2 May 1994 Physical evaluation of laser prostatectomy devices
Author Affiliations +
Proceedings Volume 2129, Lasers in Urology; (1994) https://doi.org/10.1117/12.175043
Event: OE/LASE '94, 1994, Los Angeles, CA, United States
Abstract
In this study, the optical and thermal characteristics of the various commercially available devices for treating benign prostatic hyperplasia, using transurethrally delivered Nd:YAG laser light, were evaluated. To calculate the optimal parameters for each delivery system, the created tissue effect was dynamically modeled. Optically the devices were evaluated using by measuring the 3D irradiance distribution in a water environment. The thermal behavior of the devices was studied by a specially developed thermal imaging technique. The optical and thermal characteristics of the devices together with optical and thermal parameters of prostatic tissue were used as an input for a theoretical model to predict the extent of laser induced permanent damage and thus the effect of the device. Results show the influence of beam profile, spot size and secondary reflections of each laser prostatectomy device and modeling its therapeutic effect, will contribute to obtain an optimal dosimetry with best fitted parameters for each individual device.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christiaan F. P. van Swol, Rudolf M. Verdaasdonck, Jaap Mooibroek, and Tom A. Boon "Physical evaluation of laser prostatectomy devices", Proc. SPIE 2129, Lasers in Urology, (2 May 1994); https://doi.org/10.1117/12.175043
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tissue optics

Thermography

Tissues

Instrument modeling

Light scattering

Thermal modeling

Fiber lasers

Back to Top