Thyroglobulin measurement by highly sensitive assays: focus on laboratory challenges

  • Luca Giovanella

    Luca Giovanella MD, PhD is the Chief of the Department of Nuclear Medicine-PET/CT and Thyroid Centre at Oncology Institute of Southern Switzerland (IOSI) and the Scientific Advisor of the Department of Clinical Chemistry and Laboratory Medicine at Ente Ospedaliero Cantonale in Bellinzona and Lugano (Switzerland). He serves as Privat-Dozent (PD) at University of Zürich (Switzerland). His clinical and scientific activities are mainly focused on the diagnosis and treatment of thyroid diseases.

    EMAIL logo
    , Ulla Feldt-Rasmussen

    Ulla Feldt-Rasmussen MD, is Professor at University of Copenhagen and Chief Physician of Medical Endocrinology at National University Hospital. Her research interests involve the thyroid gland, cancer and autoimmunity. She has published more than 320 papers in peer-reviewed journals, organized international meetings and postgraduate courses, and led several European thyroid-related projects. Dr. Feldt-Rasmussen is reviewer and editorial board member of many international journals, belongs to numerous international professional organizations and has served as Secretary-Treasurer of European Thyroid Association, and as President of ETA Cancer Research Network. She has received several prestigious prizes and was Haines lecturer at Mayo Clinic in 2011, and Associate Professor at University of Messina, Italy.

    , Frederik A. Verburg

    Frederik A. Verburg MD, PhD studied medicine at the universities of Leuven (Belgium) and Utrecht (The Netherlands) from 1997 to 2004. In 2008 he obtained his PhD in medicine at the University of Utrecht. He trained as a nuclear medicine physician at the University Medical Center Utrecht, The Netherlands, and the University Hospital Würzburg, Germany, from 2005 to 2010. From 2011 to the present he is working as a consultant nuclear medicine physician and Assistant Professor of Nuclear Medicine at the RWTH University Hospital Aachen in Aachen, Germany.

    , Stephan K. Grebe

    Stefan K. G. Grebe MD, PhD is Professor of Laboratory Medicine and Pathology (DLMP), Mayo Clinic College of Medicine. He trained in Internal Medicine and Endocrinology and joined the DLMP of the Mayo Clinic as a Co-Director of the Endocrine Laboratory in 2002. He served as Chair of the Division of Clinical Biochemistry and Immunology from 2006 to 2014. His main clinical and research interests are currently focused on thyroid cancer tumor markers and clinical mass spectrometry.

    , Mario Plebani

    Mario Plebani obtained his medical degree summa cum laude from the Medical School of the University of Padova in 1975. He completed residency training and specialization in Laboratory Medicine (1978), and subsequently in Gastroenterology (1983), at the same University. He is full Professor of Clinical Biochemistry and Clinical Molecular Biology at the University of Padova, School of Medicine, Chief of the Department of Laboratory Medicine at the University-Hospital of Padova, and Chief of the Center of Biomedical Research (a specialized Center for quality in laboratory medicine for the Veneto Region). Currently, he is also Director of the Postgraduate School in Clinical Biochemistry at the Medical School of the Padova University and President of the Course for Medical Technologists at the same Medical School. He served as President of the International Society of Enzymology (ISE) for 4 years (2004–2008) and as President of the Italian Society of Clinical Biochemistry and Molecular Clinical Biology for 5 years (in 2003 and from 2007 to 2009). He is the Chairman of the IFCC Working group on ‘Laboratory errors and patient safety’ (WG LEPS) and in 2008 received the AACC Award for Outstanding Clinical Laboratory Contributions to Improving Patient Safety. Dr. Plebani is Editor in Chief of CCLM (Clinical Chemistry and Laboratory Medicine), and Associate editor of CRC Clinical Laboratory Sciences, and International Journal of Biological Markers. His main areas of research are quality in laboratory medicine, biomarkers in cancer and cardiovascular diseases, and in vitro allergy diagnostics.

    and Penelope M. Clark

    Penelope M. Clark PhD, is an Honorary Consultant Clinical Scientist and Senior Lecturer at The Queen Elizabeth Hospital Birmingham and the University of Birmingham. She headed the Regional Endocrine Laboratory with responsibility for national services for thyroid disease and pituitary disease including thyroid cancer. She is a Fellow of The Royal College of Pathologists and The Association for Clinical Biochemistry and Laboratory Medicine.

Abstract

Differentiated thyroid cancer (DTC) is the most common endocrine cancer and its incidence has increased in recent decades. The initial treatment consists of total thyroidectomy followed by ablation of thyroid remnants by radioiodine in most cases. As thyroid cells are the only source of thyroglobulin (Tg), circulating Tg serves as a biochemical marker of persistent or recurrent disease in the follow-up of DTC. Due to the suboptimal clinical detection rate of older Tg assays endogenous or exogenous thyrotropin (TSH) stimulations are recommended for unmasking occult disease. However, the development of new Tg assays with improved analytical sensitivity and precision at low concentrations now allows detection of very low Tg concentrations, reflecting minimal amounts of thyroid tissue, even without the need for TSH stimulation. Even if the use of these assays still has not found its way in current clinical guidelines, such assays are now increasingly used in clinical practice. As serum Tg measurement is a technically challenging assay and criteria to define a ‘highly sensitive’ assay may be different, a good knowledge of the technical difficulties and interpretation criteria is of paramount importance for both clinical thyroidologists, laboratory physicians and scientists involved in the care of DTC patients.

Introduction

Thyroglobulin (Tg) is a large, 660-kDa, glycoprotein which normally is stored in the follicular colloid of the thyroid gland where it acts as a substrate for thyroid hormone production. As it is only produced by benign or well-differentiated malignant thyroid cells, it is a very good tumor marker for patients with differentiated thyroid carcinoma (DTC) [1], especially after removal of all benign and malignant thyroid tissue through surgery and I-131 ablation [2].

Over the years, the sensitivity and precision of Tg assays have improved by multiple orders of magnitude [3]. Currently, Tg measurement in combination with ultrasound of the neck is considered standard of care in the follow-up of DTC [4, 5]. However, until recently, optimal sensitivity of Tg assays for the detection of smaller disease foci required stimulation of endogenous Tg production by high serum thyroid stimulating hormone (TSH) concentrations [4–7], requiring expensive injections with recombinant human TSH or withdrawal of levothyroxine (LT4) medication which is associated with profound hypothyroidism. Further improvements in assay technology have however recently resulted in new highly sensitive Tg (hsTg) assays [8, 9]. From an increasing number of studies in the literature it appears that these assays are in fact sufficiently sensitive to obviate the need for TSH stimulation in most DTC patients [10, 11]. Therefore, the increasing adoption of these assays in clinical practice has considerable implications, such as a reduction of costs of DTC follow-up or avoidance of hypothyroidism. However, hsTg measurement is technically challenging and, in addition, criteria adopted to define a ‘highly sensitive’ assay by different manufacturers, laboratories and clinicians are very likely to diverge considerably.

The aim of the present paper is to review advances and challenges in Tg measurement techniques and their impact on clinical management of DTC patients. We hope this could be useful in providing a ‘sense of direction’ for laboratories and clinicians that are considering the use of hsTg assays.

Tg assays

Analytical performance

Immunoassay has been the main analytical technique used for the measurement of serum Tg, at first by competitive immunoassay and later by immunometric (reagent excess) assays. More recently, mass spectrometric methods have been developed. Each new assay format has been developed to attempt to overcome the major analytical challenges in measuring, as a tumor marker, a heterogeneous analyte of large molecular weight in the presence of endogenous and other interfering antibodies. The key performance characteristics can be listed as:

  1. The clinical requirement for a wide measuring range of five orders of magnitude with a linear response;

  2. The absence of a high dose hook effect;

  3. Comparability of results obtained by different analytical methods;

  4. Accurate results with quantitative recovery;

  5. Absence of interference by anti-thyroglobulin antibodies (TgAb) and heterophilic antibodies (HAb);

  6. Commutability of Tg in different samples matrices.

Whilst substantial efforts have been made to overcome the problems of antibody interference (discussed elsewhere in this article), it is notable that there is no Reference Method System including a Reference Method Procedure available for Tg and use of the BCR® 457 Certified Reference Material (formerly CRM 457) has not completely eliminated the notable differences in results obtained by different methods. Issues of commutability of the BCR® 457 material and the need for harmonization have yet to be addressed (discussed elsewhere in this article). Due to the lack of agreement between different immunoassays it has been recommended that patients should be followed up using the same assay [5, 6, 11, 12]. If an assay change is required then dual reporting of patient results for a period of time to allow comparison of results by re-baselining serum Tg quantitation by the new assay is suggested. In addition, it is suggested that clinical decision limits should be assay specific [11]. Laboratories will be familiar with the necessary experiments to be performed to verify assay performance [13, 14], namely assessment of linearity, measuring range, trueness (measurement bias), comparability through patient comparison studies, limit of detection/limit of quantitation/functional sensitivity. It is worth considering some particular points with regard to serum Tg [15]:

Assay diluents

Commercial assays may be provided with an assay diluent and specify a dilution value (e.g., 1 in 10 v/v). Laboratories may wish to consider whether this covers the range of concentrations that is required from a clinical perspective, such as the monitoring of metastatic disease where serum Tg may reach concentrations >1000 μg/L. The concentration at which the high dose hook effect has been excluded should also be determined. The feasibility of using an in-house human serum pool as diluent (with undetectable TgAb as measured by a suitable assay and Tg concentration <0.1 μg/L) may need to be investigated and the linearity over a wider concentration range determined.

Recovery experiments

Estimation of recovery of added Tg has been proposed as a method of assessing whether there is interference by endogenous antibodies (i.e., TgAb and HAb) though this is not currently advocated by current guidelines (see section on Interferences). Nevertheless, determination that an assay shows quantitative recovery should be performed as part of the method validation. Studies have shown that the measured recovery is dependent on the protocol used – in particular the Tg concentration, source of Tg (degree of iodination) and incubation time [16, 17]. Assessment of recovery using a source of Tg independent of the kit calibrators is suggested, though manufacturers may recommend a protocol if recovery is being determined in the context of assessing assay interference.

Analytical sensitivity

Recent years have seen growing recognition of the clinical need for improved precision of assays at low Tg concentrations and this has been paralleled by improvements in assay performance. This inter-relationship has raised a number of issues: what constitutes a significant change (rise or fall) in serum Tg from a clinical perspective and from an analytical perspective? What is the lowest serum Tg concentrations that is clinically significant? In the first case, serial measurements may be necessary in order to assess the significance of any change, rather than acting on a single result [12, 18, 19]. In the second case, there has been a gradual improvement in assay sensitivity with the original radioimmunoassays (RIAs) reporting down to 2–5 μg/L, the first immunometric assays (IMA) down to 1 μg/L and more recently IMAs with limits of 0.1 μg/L. In comparison, reported mass spectrometric methods have sensitivities of 0.5–2 μg/L. However, these broad comparisons are limited because of differences in bias between different assays and different experimental and statistical methods were used to determine the sensitivity of the assays.

In the first instance analytical sensitivity has often been determined by repeat analysis of the zero calibrator and determination of the apparent concentration equal to the zero plus (for immunometric assays) two or three standard deviations of the signal (minus for competitive assays), when it is known as the limit of the blank (LOB). There are significant limitations to this approach. In the majority of cases the measured sensitivity will be below the concentration of the lowest concentration calibrator. The assumption is often made that the fitted standard curve is close to the measured dose-response curve. This may not be the case and can be difficult to assess for automated immunoassays leading to a misleading estimate of the LOB [13]. Although of limited use in understanding the precision of low concentration samples, the LOB can be useful when optimizing assay conditions during assay development. The limit of detection (LOD) is defined as the lowest analyte concentration that can be distinguished from the LOB using replicate analysis of a sample of known low concentration. LOD has similar limitations to those of LOB.

Thus functional sensitivity (FS) was introduced as a measure of analytical sensitivity and was originally described for assessing the sensitivity of TSH assays. The National Academy of Clinical Biochemistry (NACB) published guidelines which included advice on how to determine FS though it should be noted that these have now been archived ‘per NACB and National Guideline Clearinghouse policy’ (http://www.aacc.org/SiteCollectionDocuments/Archived%20and%20Historical/ThyroidArchived2010.pdf#page=1) [20]. The NACB protocol indicated that FS may be determined from between batch precision of measurement of patient pools, in the same test mode (singleton or duplicate) as patient samples over the clinically relevant concentration range over two different lots of reagents and calibrators and over a period of 6 months. The patient pools should be TgAb negative and should cover the clinically relevant concentration range. The protocol specifies three different concentration ranges for the patient pools. Whilst it may be difficult to obtain individual patient samples to determine a precision profile and thus FS, this approach can have advantages in terms of covering the concentration range and is recommended by the British Thyroid Association [12]. From the calculated precision profile a cut-off value corresponding to a CV of 20% (somewhat arbitrarily) is taken as the FS. The difference in FS between Tg assays has created a ‘generational’ nomenclature system with each subsequent generation exhibiting a substantial improvement (i.e., 10-fold).

It should be recognized, however, that there are also limitations to this approach when determining the sensitivity of an assay. For example from an analytical perspective the sample type may influence results – control sera should not be used due to potential matrix effects. There are also differences in the statistical approach to the calculation of the precision profile, principally to do with the identification of outliers and the confidence intervals of the profile [13]. The latter authors emphasize the need to determine confidence intervals of the precision profile – these may vary significantly over the concentration range and indicate the confidence with which the calculated FS can be viewed. In addition, the published literature may not be helpful by not providing details of the procedure used to generate the precision profile (e.g., whether within or between batch precision was used, how many samples were analyzed and whether samples were analyzed in singleton or duplicate). With sparse data the relationship between imprecision and concentration can be poorly estimated. Finally, assays may be adapted over time (e.g., through reagent changes or recalibrations) even though they keep the same brand name. A good example of this is provided by the ‘high sensitivity’ troponin assays where changes in reagent lots and calibration algorithms have led to changes in FS. Lastly the limit of quantitation (LOQ) is similar to the FS but does have an additional requirement for predefined goals for bias and imprecision [14, 21] and is increasingly used as a measure of sensitivity for both immunoassay and mass spectrometric assays. Manufacturers may quote LOB, LOD and LOQ as determined by regulatory authorities and national guidelines (e.g., those of the Clinical and Laboratory Standards Institute EP17-A2) [22]. The relationship between these estimates of sensitivity is: LOB<LOD≤LOQ.

Given that assay performance can vary with time, operator, reagent lot, calibration, equipment maintenance and other factors monitoring of sensitivity whether as FS or LOQ should be ongoing and laboratories should determine their own FS/LOQ rather than just quoting manufacturers’ data.

Thus, it becomes clear that in comparing the performance of different Tg assays and in order to provide clinicians with realistic interpretations of Tg results, it is necessary to know exactly how ‘sensitivity’ has been determined such that like can be compared with like. Some examples of these approaches are illustrated in Table 1.

Table 1

Values for Tg assays for LOD, LOB, LOQ or FS as reported in the literature or quoted by manufacturers.

AssayManufacturerMethodTg values for LoB, LoD, LoQ, FS in ug/L
Dynotest Tg PlusBRAHMSThermofisher, DIRMALOB 0.16
FS 0.4 [23]
M quoted values but corrected for difference in manufacturers standardization to CRM 457 (i.e., multiplied by 2)
Tg-AccessBeckmann, USICMALOD 0.1 (M)a
FS 0.1 [24] Method not stated.
FS 0.1 [25] Precision profile determined from 6 pools, TgAb status not stated.
FS 0.1 [26] Precision profile from 10 serum pools at a CV=20%.
FS 0.11 [27] Method of determination not stated.
FS 0.05 [28] According to NACB protocol.
e-Iason TgCaIason GmbH (A)ELISAFS 0.02 (M)
ELSA-hTgCIS Bio International (F)IMAFS 0.8 [29] Method not stated.
TgII Roche ElecsysRoche Diagnostics AG (CH)ICMALOB 0.02 (M)
LOD 0.04 (M)
LOQ 0.1 (M) Total allowable error of ≤30%
All determined according to CLSI EP17-A requirements
Us-TgBRAHMS Thermofisher (D)TRACELOB=0.02 M
LOD=0.04 M
LOQ=0.1 M, total allowable error of ≤40%
All determined according to CLSI EP17-A requirements
Immulite 2000Siemens Healthcare (D)ICMALOD=0.2 (M)
FS=0.9 (M)
Method of determination not stated
FS=0.36 [30] according to NACB protocol
FS=0.6 [31] based on in-house (non-manufacturer) calibration allowing for calculation of ‘negative’ concentrations

ELISA enzyme-linked immunosorbent assay; FS, functional sensitivity; ICMA immunochemiluminometric assay; IECMA immunoelectrochemiluminometric assay; IRMA, immunoradiometric assay; LOB, limit of the blank; LOD, limit of detection; LOQ, limit of quantification; M, manufacturer’s quoted details; NA, not available; NQ, not quoted; TRACE, time-resolved amplified cryptate emission. aInsert package reports the term ‘analytical sensitivity’.

Development of mass spectrometric assays

Protein identification and quantification by mass spectrometry (MS) has a long tradition in laboratory medicine. More recently, efforts have been directed towards the development of MS-based Tg assays. The MS work-flow for protein measurements involves a digestion of the sample with trypsin. Trypsin cleaves protein in a predictable fashion into peptides, which are measured by MS and identified by protein database matching. In the case of Tg, the tryptic digest will cleave all proteins in the sample. One can then specifically look for (one or several) tryptic peptides that are proteotypic for Tg based on predicted cleavage [32]. Quantification is obtained by including a standard curve based on synthetic version of the measured peptides and adding a non-radioactive isotopic version of the peptides in defined concentrations as internal standards to the samples before analysis [33]. In practice, however, many problems may affect Tg measurement by MS. One faces formidable signal to noise problems in identifying the quite low concentration Tg peptides in a trillion-fold higher background abundance of all the other peptides from all the other proteins. To overcome this problem, some form of sample enrichment is required. Separation of proteins based on size can be used before trypsin digestion. Even more effective is immune affinity purification of the desired Tg target peptide(s) from all the other peptides. A good antibody against the target peptide can achieve dramatic enrichment. Using one or both of these enrichment techniques, Tg measurement by MS becomes possible [34]. By limiting oneself to one or two peptide targets that ionize well, optimizing chromatography, and using selective reaction monitoring on high sensitivity mass filtering tandem MSs, rather than the higher resolving, but (typically) less sensitive, high resolution instruments used in proteomics research, analytical sensitivities can be achieved that are slightly higher than modern immunometric Tg immunoassays [35] (Figure 1).

Figure 1 Schematic depiction of the workflow for current (high-pressure) liquid chromatography tandem mass spectrometry (LC-MS/MS; HPLC-MS/MS) measurement of Tg.
Figure 1

Schematic depiction of the workflow for current (high-pressure) liquid chromatography tandem mass spectrometry (LC-MS/MS; HPLC-MS/MS) measurement of Tg.

Standardization and harmonization of Tg assays

Tg is a large (660 kDa), highly glycosylated dimeric molecule that is heterogeneous in serum due to differential splicing of Tg mRNA as well as carbohydrate and iodide heterogeneity. In addition, biosynthesis of the mature Tg molecule may become deregulated in thyroid tumor cells resulting in differences in the structure of circulating Tg protein. These changes can lead to exposure or masking of epitopes and hence differences in Tg immunoreactivity [36]. Different Tg assays employ a number of antibodies against Tg with varying specificity for different epitopes. Potentially this can result in variability in the measurement of different Tg isoforms in the patient’s specimen and ultimately to differences in Tg concentration reported by the assays [20, 37]. Early international collaborative studies showed that serum Tg concentrations varied by as much as 40%–60% between methods [38]. The introduction and use of the BCR® 457 has significantly reduced inter-method variability to about 30%, but has not eliminated it completely [39, 40]. Consequently, any change in Tg assay has the potential to disrupt serial monitoring and prompt inappropriate clinical decisions. For longitudinal consistency of clinical care, consecutive measurements of Tg concentrations should be performed in the same laboratory using the same assay each time. If an assay change is unavoidable, a new baseline of the individual patient’s serum Tg concentrations should be established through parallel Tg measurements using both the old and the new assay [11]. Furthermore, internal and external quality control programs, including samples at low and very low Tg concentrations, are of pivotal importance for checking the precision, reproducibility (internal quality control) and accuracy (e.g., lack of bias of analytical results) of assays to ensure optimal patient care. Laboratories providing Tg measurement are required to participate in a certified national or international program of quality assurance [41].

Finally, it is worth bringing to attention that BCR® 457 was produced around 20 years ago, and although high temperature accelerated stability was investigated as part of the project [39] it cannot be excluded that long-term storage might have changed the epitope composition by degradation or unfolding. By using highly sensitive Tg methods a very precise calibrations could be more important than using the previous less sensitive ones, without a requirement for the low concentrations. It may thus be important to consider investigating consequences of the current assay calibration and comparability by the use of newer Tg methods, and perhaps it is time to consider producing a new calibrator.

Interferences

Depending on the population studied, the assay used and the definition of a positive test result, up to 25%–30% of patients with DTC have a positive test for TgAb at the time of diagnosis [16, 42]. Some authors proposed to measure serum Tg by RIA and IMA on the same sample as discordant result suggest TgAb interference [16, 43]. However, this method is technically demanding and time-consuming and, consequently, it is rarely employed in clinical practice. There are two approaches to detecting TgAb: recovery of added exogenous Tg or measurement of TgAb by immunoassay [17, 36, 44, 45]. Whereas the first approach has a relatively wide reference range and is very dependent on experimental conditions (e.g., concentration of added Tg, two measurements of Tg that increases overall imprecision), the second approach will, depending on the assay, detect most TgAb-positive patients. In this context, a positive result is often defined as a result above the upper limit of the reference range. However, for thyroidectomized patients, lacking antigen for TgAb generation, due to limitations, the cut-off value for detection of analytical interference is probably much lower [45, 46].

In addition, not all patients with positive TgAb will show interference by measurement of recovery or discordance between Tg results using different assays [45]. Any TgAb assay should be standardized against the First International Reference Preparation 65/93, but, due to the heterogeneous nature of TgAb, no single assay can predict with absolute certainty whether TgAb in a given sample will interfere with Tg measurement [47]. Theoretically, a simple relationship exists between Tg and TgAb and the higher the TgAb concentration, the higher the Tg concentration that can be concealed by TgAb. Sometimes, however, seemingly low concentrations of TgAb may be associated with strong interference and conversely patients with high concentrations of TgAb show no evidence of interference with the Tg measurement [45]. Recently, new recovery tests using a low Tg concentration (i.e., 5–10 μg/L) were introduced [48]. While the performance of these ‘low concentration recovery tests’ (e.g., mini-recovery tests) has yet to be investigated extensively in patients with DTC, they may prove to serve as an addition to TgAb measurement [45]. All in all, any sample with a positive TgAb result and/or abnormal recovery test/discordant Tg result by different analytical methods (e.g., immunometric assay vs. radioimmunoassay) should be considered as unreliable for measuring serum Tg concentrations in patients with DTC. In addition, a small percentage of patients (1%–3% in the literature) show interference with Tg measurement due to HAb [49, 50]. These can bind animal antigens and form a bridge between capture and detection antibody leading to a falsely elevated (or, rarely, falsely decreased) Tg measurement in immunometric assays [51]. HAb interference may be detected either by recovery measurement or measurement of Tg in serially diluted sera (providing Tg concentrations are sufficiently high). In addition, as HAb interference is generally assay specific, the use of an alternate assay may both identify a false-positive sample and provide the correct test value. However, the method of choice, which is more specifically geared towards heterophile antibody interference, is to pre-treat a serum aliquot with proprietary blocking agents then compare the Tg result with an aliquot that was not pre-treated [45].

The majority of patients with DTC show an elevated pre-operative serum Tg, but the predictive role of this measurement is debatable as immunoassays cannot detect a difference between Tg from normal thyroid tissue and that secreted by thyroid cancer tissue. In clinical practice, however, patients with clear disease foci or significant thyroid remnants but undetectable serum Tg levels despite a negative TgAb test results are occasionally encountered [52, 53]. This may occur when: the spatial conformation of Tg is changed leading to decreased immunoreactivity; the ability to secrete Tg is lost; the TgAb result is a false-negative or inappropriate cut-off values were used. To overcome this problem, an in vivo recovery test has been proposed: measurement of Tg and TgAb in any patient referred for thyroidectomy due to a suspicion of DTC [11, 45]. This strategy could provide ‘baseline’ Tg and TgAb concentrations, which, as stated by NACB guidelines [20], could theoretically allow assessment of the reliability of Tg and TgAb measurements after thyroidectomy, although the clinical value of this in vivo test has never been verified.

Several analytes have been tested as alternatives to Tg measurement in TgAb positive patients, such as Tg mRNA, thyrotropin receptor mRNA or thyroid peroxidase mRNA. However, in most studies it proved impossible to achieve clinically useful sensitivity and specificity levels [54].

Recently MS emerged as a suitable method to overcome interferences in Tg measurement; in fact trypsin digests all proteins in a sample, including Tg and any TgAb or HAb, by cleaving them at predictable sites. One can then specifically look for (one or several) tryptic peptides that are proteotypic for Tg (based on predicted cleavage), without any interference by TgAb. Kato et al. published a proof of principle for the tryptic digestion approach to overcome antibody interferences in Tg assays almost 15 years ago [32]. However, this group used an in-house ELISA, rather than MS, to detect Tg proteotypic peptides after digestion. Unfortunately, this system only achieved a LOD of approximately 10 μg/L, insufficient for modern clinical use. In 2008 Hoofnagle et al. revisited this approach, using tryptic digestion, followed by immuno-capture of proteotypic peptides and detection of the captured peptides by MS [33]. Several other groups have since replicated this work [34, 35, 55]. The LOQ of these Tg MS assays is 0.5–1 μg/L, and all have shown excellent comparability (R2>0.95, slope 0.8–1.2) to immunoassays in samples that are TgAb negative. In TgAb positive samples, which have detectable Tg by hsTg immunoassay (FS/LOQ ∼0.1 μg/L), all published MS Tg assays still correlated with a similar R2 with the immunoassay, but demonstrated a slope of ∼1.5, consistent with systematic under-recovery of Tg (of about 50%–60%) in the immunoassays. Finally, in samples that are TgAb positive, but have an undetectable Tg by sensitive immunoassay the Tg MS assays can detect Tg in 20%–25% of cases (Figure 2) [55].

Figure 2 Schematic workflow for patients with positive TgAb.IMA, immunometric assay; MS, mass spectrometry; PET/CT, positron emission tomography/computed tomography; RIA, radioimmunoassay; TgAb, thyroglobulin autoantibodies; WBS, whole body scan.
Figure 2

Schematic workflow for patients with positive TgAb.

IMA, immunometric assay; MS, mass spectrometry; PET/CT, positron emission tomography/computed tomography; RIA, radioimmunoassay; TgAb, thyroglobulin autoantibodies; WBS, whole body scan.

The principle drawbacks of MS-based Tg assays are currently the complex and manual workflows, the long turn-around-time, largely owing to the several hours of tryptic digest, and the currently slightly higher LOD and LOQ when compared to hsTg immunoassays. Due to these limitations, Tg MS assays should not replace Tg immunoassays and MS testing should be reserved for TgAb positive patients. If a patient is known to be TgAb positive, Tg MS testing may be performed instead of immunoassay testing. For patients with uncertain TgAb status, TgAb should be measured before Tg testing. For patients who prove TgAb negative, Tg testing should then be reflexed to a Tg immunoassay, while samples from TgAb positive patients should be routed to Tg measurement by MS [35, 55].

Serum levels of TgAb are not correlated with the tumor load of the patient but rather indicate the activity of the immune system. Furthermore, the mere presence of TgAb in serum has thus far not conclusively been shown to correlate with a worse or better overall prognosis. However, TgAb can be used as ‘surrogate tumor marker’ as disease-free patients with high TgAb concentrations typically display a slow progressive TgAb decline over years (but may not achieve full TgAb-negativity, possibly because of the long-lived memory of plasma cells). Although the most common change in TgAb status is positive to negative in response to successful treatment the persistence of TgAb, a rising TgAb trend, the failure of TgAb to fall or a de-novo TgAb appearance is highly suspicious for active disease [11, 12, 45].

Clinical impact of new highly sensitive Tg assays

There is a wide consensus among various guidelines that Tg is an important, sensitive method for monitoring DTC patients for the presence of residual or recurrent disease after total thyroidectomy and 131I remnant ablation. A hindrance for the preeminent position of Tg measurement thus far consisted of the need, recommended in all guidelines, for TSH stimulation in order to achieve optimal sensitivity. A ‘negative’ TSH-stimulated Tg measurement in combination with a negative clinical examination, negative neck US and, when indicated, negative additional imaging procedures, predicts a very low risk of recurrence in both low- and high-risk patients [56, 57]. Once these findings have been established, routine DTC follow-up should therefore consist of periodic clinical examination combined with neck ultrasound and Tg measurement on LT4 medication [2, 3]. An important question that needs to be answered is whether the latter is still the case for hsTg assays. Previous studies have already suggested that the additional yield of DTC recurrence through rhTSH stimulation was only 0.8% using Tg assays with a FS of about 1 μg/L [58, 59]. In a study by Giovanella et al. [60] the clinical sensitivity for the detection of DTC recurrence could be increased to 92% by employing a Tg assay with a FS of 0.4 μg/L while specificity did not change. In a further study by the same group [61], unstimulated serum Tg measurements were associated with a 96% negative predictive value (NPV) in 117 low-risk DTC, which increased to 99% when combined with neck US. In this study, rhTSH stimulated Tg measurement only detected one additional recurrence in 104 patients with an undetectable unstimulated Tg. Therefore, even using elss sensitive assay technology, the additional value of an rhTSH stimulated Tg measurement is likely more modest than guidelines led us to believe, especially in low-risk patients. More recently, a number of studies were performed to investigate the diagnostic performance of basal serum hsTg measurement in the follow-up of DTC patients [7–10, 18, 24, 25, 29, 62, 63]. Smallridge et al. [64] evaluated 194 patients who also underwent rhTSH stimulation by using the Beckmann Coulter Access Tg assay (FS 0.1 μg/L). Of the 80 patients with a basal Tg below 0.1 μg/L, two (2.5%) had an rhTSH-stimulated Tg concentration above 2 μg/L. No patient had concurrent imaging suggestive of local recurrence or distant metastasis. However, if unstimulated hsTg was 0.1–0.5 or 0.6–2.0 μg/L, rhTSH-stimulated Tg was above 2 μg/L in 24.2% and 82.4% of cases, respectively. Recently, Giovanella et al. [10] reviewed and meta-analyzed data from nine studies including 3178 DTC patients and confirmed the very high NPV (98%–100%) of an undetectable basal hsTg (e.g., <0.1 μg/L). Nonetheless, these assays also have an adequate sensitivity for detection of recurrent disease (88%–98%). Therefore, hsTg assays seem to be able to obviate the need for measurement of TSH-stimulated Tg concentrations. Unfortunately, the improved sensitivity is associated with an unsatisfactory clinical specificity and positive predictive value (PPV). Patients with a low detectable basal hsTg (i.e., between 0.1 and 1 μg/L) can likely be considered as ‘disease-free’ after a negative TSH-stimulated Tg measurement. This means that the latter group of patients (who fortunately are a minority of the total DTC population) should be submitted to a TSH-stimulated Tg measurement [10–12] (Figure 3). However, although the low frequency of DTC recurrences impacts the ability to study PPVs, the PPV of an rhTSH-stimulated Tg above 1–2 μg/L appears comparable to a basal hsTg above 0.10–0.20 μg/L [19]. In addition, the trend in basal hsTg, measured when TSH is lowered/suppressed at constant level, should reflect changes in thyroid tissue mass and thus provide a sensitive parameter for disease [19, 24, 61, 65]. This is also supported by a growing number of studies showing the prognostic utility of monitoring the basal hsTg trend and Tg doubling time [66–69]. A further limitation to the available evidence is that most patients who were enrolled in the studies were affected by low-risk DTC. Data on patients with intermediate- and high-risk tumors are less robust, indicating that the above described approach could be restricted, at least for the moment, to low-risk DTC patients [11]; further studies should be performed in a broader spectrum of high-risk patients.

Figure 3 Proposed flowchart for using highly sensitive Tg testing into differentiated thyroid carcinoma follow-up.DTC, differentiated thyroid carcinoma; TgAb, thyroglobulin autoantibodies. hsTg expressed in μg/L.
Figure 3

Proposed flowchart for using highly sensitive Tg testing into differentiated thyroid carcinoma follow-up.

