A Repetitional Pulsed X-Ray Diffractometer System

, , , , , , , , , , and

Copyright (c) 1983 The Japan Society of Applied Physics
, , Citation Satoshi Iida et al 1983 Jpn. J. Appl. Phys. 22 1444 DOI 10.1143/JJAP.22.1444

1347-4065/22/9R/1444

Abstract

A pulsed X-ray source and a diffractometer used with this source have been constructed. The pulsed X-ray tube can be driven at a peak voltage of 300 kV and a peak anode current of 5 kA to give X-ray pulses with a duration of about 40 ns at a repetition rate of up to 10 Hz. The high voltage is generated by a Marx generator coupled to a pulse forming line. One X-ray pulse contains about 3×106 Cu Kα photons. Some preliminary observations of the diffraction patterns from an LiF crystal have been made using a multichannel photodiode array as a detector. The use of this apparatus as a time-resolved X-ray diffractometer are discussed in this report.

Export citation and abstract BibTeX RIS

10.1143/JJAP.22.1444