Pre-concentration and determination of trace amounts of lead in water by continuous precipitation in an unsegmented-flow atomic absorption spectrometric system

Abstract

A continuous precipitation and filtration flow system coupled on-line with an atomic absorption spectrometer for the quantitative pre-concentration and determination of lead at the micrograms per litre level is described. PbII forms a precipitate with ammonia, which is retained on a stainless-steel filter and then dissolved by a stream of nitric acid. A concentration factor of up to 700 is achieved. Several calibration graphs suited to the degree of enrichment are proposed for the determination of lead(II) in the range 1.2–1500 ng ml–1; sampling rates of 1–15 h–1 and relative standard deviations of less than 3.6% are afforded by the proposed method, which has been satisfactorily applied to the determination of lead in tap water.

Article information

Article type
Paper

Analyst, 1987,112, 1233-1236

Pre-concentration and determination of trace amounts of lead in water by continuous precipitation in an unsegmented-flow atomic absorption spectrometric system

P. Martínez-Jiménez, M. Gallego and M. Valcárcel, Analyst, 1987, 112, 1233 DOI: 10.1039/AN9871201233

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements