In the first phase of this work, surface defects (metalworking) in stainless steel implantable prostheses and their possible relation to infections that can be generated after surgery was studied. In a second phase, the results obtained in the aforementioned stage were applied to knee cruciate ligaments surgery screws, considering the fact that a substantial number of Mucormycetes infections have been reported after arthroscopic surgery in Argentina since the year 2005. Two types of screws, transverse and interference screws, were analyzed. The Allen heads presented defects such as burrs and metalworking bending as a result of the machining process. These defects allow the accumulation of machining oil, which could be contaminated with fungal spores. When this is the case, the gaseous sterilization by ethylene oxide may be jeopardized. Cortical screws were also analyzed and were found to present serious metalworking defects inside their heads. To reduce the risk of infection in surgery, the use of screws with metalworking defects on the outer surface, analyzed with stereomicroscope and considering the inside part of the Allen as an outer surface, should be avoided altogether.