DTC, differentiated thyroid carcinoma; TgAb, thyroglobulin autoantibodies. hsTg expressed in μg/L.

FNAC-Tg from lymph nodes: clinical role, validation and quality assessment

Despite the excellent prognosis of most DTC patients, a minority of them will develop a recurrence (or multiple recurrences in the long term). Whereas follicular thyroid carcinoma metastatizes to distant organs as lung or bone, the papillary thyroid carcinoma primarily disseminates to regional lymph nodes via the lymphatic vessels. Cervical recurrences may be detected clinically but are most frequently discovered on neck ultrasound (US); additional imaging procedures (i.e., radioiodine scan, PET/CT scan) are performed when distant metastases are expected. In case of suspicious US findings, fine-needle aspiration cytology (FNAC) is generally required to confirm or exclude recurrence by cytology [2, 4, 5]. Unfortunately, FNAC samples may be inadequate or even false-negative, especially in lymph nodes with small metastases, partial involvement or cystic changes. To improve the diagnostic yeld of FNAC, several authors proposed measurement of Tg in aspirates (FNAC-Tg) [70–73]. In fact, FNAC-Tg measurement using a cut-off level of 1 μg/L provides higher sensitivity and specificity than cytology in athyrotic patients with papillary thyroid carcinoma and concurrent suspicious US and increased serum Tg [23, 26].

The measurement of Tg in non-serum/plasma samples is problematic from an analytical perspective, not least because of the lack of experimental data to support the validity of results and the absence of formal support for this application by commercial manufacturers of Tg assays. The onus is thus on laboratories to determine whether analysis of non-serum/plasma samples is listed as an ‘Intended use’ in the Information for Users provided by manufacturers. In its absence, full analytical validation to regulatory standards by laboratories is required. Several potential confounding factors should be considered [27].

There may be differences in antigen structure and immunoreactivity. Furthermore, there may be matrix effects associated with the washout fluid not found in serum/plasma. Also the high dose hook effect requires exclusion. The TgAb can potentially be present in the washout fluid and affect the immunoassay. As a consequence laboratory investigation might include the following actions.

Pre-analytical

Samples are collected by fine needle aspiration for cytology and after dispensation of the sample onto the appropriate slides, the needle is washed out. Published protocols vary, often giving minimal detail, such as volume of washout and fluid as ‘normal saline’ or phosphate-buffered saline. Where a serous fluid has been aspirated, dilution with saline may be omitted. Samples are collected from cervical lymph nodes or space occupying lesions in the thyroid bed [72].

  1. Is the sample representative of the lymph node/thyroid bed? Is the first, second, etc needle washout? How many lymph nodes were aspirated? (Action – Document).

  2. Is the sample homogenous or does it contain cellular/particulate matter? (Action – Mix, centrifuge sample).

  3. Does the sample contain hemoglobin? (Action – Discard, Tg contamination from peripheral blood during the collection procedure may invalidate the result).

  4. Is the sample stable in the washout fluid and in any assay diluents? Is it stable in the collection tube used? And at what temperature? (Action – Establish by experiment, may be assay specific).

Analytical

  1. Are there any matrix effects from the fluid used to wash-out the collection needle? This may be assay specific. (Action – Determine the optimal fluid).

  2. Concentrations of Tg reported in FNAC-washout fluid range from 0.1–>1000 ug/L. Is there a high dose hook effect? (Action – Establish linearity, assay all samples before and after dilution in an appropriate diluent).

  3. Assay recovery from washout fluid of added Tg. Over-recovery of 100%–140% has been reported [6].

  4. Endogenous Tg antibodies have been reported to affect and also to have no effect on the measurement of Tg in FNAC samples [28, 65, 71]. However, some of these studies are based on TgAb measurements in the patient’s serum or by inappropriate analytical methods for TgAb, such as agglutination methods. The analytically relevant question is whether there are TgAb in the sample being analyzed and whether these interfere in the immunoassay [30]. Measurement of TgAb themselves in washout fluid may equally be problematic, but attempts should be made to determine/validate the assays before use [26].

  5. Quality assessment. Serum quality control materials may not adequately reflect the performance of Tg assays in washout fluids unless a serum-based diluent has been used as the washout fluid. Laboratories may wish to exchange washout samples with other laboratories to compare results, having established the stability of the samples.

Post-analytical factors

  1. Consider whether results are reported per volume of washout fluid or per aspirate.

  2. Consider patient-related factors – serum TSH concentration, pre/post-thyroidectomy, histopathological diagnosis.

Close collaboration between laboratory and clinician is required in establishing protocols for the sampling and collection of samples and their accurate analysis.

Conclusions

The post-surgical follow-up of DTC is aimed at early identification of the small proportion of patients who have residual disease or will develop recurrence. In the absence of TgAb and heterophile antibodies, Tg measurements are nowadays the reference standard for clinical management of patients previously treated for DTC. Current clinical guidelines are still based on study results obtained using Tg assays with a FS of about 1 μg/L. Using such assays a TSH-stimulated Tg measurement was mandatory to achieve sufficient clinical sensitivity for the detection of persistent and/or recurrent disease. Even though the introduction of hsTg assays is not without challenges, especially where it comes to the sensitivity of current assays for the detection of interfering autoantibodies against Tg, there is an increasing body of evidence that an undetectable highly sensitive Tg during LT4 treatment is sufficient with a high NPV to forgo TSH stimulation in low-risk DTC patients. Patients with slightly increased basal hsTg (e.g., 0.1–1 μg/L) concentration should undergo a TSH-stimulated Tg measurement.

Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

Financial support: LG, FAV and PC served on an advisory board of Roche AG (September 2012–September 2013); LG received research support and accepted speaker’s fees from Roche AG; UFR serves on an advisory board of Genzyme; UFR has received research support from Genzyme; LG, FAV and UFR have accepted speaker’s fees from Genzyme.

Employment or leadership: None declared.

Honorarium: None declared.

Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication.


Corresponding author: PD Dr. med. Luca Giovanella, Department of Nuclear Medicine and Thyroid Centre, Oncology Institute of Southern Switzerland Via Ospedale 12, 6500 Bellinzona, Switzerland, Phone: +41 91 8118672, Fax: +41 91 8118250, E-mail:

About the authors

Luca Giovanella

Luca Giovanella MD, PhD is the Chief of the Department of Nuclear Medicine-PET/CT and Thyroid Centre at Oncology Institute of Southern Switzerland (IOSI) and the Scientific Advisor of the Department of Clinical Chemistry and Laboratory Medicine at Ente Ospedaliero Cantonale in Bellinzona and Lugano (Switzerland). He serves as Privat-Dozent (PD) at University of Zürich (Switzerland). His clinical and scientific activities are mainly focused on the diagnosis and treatment of thyroid diseases.

Ulla Feldt-Rasmussen

Ulla Feldt-Rasmussen MD, is Professor at University of Copenhagen and Chief Physician of Medical Endocrinology at National University Hospital. Her research interests involve the thyroid gland, cancer and autoimmunity. She has published more than 320 papers in peer-reviewed journals, organized international meetings and postgraduate courses, and led several European thyroid-related projects. Dr. Feldt-Rasmussen is reviewer and editorial board member of many international journals, belongs to numerous international professional organizations and has served as Secretary-Treasurer of European Thyroid Association, and as President of ETA Cancer Research Network. She has received several prestigious prizes and was Haines lecturer at Mayo Clinic in 2011, and Associate Professor at University of Messina, Italy.

Frederik A. Verburg

Frederik A. Verburg MD, PhD studied medicine at the universities of Leuven (Belgium) and Utrecht (The Netherlands) from 1997 to 2004. In 2008 he obtained his PhD in medicine at the University of Utrecht. He trained as a nuclear medicine physician at the University Medical Center Utrecht, The Netherlands, and the University Hospital Würzburg, Germany, from 2005 to 2010. From 2011 to the present he is working as a consultant nuclear medicine physician and Assistant Professor of Nuclear Medicine at the RWTH University Hospital Aachen in Aachen, Germany.

Stephan K. Grebe

Stefan K. G. Grebe MD, PhD is Professor of Laboratory Medicine and Pathology (DLMP), Mayo Clinic College of Medicine. He trained in Internal Medicine and Endocrinology and joined the DLMP of the Mayo Clinic as a Co-Director of the Endocrine Laboratory in 2002. He served as Chair of the Division of Clinical Biochemistry and Immunology from 2006 to 2014. His main clinical and research interests are currently focused on thyroid cancer tumor markers and clinical mass spectrometry.

