diagonals of the indent, θ = 68, and h p = the plastic depth of the indent.

Statistical analysis

Surface and subsurface Vickers hardness of various groups were compared using one-way ANOVA at a significance level of 0.05 followed by a Tukey's Studentized post hoc test. Paired t test was used to compare the hardness of laser-treated surface with untreated dentin surfaces for each group.

Results

The average values of Vickers hardness number of surface dentin for the baseline and post-irradiation are shown in Table 1. Surface hardness values of sound dentin (baseline) were not significantly different among the experimental groups (ANOVA, p = 0.25). However, dentin surface hardness decreased significantly after laser irradiation with either Nd:YAG or Er,Cr:YSGG laser (t test, p < 0.05). The surface hardness values of irradiated dentin were significantly different among experimental groups. The surfaces irradiated with 3.5 W Er,Cr:YSGG laser were significantly softer than surfaces irradiated with 4.5 W Er,Cr:YSGG laser and Nd:YAG, 2 W power output. There were significant differences among all the groups (ANOVA, p < 0.05).

Table 1 The means (SD) of Vickers microhardness, subsurface hardness, and h P (in micrometers) value

The surface hardness values measured with load of 4 N (subsurface hardness) are illustrated in Table 1. There were no significant differences between the groups and neither between the subsurface hardness and the baseline surface hardness. The h p value, related to load 4 N, which is considered as the average thickness of the laser-irradiated layer, was 15.5 μm for all the groups. A significant correlation was found between the increasing hardness and h p value (R 2 = .18, p < 0.05).

Discussion

Hardness is the resistance of the material to indentation, and it is a property directly related to the susceptibility of a material to deformation and fracture. The micromechanical properties such as microhardness of irradiated human teeth should be thoroughly investigated for a better understanding that would possibly improve clinical treatment of tooth fracture and predict susceptibility of teeth to fracture [23, 25].

The mean hardness of mineralized dentin found in this study was in accordance with previously reported values ranging from 50 to 70 [27]. In the current study, both Nd:YAG and Er,Cr:YSGG laser irradiation had significantly reduced the hardness of irradiated dentin; thus, the null hypothesis was rejected. The degree of reduction in microhardness was also dependent on the power output of the Er,Cr:YSGG laser used. The use of higher power output with Er,Cr:YSGG laser resulted in dentin surfaces of significantly greater microhardness values (Table 1).

The response of target tissues to laser may be affected by various irradiation parameters but mainly on the distribution of the laser energy within the irradiated object. If the energy source is high enough, it will heat up the irradiated dentin to a point where melting, charring, and recrystallization occurred despite the fact that the incident energy is partially dispersed in or reflected off the targeted dentinal surface. It was concluded that Nd:YAG may raise the temperature on the dentin surface to more than 1,125 °C [28]. However, decomposition of pure hydroxyapatite, the main mineral content of the inorganic components of dentin, has been reported to occur at a temperature of up to 1,500 °C causing multiple cracks [25]. On the other hand, it was found that heating dental hard tissues to a temperature of more than 400 °C causes decomposition of mineral components and formation of a new mineral phase that may enhance the resistance to acid dissolution [8].

Providing that 20 % of dentin is a collagen matrix [29] and that collagen bundles contribute to increasing resistance of dentin to crack propagation and relieving stress [30], it was postulated that the high temperature produced by Nd:YAG irradiation could have vaporized the organic matrix, such as collagen, leaving pores and voids that are responsible for the significant reduction in hardness [25]. This could explain the findings of the current investigation which concurs with previous findings [25, 31]. However, the hardness of dentin has been reported to have increased after irradiation with Nd:YAG laser and was explained by the melting and resolidification of dental hard tissues [32]. Moreover, the effect on hardness could have been related to alteration in mineral content and compositional features of irradiated dental hard tissues [8] as a correlation between mechanical properties and mineral contents in tooth structure was established [32]. It was found that after Nd:YAG irradiation, the Ca/P ratio was low when compared to non-irradiated dentin [1], which might contribute to the progressive deterioration of the dentin hardness after laser irradiation that was observed in the current and previous studies [25, 31].

