True colour imaging of the fundus using a scanning laser ophthalmoscope

, , and

Published 19 December 2001 Published under licence by IOP Publishing Ltd
, , Citation P Vieira et al 2002 Physiol. Meas. 23 1 DOI 10.1088/0967-3334/23/1/301

0967-3334/23/1/1

Abstract

Currently retinal imaging is performed with the fundus camera. This has a number of limitations, in particular the high level of illuminations required for imaging. The scanning laser ophthalmoscope (SLO) has been proposed as an alternative imaging device but to date one of its main limitations has been that it gives only monochromatic images. In this paper we describe an SLO which uses low power red, green and blue lasers to image the human fundus. Using three lasers simultaneously to produce a colour image will increase the fundus exposure by a factor of three. To overcome this problem, a technique has been developed for multiplexing the lasers so that each point on the retina is imaged by the three lasers pulsed rapidly in sequence. The total exposure is thus kept to the same level as for a single laser and total imaging time is not increased. An example is shown of the image from a patient with diabetic retinopathy.

Export citation and abstract BibTeX RIS