Calibrator Design for the COBE* Far Infrared Absolute Spectrophotometer (FIRAS)

, , , , and

© 1999. The American Astronomical Society. All rights reserved. Printed in U.S.A.
, , Citation J. C. Mather et al 1999 ApJ 512 511 DOI 10.1086/306805

0004-637X/512/2/511

Abstract

The photometric errors of the external calibrator for the Far Infrared Absolute Spectrophotometer (FIRAS) instrument on the COBE are smaller than the measurement errors on the cosmic microwave background radiation (CMBR) spectrum (typically 0.02 MJy sr-1, 1 σ) and smaller than 0.01% of the peak brightness of the CMB. The calibrator is a reentrant cone, shaped like a trumpet mute, made of Eccosorb iron-loaded epoxy. It fills the entire beam of the instrument and is the source of its accuracy. Its known errors are caused by reflections, temperature gradients, and leakage through the material and around the edge. Estimates and limits are given for all known error sources. Improvements in understanding the temperature measurements of the calibrator allow an improved CMB temperature determination of 2.725±0.002 K.

Export citation and abstract BibTeX RIS

Footnotes

  • The National Aeronautics and Space Administration Goddard Space Flight Center (NASA GSFC) was responsible for the design, development, and operation of the Cosmic Background Explorer (COBE). GSFC was also responsible for the development of the analysis software and for the production of the mission data sets. The COBE program was supported by the Astrophysics Division of NASA's Office of Space Science and Applications.