Mario Plebani

Mario Plebani obtained his medical degree summa cum laude from the Medical School of the University of Padova in 1975. He completed residency training and specialization in Laboratory Medicine (1978), and subsequently in Gastroenterology (1983), at the same University. He is full Professor of Clinical Biochemistry and Clinical Molecular Biology at the University of Padova, School of Medicine, Chief of the Department of Laboratory Medicine at the University-Hospital of Padova, and Chief of the Center of Biomedical Research (a specialized Center for quality in laboratory medicine for the Veneto Region). Currently, he is also Director of the Postgraduate School in Clinical Biochemistry at the Medical School of the Padova University and President of the Course for Medical Technologists at the same Medical School. He served as President of the International Society of Enzymology (ISE) for 4 years (2004–2008) and as President of the Italian Society of Clinical Biochemistry and Molecular Clinical Biology for 5 years (in 2003 and from 2007 to 2009). He is the Chairman of the IFCC Working group on ‘Laboratory errors and patient safety’ (WG LEPS) and in 2008 received the AACC Award for Outstanding Clinical Laboratory Contributions to Improving Patient Safety. Dr. Plebani is Editor in Chief of CCLM (Clinical Chemistry and Laboratory Medicine), and Associate editor of CRC Clinical Laboratory Sciences, and International Journal of Biological Markers. His main areas of research are quality in laboratory medicine, biomarkers in cancer and cardiovascular diseases, and in vitro allergy diagnostics.

Penelope M. Clark

Penelope M. Clark PhD, is an Honorary Consultant Clinical Scientist and Senior Lecturer at The Queen Elizabeth Hospital Birmingham and the University of Birmingham. She headed the Regional Endocrine Laboratory with responsibility for national services for thyroid disease and pituitary disease including thyroid cancer. She is a Fellow of The Royal College of Pathologists and The Association for Clinical Biochemistry and Laboratory Medicine.

References

1. Grebe SK. Diagnosis and management of thyroid carcinoma: focus on serum thyroglobulin. Expert Rev Endocrinol Metab 2009;4:25–43.10.1586/17446651.4.1.25Search in Google Scholar

2. Pacini F. Follow-up of differentiate thyroid cancer. Eur J Nucl Med 2002;29:492–6.10.1007/s00259-002-0847-9Search in Google Scholar PubMed

3. Giovanella L. Highly sensitive thyroglobulin measurements in differentiated thyroid carcinoma management. Clin Chem Lab Med 2008;46:1067–73.10.1515/CCLM.2008.212Search in Google Scholar PubMed

4. Pacini F, Schlumberger M, Dralle H, Elisei R, Smit JW, Wiersinga W, et al. European consensus for the management of patients with differentiated thyroid carcinoma of the follicular epithelium. Eur J Endocrinol 2006;154:787–803.10.1530/eje.1.02158Search in Google Scholar PubMed

5. Cooper DS, Doherty GM, Haugen BR, Kloos RT, Lee SL, Mandel SJ, et al. Revised American Thyroid Association management guidelines for patients with thyroid nodules and differentiated thyroid cancer. Thyroid 2009;19:1167–214.10.1089/thy.2009.0110Search in Google Scholar PubMed

6. Haugen BR, Ridgway EC, McLaughlin BA, McDermott MT. Clinical comparison of whole-body radioiodine scan and serum thyroglobulin after stimulation with recombinant human thyrotropin. Thyroid 2002;12:37–43.10.1089/105072502753451959Search in Google Scholar PubMed

7. Wartofsky L. Using baseline and recombinant human TSH-stimulated Tg measurements to manage thyroid carcinoma without diagnostic (131)I scanning. J Clin Endocrinol Metab 2000;87:1486–9.10.1210/jcem.87.4.8445Search in Google Scholar PubMed

8. Wunderlich G, Zöphel K, Crook L, Smith S, Smith BR, Franke WG. A high-sensitivity enzyme-linked immunosorbent assay for serum thyroglobulin. Thyroid 2001;11:819–24.10.1089/105072501316973064Search in Google Scholar PubMed

9. Iervasi A, Iervasi G, Bottoni A, Boni G, Annicchiarico C, Di Cecco P, et al. Diagnostic performance of a new highly sensitive thyroglobulin immunoassay. J Endocrinol 2004;182:287–94.10.1677/joe.0.1820287Search in Google Scholar PubMed

10. Giovanella L, Treglia G, Sadeghi R, Trimboli P, Ceriani L, Verburg FA. Unstimulated high-sensitive thyroglobulin in follow-up of differentiated thyroid cancer patients: a meta-analysis. J Clin Endocrinol Metab 2014;99:440–7.10.1210/jc.2013-3156Search in Google Scholar PubMed

11. Giovanella L, Clark P, Chiovato L, Duntas LH, Elisei R, Feldt-Rasmussen U, et al. Thyroglobulin measurement using highly sensitive assays in patients with differentiated thyroid cancer: a clinical position paper. Eur J Endocrinol 2014;171:R33–46.10.1530/EJE-14-0148Search in Google Scholar PubMed PubMed Central

12. Perros P, Boelaert K, Colley S, Evans C, Evans RM, Gerrard G, et al. British Thyroid Association Guidelines for the management of thyroid cancer, 3rd ed. Available from: www.british-thyroid-association.org/Guidelines/Docs/BTA_DTC_guidlines.pdf Accessed 7 June, 2014.10.1111/cen.12515Search in Google Scholar PubMed

13. Sadler WA, Smith MH. Use and abuse of imprecision profile. Some pitfall illustrated by comparing and plotting confidence intervals. Clin Chem 1990;36:1346–50.10.1093/clinchem/36.7.1346Search in Google Scholar

14. Sheehan C, He J, Smith M. Method evaluation – a practical guide. In: Wild D, editor. The immunoassay handbook, 4th ed. Oxford: Elsevier, 2013:395–402.Search in Google Scholar

15. Ross HA, Netea-Maier RT, Schakenraad E, Bravenboer B, Hermus AR, Sweep FC. Assay bias may invalidate decision limits and affect comparability of serum thyroglobulin assay methods: an approach to reduce interpretation differences. Clin Chim Acta 2008;394:104–9.10.1016/j.cca.2008.04.020Search in Google Scholar PubMed

16. Spencer CA, Takeuchi M, Kazarosyan M, Wang CC, Gutler RB, Singer PA, et al. Serum thyroglobulin autoantibodies: prevalence, influence on serum thyroglobulin measurement and prognostic significance in patients with differentiated thyroid carcinoma. J Clin Endocrinol Metab 1998;83:1121–7.10.1210/jc.83.4.1121Search in Google Scholar

17. Verburg FA, Hartmann D, Grelle I, Giovanella L, Buck AK, Reiners C. Relationship between antithyroglobulin autoantibodies and thyroglobulin recovery rates using different thyroglobulin concentrations in the recovery buffer. Horm Metab Res 2013;45:728–35.10.1055/s-0033-1349890Search in Google Scholar PubMed

18. Castagna MG, Tala Jury HP, Cipri C, Belardini V, Fioravanti C, Pasqui L, et al. The use of ultrasensitive thyroglobulin assays reduces but does not abolish the need for TSH stimulation in patients with differentiated thyroid carcinoma. J Endocrin Invest 2011;34:e219–23.Search in Google Scholar

19. Schlumberger M, Hitzel A, Toubert ME, Corone C, Troalen F, Schlageter MH, et al. Comparison of seven serum thyroglobulin assays in the follow-up of papillary and follicular thyroid cancer patients. J Clin Endocrinol Metab 2007;92:2487–95.10.1210/jc.2006-0723Search in Google Scholar PubMed

20. Demers LM, Spencer CA, editors. National Academy of Clinical Biochemistry Guideline (Archived). Laboratory Support for the diagnosis and management of thyroid disease. Available from: http://www.aacc.org/members/nacb/Archive/LMPG/ThyroidDisease/Pages/ThyroidDiseasePDF.aspx. Accessed 16 June, 2014.Search in Google Scholar

21. Armbruster DA, Pry T. Limit of blank, limit of detection and limit of quantitation. Clin Biochem Rev 2008;29(Suppl 1):S49–52.Search in Google Scholar

