Approaches for calibration and validation of near-infrared optical methods for oxygenation monitoring

  • Christoph Hornberger EMAIL logo and Heidrun Wabnitz

Abstract

Pulse oximetry for arterial oxygenation monitoring and tissue oximetry for monitoring of cerebral oxygenation or muscle oxygenation are based on quantitative in vivo diffuse optical spectroscopy. However, in both cases the information on absolute or relative concentration of human tissue constituents and especially on hemoglobin oxygenation can often not be retrieved by model-based analysis. An in vivo calibration against an accepted reference measurement can be a practical alternative. Pulse oximeters and most of commercial cerebral tissue oximeters rely on empirical calibration based on invasive controlled human desaturation studies. As invasive in vivo tests on healthy subjects are ethically disputable and should be limited to exceptional cases this calibration practice is unsatisfactory. We present the current status and problems of calibration and validation in pulse oximetry and cerebral tissue oximetry including the pros and cons of in vivo as well as in vitro methods. We emphasize various digital and physical phantom approaches and discuss the prospects of their application and possible further developments.

References

[1] Akl TJ, King TJ, Long R, et al. In vitro performance of a perfusion and oxygenation optical sensor using a unique liver phantom. Proc SPIE 2012; 8229: 822904.10.1117/12.908645Search in Google Scholar

[2] Arridge SR. Photon-measurement density functions. Part I: analytical forms. Appl Opt 1995; 34: 7395–7409.10.1364/AO.34.007395Search in Google Scholar PubMed

[3] Batchelder PB, Raley DM. Maximizing the laboratory setting for testing devices and understanding statistical output in pulse oximetry. Anesth Analg 2007; 105(Suppl 6): S85–S94.10.1213/01.ane.0000268495.35207.abSearch in Google Scholar PubMed

[4] Bickler PE, Feiner JR, Lipnick MS, Batchelder P, MacLeod DB, Severinghaus JW. Effects of acute, profound hypoxia on healthy humans. Anesth Analg 2017; 124: 146–153.10.1213/ANE.0000000000001421Search in Google Scholar PubMed

[5] Bickler PE, Feiner JR, Rollins MD. Factors affecting the performance of 5 cerebral oximeters during hypoxia in healthy volunteers. Anesth Analg 2013; 117: 813–823.10.1213/ANE.0b013e318297d763Search in Google Scholar PubMed

[6] Calabro K, Bigio I. Influence of the phase function in generalized diffuse reflectance models: review of current formalisms and novel observations. J Biomed Opt 2014; 19: 075005.10.1117/1.JBO.19.7.075005Search in Google Scholar PubMed PubMed Central

[7] Chen C, Ahmed M, Häfner T, Klämpfl F, Stelzle F, Schmidt M. Fabrication of a turbid optofluidic phantom device with tunable μa and μ′s to simulate cutaneous vascular perfusion. Sci Rep 2016; 6: 30567.10.1038/srep30567Search in Google Scholar PubMed PubMed Central

[8] Chon B, Tokumasu F, Lee JY, Allen DW, Rice JP, Hwang J. Digital phantoms generated by spectral and spatial light modulators. J Biomed Opt 2015; 20: 121309–121328.10.1117/1.JBO.20.12.121309Search in Google Scholar PubMed

[9] Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen saturation. Anesthesiology 2012; 116: 834–840.10.1097/ALN.0b013e31824c00d7Search in Google Scholar PubMed

[10] Farina A, Torricelli A, Bargigia I, et al. In-vivo multilaboratory investigation of the optical properties of the human head. Biomed Opt Express 2015; 6: 2609–2623.10.1364/BOE.6.002609Search in Google Scholar PubMed PubMed Central

[11] Ferrari M, Quaresima V. Review: near infrared brain and muscle oximetry: from the discovery to current applications. J Near Infrared Spectrosc 2012; 20: 1–14.10.1255/jnirs.973Search in Google Scholar

