1 June 1995 Mueller matrix imaging polarimetry
Author Affiliations +
Abstract
The design and operation of a Mueller matrix imaging polarimeter is presented. The instrument is configurable to make a wide variety of polarimetric measurements of optical systems and samples. In one configuration, it measures the polarization properties of a set of ray paths through a sample. The sample may comprise a single element, such as a lens, polarizer, retarder, spatial light modulator, or beamsplitter, or an entire optical system containing many elements. In a second configuration, it measures an optical system's point spread matrix, a Mueller matrix relating the polarization state of a point object to the distribution of intensity and polarization across the image. The instrument is described and a number of example measurements are provided that demonstrate the Mueller matrix imaging polarimeter's unique measurement capability.
J. Larry Pezzaniti and Russell A. Chipman "Mueller matrix imaging polarimetry," Optical Engineering 34(6), (1 June 1995). https://doi.org/10.1117/12.206161
Published: 1 June 1995
Lens.org Logo
CITATIONS
Cited by 165 scholarly publications and 10 patents.