Unlike Nd:YAG laser, the Er,Cr:YSGG laser is a hydrokinetic system where ablation is water mediated, and flooding the cutting tip with a spray of water minimizes the generated heat [7, 12, 13, 15]. And unlike the use of shorter wavelengths of Nd:YAG (1,064 nm) laser which has no appreciable absorption in dental hard tissue and can lead to thermal cracking and amorphous change in the hydroxyapatite crystal structure, the erbium wavelengths have a considerable affinity for hydroxyapatite and water chromophores.

The Er,Cr:YSGG laser cutting mechanism is called “thermomechanical process” in which emitted laser pulsed-wave is readily absorbed by the water in the hydroxyapatite of enamel and dentin and caused the vaporization of the heated water and other hydrated organic components [7]. Vaporization results in increased internal pressure within the target tissues that eventually caused instantaneous microexplosions that destroy the inorganic components before the melting point of tooth tissue and leads to ejection of micro-fragments [1].

In contrast to Nd:YAG laser, Er,Cr:YSGG laser-treated surfaces demonstrated microirregularities, corrugated or wavy profile, opened protruded dentinal tubules due to preferential ablation of the less-mineralized intertubular dentin, absence of smear layer, and also an absence of thermal changes such as melting or carbonation [6, 13, 16, 33, 34, 35]. At low-power parameter, dentin debris was observed with tags attaching it to the underlying lased dentin [13]. Moreover, it was found that the glassy molten substance without any debris on the dentinal surfaces ablated at high-power settings was composed of the same elements as hydroxyapatite [12, 13]. This was claimed to indicate that dentin constituents undertake thermoplastic properties as the molten heated elements may immediately return to solid state apparently enhanced by water cooling [12].

The high absorption of Er,Cr:YSGG laser wavelength by the water-rich collagen fibers and hydroxyapatite causes simultaneous ablation, besides rapid heating of the mineral contents, contributes to the explosive dissociation of the target hard tissue [8], and might be responsible for the reduction of hardness of irradiated dentin in the current investigation. The findings are in agreement with a previous study where the dentin hardness was significantly reduced after irradiation with Er,Cr:YSGG laser irrespective of the power setting [12]. Similarly, it was found that the laser power setting did not considerably affect the extent of reduction in microhardness although it was directly proportional to the amount of ablation [12]. However, the current findings revealed that the lower power setting (3.5 W) of Er,Cr:YSGG laser reduced the hardness more than the higher power setting of 4.5 W. Probably, the higher power may have induced a more pronounced thermal effect, although minimized by the water spray, which probably resulted in rapid recrystallization similar to the effect produced by Nd:YAG laser. Thermal effect produces resolidified surfaces with an increased hardness.

Contrary to the findings of the current investigation, it was reported that Er,Cr:YSGG laser irradiation did not affect the hardness of the cavity floors compared to bur-prepared cavities [16, 17]. However, unlike the current investigation, in the previous studies, the cavity floor microhardness was measured at a point 25 μm [9] and 50 μm [17] below the cavity floor. Hence, the findings of Hossain et al. [16] and Celik et al. [17] of non-significant differences between the irradiated and bur-cut surfaces could be misleadingly interpreted, because the hardness measurement was not made for the outmost laser-affected layer [25].

Based on the current findings, the indentor loading force of 4 N was employed, based on the pilot study, to penetrate into the laser-unaffected subsurface zone. The plastic depth of the indent was 15.5 μm for all laser groups. This may represent the critical depth of laser irradiation influencing the microhardness of dentin. That critical depth was about 7 μm for Nd:YAG laser as reported by Lee and co-workers [25]. However, different laser parameters were used; thus, a direct comparison cannot be accurately made.

Generally, it is difficult to compare the varied results obtained from the different studies due to the wide diversity in the type of laser energy, power parameters, and presence or absence of water cooling that may render them incomparable.

Conclusions

Within the limitations of the current study, both laser systems used have reduced the microhardness of irradiated dentin significantly compared to the control. Increasing the power setting of Er,Cr:YSGG laser seems to increase the hardness of dentin in a comparable extent to the thermal effect of Nd:YAG. The Er,Cr:YSGG laser reduced the hardness of dentin significantly more than Nd:YAG laser. Further studies are needed to understand the mechanism that affected the mechanical properties of irradiated dentin. It was also concluded that the microhardness measurement should be recorded for the most superficial laser-affected layer to report realistic findings.