[12] Gehring H, Duembgen L, Peterlein M, Hagelberg S, Dibbelt L. Hemoximetry as the “gold standard”? Error assessment based on differences among identical blood gas analyzer devices of five manufacturers. Anesth Analg 2007; 105(6 Suppl): 24–30.10.1213/01.ane.0000268713.58174.ccSearch in Google Scholar PubMed

[13] Ghassemi P, Wang J, Melchiorri AJ, et al. Rapid prototyping of biomimetic vascular phantoms for hyperspectral reflectance imaging. J Biomed Opt 2015; 20: 121312.10.1117/1.JBO.20.12.121312Search in Google Scholar PubMed PubMed Central

[14] Greisen G, Andresen B, Plomgaard AM, Hyttel-Sørensen S. Cerebral oximetry in preterm infants: an agenda for research with a clear clinical goal. Neurophotonics 2016; 3: 031407-1–031407-6.10.1117/1.NPh.3.3.031407Search in Google Scholar PubMed PubMed Central

[15] Gunadi S, Leung TS, Elwell CE, Tachtsidis I. Spatial sensitivity and penetration depth of three cerebral oxygenation monitors. Biomed Opt Express 2014; 5: 2896–2912.10.1364/BOE.5.002896Search in Google Scholar PubMed PubMed Central

[16] Hessel TW, Hyttel-Sorensen S, Greisen G. Cerebral oxygenation after birth – a comparison of INVOS and FORE-SIGHT near-infrared spectroscopy oximeters. Acta Paediatr 2014; 103: 488–493.10.1111/apa.12567Search in Google Scholar PubMed PubMed Central

[17] Hornberger C, Knoop P, Matz H, et al. A prototype device for standardized calibration of pulse oximeters II. J Clin Monit Comput 2002; 17: 203–209.10.1023/A:1020795307742Search in Google Scholar

[18] Hornberger C, Knoop P, Nahm W, et al. A prototype device for standardized calibration of pulse oximeters. J Clin Monit Comput 2000; 16: 161–169.10.1023/A:1009931527538Search in Google Scholar

[19] Hwang J, Ramella-Roman JC, Nordstrom R. Introduction: feature issue on phantoms for the performance evaluation and validation of optical medical imaging devices. Biomed Opt Express 2012; 3: 1399–1403.10.1364/BOE.3.001399Search in Google Scholar PubMed PubMed Central

[20] Hyttel-Sorensen S, Kleiser S, Wolf M, Greisen G. Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantom. Biomed Opt Express 2013; 4: 1662–1672.10.1364/BOE.4.001662Search in Google Scholar PubMed PubMed Central

[21] Imonot L, Hersch RD, Hébert MH, Mazauric SM. Multilayer four-flux matrix model accounting for directional-diffuse light transfers. Appl Opt 2016; 55: 27–37.10.1364/AO.55.000027Search in Google Scholar PubMed

[22] ISO 80601-2-61:2011 Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.Search in Google Scholar

[23] Jang HY, Pfefer TJ, Chen Y. Solid hemoglobin-polymer phantoms for evaluation of biophotonic systems. Opt Lett 2015; 40: 4321–4324.10.1364/OL.40.004321Search in Google Scholar PubMed

[24] Kleiser S, Nasseri N, Andresen B, Greisen G, Wolf M. Comparison of tissue oximeters on a liquid phantom with adjustable optical properties. Biomed Opt Express 2016; 7: 2973–2992.10.1364/BOE.7.002973Search in Google Scholar PubMed PubMed Central

[25] Korhonen V, Myllylä S, Kirillin M, et al. Light propagation in NIR spectroscopy of the human brain. IEEE J Sel Top Quantum Electronics 2014; 20: 289–298.10.1109/JSTQE.2013.2279313Search in Google Scholar

[26] Kraitl J, Klinger D, Fricke D, Timm U, Ewald H. Non-invasive measurement of blood components. In: Mukhopadhyay SC, Jayasundera KP, Fuchs A, editors. Advancement in Sensing Technology. Heidelberg: Springer 2013: 237–262.10.1007/978-3-642-32180-1_14Search in Google Scholar

