Performance of a position-sensitive scintillation detector

and

Published under licence by IOP Publishing Ltd
, , Citation J S Karp and G Muehllehner 1985 Phys. Med. Biol. 30 643 DOI 10.1088/0031-9155/30/7/003

0031-9155/30/7/643

Abstract

The spatial resolution of a NaI(Tl), 25 mm thick bar detector designed for use in positron emission tomography has been studied. The position along the 500 mm long detector is determined from the centroid of the light distribution in the crystal as measured by a linear array of photomultiplier tubes. A Monte Carlo computer simulation was performed to investigate the factors limiting the spatial resolution. The program allowed the authors to study the effect of various phototube configurations and crystal surfaces. Since the resolution is affected by the width of the light distribution, the authors studied the effect of sharpening the distribution by modifying the front crystal surface with grooves cut perpendicular to the long axis of the crystal and by using nonlinear preamplifiers. The simulation predicts a spatial resolution (FWHM) of 3 mm with this crystal. Experimental measurements of spatial resolution were performed concurrently with the simulations. In particular, a modified grooved crystal was measured to have 4.0 mm spatial resolution, an improvement over the original crystal without grooves. With delay line pulse shortening, which increases the count rate capability of the detector, the grooved crystal was measured to have 5.5 mm spatial resolution.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/30/7/003