Safety and Effectiveness of a Next-Generation Contact Force Catheter: Results of the TactiSense Trial

https://doi.org/10.1016/j.jacep.2021.01.019Get rights and content
Under a Creative Commons license
open access

Abstract

Objectives

The objective of this clinical trial was to evaluate the safety and effectiveness of the TactiCath Contact Force Ablation Catheter, Sensor Enabled (TactiCath SE) (Abbott, Abbott Park, Illinois) for the treatment of drug-refractory, symptomatic paroxysmal atrial fibrillation (PAF).

Background

Catheter ablation of atrial fibrillation (AF) is an established therapy for the treatment of PAF. Ablation technology is evolving with the primary goals of improving efficacy and safety of the procedure.

Methods

This was a multicenter single-arm trial evaluating a novel ablation catheter for the treatment of PAF. A total of 156 subjects were enrolled at 19 sites in the United States, Europe, and Australia. The primary safety endpoint was the rate of device- or procedure-related serious adverse events occurring within 7 days. The primary effectiveness endpoint was acute success defined as pulmonary vein isolation at 30 min after ablation. Two descriptive endpoints were prospectively captured: 1) 1-year freedom from recurrence of symptomatic AF, atrial flutter (AFL), and atrial tachycardia (AT) lasting ≥30 s without a new or increased dose of Class I/III antiarrhythmic drugs; and 2) 1-year drug-free success defined by the absence of any recurrent AF/AFL/AT lasting ≥30 s without using Class I/III antiarrhythmic drugs.

Results

Primary safety events occurred in 4.7% of patients (95% confidence interval [CI]: 2.23% to 8.64%), and the procedure was acutely successful in 98.0% of patients (95% CI: 94.95% to 99.46%). According to Kaplan-Meier estimates at 1 year, 82.2% (95% CI: 74.7% to 87.6%) were free from symptomatic recurrence, and 1-year drug-free success was 68.2% (95% CI: 59.9% to 75.1%).

Conclusions

The TactiCath™ Contact Force Ablation Catheter, Sensor Enabled™ is safe and effective for the treatment of paroxysmal AF.

Key Words

atrial fibrillation
contact force
efficacy
pulmonary vein isolation
radiofrequency catheter ablation
safety

Abbreviations and Acronyms

AAD
antiarrhythmic drugs
AF
atrial fibrillation
AFL
atrial flutter
AT
atrial tachycardia
PAF
paroxysmal atrial fibrillation
PVI
pulmonary vein isolation

Cited by (0)

William Stevenson, MD, served as Guest Editor-in-Chief for this paper.

The authors attest they are in compliance with human studies committees and animal welfare regulations of the authors’ institutions and Food and Drug Administration guidelines, including patient consent where appropriate. For more information, visit the Author Center.