[27] Mannheimer PD. The light-tissue interaction of pulse oximetry. Anesth Analg 2007; 105(6 Suppl): 10–17.10.1213/01.ane.0000269522.84942.54Search in Google Scholar PubMed

[28] Moyle JT. Pulse oximetry. London: BMJ Books 2002.Search in Google Scholar

[29] Nasseri N, Kleiser S, Ostojic D, Karen T, Wolf M. Quantifying the effect of adipose tissue in muscle oximetry by near infrared spectroscopy. Biomed Opt Express 2016; 7: 4605–4619.10.1364/BOE.7.004605Search in Google Scholar PubMed PubMed Central

[30] Nitzan M, Romem A, Koppel R. Pulse oximetry: fundamentals and technology update. Med Devices (Auckl) 2014; 7: 231–239.10.2147/MDER.S47319Search in Google Scholar PubMed PubMed Central

[31] Pifferi A, Contini D, Mora AD, Farina A, Spinelli L, Torricelli A. New frontiers in time-domain diffuse optics, a review. J Biomed Opt 2016; 21: 091310-1–091310-17.10.1117/1.JBO.21.9.091310Search in Google Scholar PubMed

[32] Pifferi A, Torricelli A, Bassi A, et al. Performance assessment of photon migration instruments: the MEDPHOT protocol. Appl Opt 2005; 44: 2104–2114.10.1364/BIO.2004.WD7Search in Google Scholar

[33] Pogue BW, Patterson MS. Review of tissue simulating phantoms for optical spectroscopy, imaging and dosimetry. J Biomed Opt 2006; 11: 041102.10.1117/1.2335429Search in Google Scholar PubMed

[34] Prahl SA, van Gemert M, Welch A. Determining the optical properties of turbid media by using the adding–doubling method. Appl Opt 1993; 32: 559–568.10.1364/AO.32.000559Search in Google Scholar PubMed

[35] Rackebrandt K, Gehring H. Calibration and evaluation of a continuous wave multi-distance NIRS system in simulated desaturation investigations. Biomed Phys Eng Express 2016; 2: 035017.10.1088/2057-1976/2/3/035017Search in Google Scholar

[36] Schell RM, Cole DJ. Cerebral monitoring: jugular venous oximetry. Anesth Analg 2000; 90: 559–566.10.1097/00000539-200003000-00012Search in Google Scholar PubMed

[37] Suzuki S, Takasaki S, Ozaki T, Kobayashi Y. A tissue oxygenation monitor using NIR spatially resolved spectroscopy. SPIE Proc 1999; 3597: 582–592.10.1117/12.356862Search in Google Scholar

[38] Tomlin KL, Neitenbach A-M, Borg U. Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in healthy volunteers. BMC Anesthesiol 2017; 17: 6.10.1186/s12871-016-0298-7Search in Google Scholar PubMed PubMed Central

[39] Wabnitz H, Jelzow A, Mazurenka M, et al. Performance assessment of time-domain optical brain imagers, part 2: nEuroPt protocol. J Biomed Opt 2014; 19: 086012.10.1117/1.JBO.19.8.086012Search in Google Scholar PubMed

[40] Wang J, Coburn J, Liang C-P, et al. Three-dimensional printing of tissue phantoms for biophotonic imaging. Opt Lett 2014; 39: 3010–3013.10.1364/OL.39.003010Search in Google Scholar PubMed

[41] Yudovsky D, Durkin A. Hybrid diffusion and two-flux approximation for multilayered tissue light propagation modeling. Appl Opt 2011; 50: 4237–4245.10.1364/AO.50.004237Search in Google Scholar PubMed PubMed Central

Received: 2017-07-10
Accepted: 2017-08-10
Published Online: 2018-02-09
Published in Print: 2018-10-25

©2018 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 14.5.2024 from https://www.degruyter.com/document/doi/10.1515/bmt-2017-0116/html
Scroll to top button