22. CLSI. Protocols for determination of limits of detection and limits of quantification: approved guidelines, 2nd ed. EP17-A. CLSI: Wayne, PA, 2012.Search in Google Scholar

23. Giovanella L, Bongiovanni M, Trimboli P. Diagnostic value of thyroglobulin assay in cervical lymph node fine-needle aspirations for metastatic differentiated thyroid cancer. Curr Opin Oncol 2013;25:6–13.10.1097/CCO.0b013e32835a9ab1Search in Google Scholar PubMed

24. Rosario PW, Dos Santos JB, Calsolari MR. Follow-up of patients with low-risk papillary thyroid carcinoma and undetectable basal serum thyroglobulin after ablation measured with a sensitive assay: a prospective study. Horm Metab Res 2013;45:911–4.10.1055/s-0033-1347264Search in Google Scholar PubMed

25. Malandrino P, Latina A, Marescalco S, Spadaro A, Regalbuto C, Fulco RA, et al. Risk-adapted management of differentiated thyroid cancer assessed by a sensitive measurement of basal serum thyroglobulin. J Clin Endocrinol Metab 2011;96:1703–9.10.1210/jc.2010-2695Search in Google Scholar PubMed

26. Grani G, Fumarola A. Thyroglobulin in lymph node fine-needle aspiration washout: a systematic review and meta-analysis of diagnostic accuracy. J Clin Endocrinol Metab 2014;99:1970–62.10.1210/jc.2014-1098Search in Google Scholar

27. Boot CS, Mahon BS, Bramhall SR, Clark PM. Validation of carcinoembryonic antigen and carbodhydrate antigen 19-9 measurements in pancreatic cyst fluid with a serum-based immunoassay. Clin Chem 2010;56:1351–61.10.1373/clinchem.2010.146373Search in Google Scholar

28. Jeon MJ, Park JW, Han JM, Yim JH, Song DE, Gong G, et al. Serum antithyroglobulin antibodies interfere with thyroglobulin detection in fine-needle aspirates of metastatic neck nodes in papillary thyroid carcinoma. J Clin Endocrinol Metab 2013;98:155–60.10.1210/jc.2012-2369Search in Google Scholar

29. Rosario PW, Purisch S. Does a highly sensitive thyroglobulin (Tg) assay change the clinical management of low-risk patients with thyroid cancer with Tg on T4 <1 ng/ml determined by traditional assays? Clin Endocrinol 2008;68:338–42.10.1111/j.1365-2265.2007.03043.xSearch in Google Scholar

30. Giovanella L, Ceriani L, Suriano S, Crippa S. Thyroglobulin measurement on fine-needle washout fluids: influence of sample collection methods. Diagn Cytopathol 2009;37:42–4.10.1002/dc.20964Search in Google Scholar

31. Giovanella L, Ceriani L, Ghelfo A. Redefining functional sensitivity of thyroglobulin assay on Immulite platform: implications in thyroid cancer management. Clin Chem Lab Med 2007;45:1523–4.10.1515/CCLM.2007.309Search in Google Scholar

32. Kato R, Maruyama M, Sekino T, Kasuga Y. A new assay for thyroglobulin concentration in serum using monoclonal antibodies against synthetic peptides. Clin Chim Acta 2000;298:69–84.10.1016/S0009-8981(00)00258-8Search in Google Scholar

33. Hoofnagle AN, Becker JO, Wener MH, Heinecke JW. Quantification of thyroglobulin, a low-abundance serum protein, by immunoaffinity peptide enrichment and tandem mass spectrometry. Clin Chem 2008;54:1796–804.10.1373/clinchem.2008.109652Search in Google Scholar PubMed PubMed Central

34. Clarke NJ, Zhang Y, Reitz RE. A novel mass spectrometry-based assay for the accurate measurement of thyroglobulin from patient samples containing antithyroglobulin autoantibodies. J Investig Med 2012;60:1157–63.10.2310/JIM.0b013e318276deb4Search in Google Scholar PubMed

35. Grebe SK. Soluble thyroid tumor markers – old and new challenges and potential solutions. NZ Instit Med Lab Sci J 2013;67:76–87.Search in Google Scholar

36. Spencer CA, Lopresti JS. Measuring thyroglobulin and thyroglobulin autoantibody in patients with differentiated thyroid cancer. Nature Clin Pract Endocrinol Metab 2008; 4:223–33.10.1038/ncpendmet0757Search in Google Scholar PubMed

37. Clark PM. Laboratory services for thyroglobulin and implications for monitoring of differentiated thyroid cancer. J Clin Pathol 2009;62:402–6.10.1136/jcp.2008.058024Search in Google Scholar PubMed

38. Feldt-Rasmussen U, Schlumberger M. European inter-laboratory comparison of serum thyroglobulin measurement. J Endocrinol Invest 1988;11:175–81.10.1007/BF03350129Search in Google Scholar PubMed

39. Feldt-Rasmussen U, Profilis C, Colinet E, Black E, Bornet H, Bourdoux P, et al. Human thyroglobulin reference material (CRM 457). 1st Part: assessment of homogeneity, stability and immunoreactivity. Ann Biol Clin (Paris) 1996;54:337–42.Search in Google Scholar

40. Feldt-Rasmussen U, Profilis C, Colinet E, Black E, Bornet H, Bourdoux P, et al. Human thyroglobulin reference material (CRM 457). 2nd Part: physicochemical characterization and certification. Mol Cell Endocrinol 1966;116:165–72.Search in Google Scholar

41. Clark P, Franklyn J. Can we interpret serum thyroglobulin results? Ann Clin Biochem 2012;49:313–22.10.1258/acb.2011.011248Search in Google Scholar PubMed

42. Feldt-Rasmussen U, Rasmussen AK. Autoimmunity in differentiated thyroid cancer: significance and related clinical problems. Hormones (Athens) 2010;9:109–17.10.14310/horm.2002.1261Search in Google Scholar PubMed

43. Crane MS, Strachan MW, Toft AD, Beckett GJ. Discordance in thyroglobulin measurements by radioimmunoassay and immunometric assay: a useful means of identifying thyroglobulin assay interference. Ann Clin Biochem 2013; 50:421–32.10.1177/0004563213480492Search in Google Scholar PubMed

44. Persoon AC, Links TP, Wilde J, Sluiter WJ, Wolffenbuttel BH, van den Ouweland JM. Thyroglobulin (Tg) recovery testing with quantitative Tg antibody measurement for determining interference in serum Tg assays in differentiated thyroid carcinoma. Clin Chem 2006;52:1196–9.10.1373/clinchem.2005.060103Search in Google Scholar PubMed

45. Verburg FA, Luster M, Cupini C, Chiovato L, Duntas LH, Elisei R, et al. Implications of thyroglobulin antibody positivity in patients with differentiated thyroid cancer: a clinical position paper. Thyroid 2013;23:1211–25.10.1089/thy.2012.0606Search in Google Scholar PubMed

46. Spencer C, Petrovic I, Fatemi S. Current thyroglobulin autoantibody (TgAb) assays often fail to detect interfering TgAb that can result in the reporting of falsely low/undetectable serum Tg IMA values for patients with differentiated thyroid cancer. J Clin Endocrinol Metab 2011;96:1283–91.10.1210/jc.2010-2762Search in Google Scholar PubMed

47. Taylor KP, Parkington D, Bradbury S, Simpson HL, Jefferies SJ, Halsall DJ. Concordance between serum thyroglobulin antibody assays. Ann Clin Biochem 2011;48:367–9.10.1258/acb.2011.010248Search in Google Scholar PubMed

48. Giovanella L, Imperiali M, Verburg FA, Ceriani L. Evaluation of the BRAHMS Kryptor(®) thyroglobulin minirecovery test in patients with differentiated thyroid carcinoma. Clin Chem Lab Med 2013;51:449–53.10.1515/cclm-2012-0378Search in Google Scholar PubMed

49. Preissner CM, O’Kane DJ, Singh RJ, Morris JC, Grebe SK. Phantoms in the assay tube: heterophile antibody interferences in serum thyroglobulin assay. J Clin Endocrinol Metab 2003;88:3069–74.10.1210/jc.2003-030122Search in Google Scholar PubMed

