Determination of iodide by detection of iodine using gas-diffusion flow injection and chemiluminescence

https://doi.org/10.1016/j.talanta.2004.08.002Get rights and content

Abstract

This work describes development of a flow injection (FI) system for determination of iodide, based on the chemiluminescence (CL) reaction between iodine and luminol. Iodide in the sample zone is oxidized to iodine. Employment of a gas-diffusion (GD) unit allows for selective detection of the generated CL (425 nm). Preliminary results showed for concentrations of less than 2 mg L−1, that signals were irreproducible and that the calibration was not linear.

In order to solve these problems, a method of ‘membrane conditioning’ was investigated, in which iodide stream was continuously merged with oxidant to generate I2 that conditioned the GD membrane and tubing. This minimized surface interaction between the active surface and the I2 generated from the samples, thus improving both precision and sensitivity. By employing membrane conditioning, it has been possible to reliably detect concentrations down to 0.1 mg L−1.

At the optimized condition, an excellent linear calibration (r2 = 0.999) was obtained from 0.1 to 1.0 mg L−1. The method was successfully applied to determine iodide in some pharmaceutical products such as potassium iodide tablets and a liquid patent medicine. However, for vitamin tablets, ascorbic acid was found to interfere seriously by causing a negative signal.

Introduction

Iodine is an essential nutrient in the human diet. Iodine compounds are found in many foods, both naturally and added as supplements. Iodine compounds are also used in preparations of some pharmaceutical products. For example, iodine in the form of tri-iodide is used as antiseptic and disinfectant, while potassium iodide is thought to act as an expectorant. Iodine, mostly in the form of potassium iodide, is also used as an ingredient in multi-vitamin supplements. In the United States, potassium iodide is available in tablet form and sold in drug store for thyroid protection, in the event of nuclear emergency.

However, the level of iodine should be used with extreme caution in cases where patients are markedly sensitive to iodide [1]. In patients with hyperthyroidism, iodide rapidly inhibits the synthesis of thyroid hormones. Therefore, it is essential to have accurate and precise methods available to determine the iodine content of these pharmaceutical products.

Flow injection (FI) analysis [2] is a technique that provides the means of achieving reproducible automated analyses, and there have been several reports in the literature of the use of FI methods for the determination of iodide based on spectrophotometric [3], potentiometric [4], catalytic spectrophotometric [5], [6], [7], [8], [9], [10] and flame atomic absorption spectrophotometric [11] detection methods. Determinations based on chemiluminescence (CL) emission from the iodine–luminol reaction have also been reported [12], [13], [14], [15]. CL detection is attractive in terms of the relatively low cost of the equipment involved and the simplicity of the detection system. Moreover, the CL method permits very low detection limit for iodine down to 1 × 10−7 M [12]. Iodide does not react with luminol, and it is therefore necessary to first oxidize iodide to elementary iodine, in order to initiate chemiluminescence. According to previous reports, this CL detection may be susceptible to a number of interferences including metal ions, and for this reason, a separation technique is usually carried out before the analysis of real samples [13], [14].

Burguera and others have reported use of a headspace device coupled to a FI system to separate iodine from the matrix before CL detection [13]. Fujiwara et al. proposed the use of on-line oxidation and solvent extraction coupled with reversed micellar mediated CL detection for determination of iodine and iodide in commercial gargle products [15]. In their work, oxidation of iodide to iodine and solvent extraction of iodine were performed simultaneously before CL detection using the reaction of iodine with luminol in a reversed micellar solution of hexadecyltrimethylammonium chloride.

Normally, by incorporating gas-diffusion (GD) into a FI system, volatile analytes can be separated from interferents in the sample matrix via diffusion across the hydrophobic membrane. This process is fairly selective because fewer species are converted to the gaseous form at room temperature [16]. Motomitzu and Yoden, for example, have reported the use of a tubular GD unit for determination of iodide and other halides [17], and Hakedal and Egeberg [18] used a GD–FI system for determination of iodide in brine. In this latter system, the absorbance of tri-iodide in the UV-region was measured for calibration.

We recently reported work based on the use of GD with a flow injection system for quantitative analysis of iodide [19]. However, the sensitivity of this system was limited by the photometric detection of the I3–starch complex, and therefore in this present work, we have utilized the CL reaction between iodine and luminol as the means of detection. Some problems, which are caused by the chemical properties of iodine and the membrane, are described along with appropriate solutions. The potential of the CL–FI system for quantitative analysis of iodide in pharmaceutical samples is presented.

Section snippets

The FI manifold

Fig. 1 depicts two FI systems, which were used in the method development. An AS-90 series autosampler (Perkin-Elmer, USA) was used for automatically loading standard or sample solutions into a 300 μl PTFE loop (1.0 mm i.d.). A FIAS-300 module (Perkin-Elmer, USA) was employed for pumping the reagents. A Metrohm gas-diffusion unit (model 754, Switzerland), fitted inside with a circular PTFE membrane (47 mm i.d., 0.8 mm thickness with pore size 0.45 μm; Sartorius, Germany), was employed. The unit

Manifold design

A gas-diffusion flow injection (GD–FI) system, shown in Fig. 1a, was investigated for determination of iodide. By means of this system, iodide (I) in a sample is oxidized to iodine (I2) in a donor stream (oxidant + water). Iodine thus formed diffuses through the PTFE membrane into an acceptor stream of iodide solution, and then reacts with a stream of luminol to produce chemiluminescence. While this chemiluminescence reaction is not selective for iodine, the use of membrane diffusion should

Conclusions

In this work, a flow injection system, which employs gas-diffusion unit for selective detection of iodide, was developed. Dynamic conditioning of the system by continuous oxidation of iodide to provide a constant background of iodine was found necessary to achieve a linear calibration. This type of system conditioning can be optimized to obtain the desired sensitivity. Generally, this method is sensitive and simple. The present method was successfully applied to determine iodide in KI tablets

Acknowledgements

This work was supported by grants from the Thailand Research Fund, Royal Golden Jubilee Scholarship and the Postgraduate Education and Research in Chemistry.

References (20)

  • F.M. Najib et al.

    Talanta

    (1992)
  • N. Choengchan et al.

    Anal. Chim. Acta

    (2003)
  • M.C. Yebra et al.

    Anal. Chim. Acta

    (2000)
  • W.M. Hardy et al.

    Talanta

    (1977)
  • J.L. Burguera et al.

    Talanta

    (1996)
  • S. Motomizu et al.

    Anal. Chim. Acta

    (1992)
  • USP, DI Drug Information for the Health Care Provider, Kingsport Press, Marryland,...
  • J. Ruzicka et al.

    Flow Injection Analysis

    (1998)
  • C.D. Monks et al.

    Analyst

    (1993)
  • W. Chantore et al.

    Lab. Robotics. Automat.

    (1999)
There are more references available in the full text version of this article.

Cited by (0)

View full text