Multimodal simulation of large area silicon photomultipliers for time resolution optimization

https://doi.org/10.1016/j.nima.2021.165247Get rights and content
Under a Creative Commons license
open access

Highlights

  • Monte Carlo framework capable to simulate both parts, physics and electronics timing detector.

  • Application-Specific Integrated Circuit design optimized for high resolution time resolution.

  • Time-of-Flight Positron Emission Tomography towards 10 ps on Coincidence Time Resolution.

  • Silicon Photomultiplier segmentation approach implies better timing.

Abstract

The Silicon Photomultiplier (SiPM) sensor is replacing the extensive use of the Photomultiplier Tube (PMT) in fast timing applications. These photo-sensors can be applied in different fields such as medical imaging systems like Positron Emission Tomography (PET), LIDAR technologies or High Energy Physics (HEP) experiments. More specific, Time-of-Flight PET (ToF-PET) requires further developments to achieve a Coincidence Time Resolution (CTR) of 10ps, this enabling the real time reconstruction and in vivo molecular examination. The most recent state-of-the-art ToF-PET systems can reach 200 ps in CTR. Lowering this value will require a cross-optimization of the scintillator crystal, the sensor and the electronics at the same time. These three elements optimization will be the key to boost the timing resolution of the complete system. The aim of this work is to provide a simulation framework that enables this cross-optimization of the PET system taking into consideration the photon physics interaction in the scintillator crystal, the sensor response (size, dead area, capacitance) and the readout electronics behavior (input impedance, noise, bandwidth). This framework has allowed us to study a new promising approach that helps reducing the CTR parameter by segmenting a large area SiPM into “m” smaller SiPMs and then, summing the signals to recover all the signal spread along these smaller sensors. A 15% improvement on time resolution is expected by segmenting a 4 mm × 4 mm single sensor into 9 sensors of 1.3 mm × 1.3 mm with respect to the case where no segmentation is applied.

Keywords

ASIC
Fast timing
Medical imaging
Front-end electronics
Silicon photomultiplier
Simulation

Cited by (0)