Time of flight positron emission tomography towards 100ps resolution with L(Y)SO: an experimental and theoretical analysis

, , , , , , and

Published 29 July 2013 © CERN 2013
, , Citation S Gundacker et al 2013 JINST 8 P07014 DOI 10.1088/1748-0221/8/07/P07014

1748-0221/8/07/P07014

Abstract

Scintillation crystals have a wide range of applications in detectors for high energy and medical physics. They are recquired to have not only good energy resolution, but also excellent time resolution. In medical applications, L(Y)SO crystals are commonly used for time of flight positron emission tomography (TOF-PET). This study aims at determining the experimental and theoretical limits of timing using L(Y)SO based scintillators coupled to silicon photomultipliers (SiPMs). Measurements are based on the time-over-threshold method in a coincidence setup utilizing the ultra-fast amplifier-discriminator NINO and a fast oscilloscope. Using a 2 × 2 × 3 mm3 LSO:Ce codoped 0.4% Ca crystal coupled to a commercially available SiPM (Hamamatsu S10931-050P MPPC), we achieve a coincidence time resolution (CTR) of 108±5ps FWHM measured at E=511keV. We determine the influence of the data acquisition system to 27±2ps FWHM and thus negligible as compared to the CTR. This shows that L(Y)SO scintillators coupled to SiPM photodetectors are capable of achieving very good time resolution close to the desired 100ps FWHM for TOF-PET systems. To fully understand the measured values, we developed a simulation tool in MATLAB that incorporates the timing properties of the photodetector, the scintillation properties of the crystal and the light transfer within the crystal simulated by SLITRANI. The simulations are compared with measured data in order to determine their predictive power. Finally we use this model to discuss the influence of several important parameters on the time resolution like scintillation rise- and fall time and light yield, as well as single photon time resolution (SPTR) and the detection efficiency of the SiPM. In addition we find the influence of photon travel time spread in the crystal not negligible on the CTR, even for the used 2 × 2 × 3 mm3 geometry.

Export citation and abstract BibTeX RIS

published under the terms of the Creative Commons Attribution 3.0 License by IOP Publishing Ltd and Sissa Medialab srl. Any further distribution of this work must maintain attribution to the author(s) and the published article's title, journal citation and DOI.

Please wait… references are loading.