50. Verburg FA, Wäschle K, Reiners C, Giovanella L, Lentjes EG. Heterophile antibodies rarely influence the measurement of thyroglobulin and thyroglobulin antibodies in differentiated thyroid cancer patients. Horm Metab Res 2010;42:736–9.10.1055/s-0030-1254132Search in Google Scholar PubMed

51. Giovanella L, Keller F, Ceriani L, Tozzoli R. Heterophile antibodies may falsely increase or decrease thyroglobulin measurement in patients with differentiated thyroid carcinoma. Clin Chem Lab Med 2009;47:952–4.10.1515/CCLM.2009.230Search in Google Scholar PubMed

52. Giovanella L, Suriano S, Ceriani L, Verburg FA. Undetectable thyroglobulin in patients with differentiated thyroid carcinoma and residual radioiodine uptake on a postablation whole-body scan. Clin Nucl Med 2012;36:109–12.10.1097/RLU.0b013e318203bb84Search in Google Scholar PubMed

53. Giovanella L. False-negative thyroglobulin measurement in recurrent/metastatic thyroid carcinomas. Eur J Nucl Med Mol Imaging 2009;36:326–7.10.1007/s00259-008-1005-9Search in Google Scholar PubMed

54. Verburg FA, Lips CJ, Lentjes EG, de Klerk JM. Detection of circulating Tg-mRNA in the follow-up of papillary and follicular thyroid cancer: how useful is it? Br J Cancer 2004;91:200–4.10.1038/sj.bjc.6601991Search in Google Scholar PubMed PubMed Central

55. Kushnir MM, Rockwood AL, Roberts WL, Abraham D, Hoofnagle AN, Meikle AW. Measurement of thyroglobulin by liquid chromatography-tandem mass spectrometry in serum and plasma in the presence of antithyroglobulin autoantibodies. Clin Chem 2013;59:982–90.10.1373/clinchem.2012.195594Search in Google Scholar PubMed PubMed Central

56. Verburg FA, Stokkel MP, Duren C, Verkooijen RB, Mader U, van Isselt JW, et al. No survival difference after successful (131)I ablation between patients with initially low-risk and high-risk differentiated thyroid cancer. Eur J Nucl Med Mol Imaging 2010;37:276–83.10.1007/s00259-009-1315-6Search in Google Scholar PubMed

57. Pacini F, Molinaro E, Lippi F, Castagna MG, Agate L, Ceccarelli C, et al. Prediction of disease status by recombinant human TSH stimulated serum Tg in the postsurgical follow-up of differentiated thyroid carcinoma. J Clin Endocrinol Metab 2001;86:5686–90.10.1210/jcem.86.12.8065Search in Google Scholar PubMed

58. Souza do Rosario PW, Ribeiro Borges MA, Fagundes TA, Franco AC, Purisch S. Is stimulation of thyroglobulin (Tg) useful in low-risk patients with thyroid carcinoma and undetectable Tg on thyroxin and negative neck ultrasound? Clin Endocrinol 2005;62:121–5.10.1111/j.1365-2265.2005.02212.xSearch in Google Scholar PubMed

59. Persoon AC, Jager PL, Sluiter WJ, Plukker JT, Wolffenbuttel BH, Links TP. A sensitive Tg assay or rhTSH stimulated Tg: what’s the best in the long-term follow-up of patients with differentiated thyroid carcinoma? PLoS One 2007;2:e816.10.1371/journal.pone.0000816Search in Google Scholar PubMed PubMed Central

60. Giovanella L, Ceriani L. High-sensitivity human thyroglobulin (hTG) immunoradiometric assay in the follow-up of patients with differentiated thyroid cancer. Clin Chem Lab Med 2002;40:480–4.10.1515/CCLM.2002.083Search in Google Scholar PubMed

61. Giovanella L, Ceriani L, Ghelfo A, Keller F, Sacchi A, Maffioli M, et al. Thyroglobulin assay during thyroxine treatment in low-risk differentiated thyroid cancer management: comparison with recombinant human thyrotropin-stimulated assay and imaging procedures. Clin Chem Lab Med 2006;44:648–52.10.1515/CCLM.2006.107Search in Google Scholar PubMed

62. Heemstra KA, Lyu YY, Stokkel M, Kievit J, Corssmit E, Pereira AM, et al. Serum thyroglobulin concentrations predict disease-free remission and death in differentiated thyroid carcinoma. Clin Endocrinol 2007;66:58–64.Search in Google Scholar

63. Chindris AM, Diehl NN, Crook JE, Fatourechi V, Smallridge RC. Undetectable sensitive serum thyroglobulin (<0.1 ng/ml) in 163 patients with follicular cell-derived thyroid cancer: results of rhTSH stimulation and neck ultrasonography and long-term biochemical and clinical follow-up. J Clin Endocrinol Metab 2012;97:2714–23.10.1210/jc.2011-3017Search in Google Scholar PubMed

64. Smallridge RC, Meek SE, Morgan MA, Gates GS, Fox TP, Grebe S, et al. Monitoring thyroglobuin in a sensitive immunoassay has comparable sensitivity to recombinant human TSH-stimulated thyroglobulin in follow-up of thyroid cancer patients. J Clin Endocrinol Metab 2007;92:82–7.10.1210/jc.2006-0993Search in Google Scholar PubMed

65. Spencer C, Lopresti J, Fatemi S. How sensitive (second-generation) thyroglobulin measurement is changing paradigms for monitoring patients with differentiated thyroid cancer, in the absence or presence of thyroglobulin autoantibodies. Curr Opinion Endocrinol Diabetes Obes 2014;21:394–404.10.1097/MED.0000000000000092Search in Google Scholar PubMed PubMed Central

66. Miyauchi A, Kudo T, Miya A, Kobayashi K, Ito Y, Takamura Y, et al. Prognostic impact of serum thyroglobulin doubling-time under thyrotropin suppression in patients with papillary thyroid carcinoma who underwent total thyroidectomy. Thyroid 2011;21:707–16.10.1089/thy.2010.0355Search in Google Scholar PubMed

67. Pacini F, Sabra MM, Tuttle RM. Clinical relevance of thyroglobulin doubling time in the management of patients with differentiated thyroid cancer. Thyroid 2011;21:691–2.10.1089/thy.2011.2107.ed1Search in Google Scholar PubMed

68. Wong H, Wong KP, Yau T, Tang V, Leung R, Chiu J, et al. Is there a role for unstimulated thyroglobulin velocity in predicting recurrence in papillary thyroid carcinoma patients with detectable thyroglobulin after radioiodine ablation? Ann Surg Oncol 2012;19:3479–85.10.1245/s10434-012-2391-6Search in Google Scholar PubMed PubMed Central

69. Giovanella L, Trimboli P, Verburg FA, Treglia G, Piccardo A, Foppiani L, et al. Thyroglobulin levels and thyroglobulin doubling time independently predict a positive 18F-FDG PET/CT scan in patients with biochemical recurrence of differentiated thyroid carcinoma. Eur J Nucl Med Mol Imaging 2013;40:874–80.10.1007/s00259-013-2370-6Search in Google Scholar PubMed

70. Boi F, Baghino G, Atzeni F, Lai ML, Faa G, Mariotti S. The diagnostic value for differentiated thyroid carcinoma metastases of thyroglobulin (Tg) measurement in washout fluid from fine-needle aspiration biopsy of neck lymph nodes is maintained in the presence of circulating anti-Tg antibodies. J Clin Endocrinol Metab 2006;91:1364–9.10.1210/jc.2005-1705Search in Google Scholar PubMed

71. Sigstad E, Heilo A, Paus E, Holgersen K, Grøholt KK, Jørgensen LH. The usefulness of detecting thyroglobulin in fine-needle aspirates from patients with neck lesions using a sensitive thyroglobulin assay. Diagn Cytopathol 2007;35:761–7.10.1002/dc.20726Search in Google Scholar PubMed

72. Suh YJ, Son EJ, Moon HJ, Kim EK, Han KH, Kwak JY. Utility of thyroglobulin measurements in fine-needle aspirates of space occupying lesions in the thyroid bed after thyroid cancer operations. Thyroid 2013;23:280–8.10.1089/thy.2011.0303Search in Google Scholar PubMed

73. Snozek CL, Chambers EP, Reading CC, Sebo TJ, Sistrunk JW, Singh RJ, et al. Serum thyroglobulin, high-resolution ultrasound and lymph node thyroglobulin in diagnosis of differentiated thyroid carcinoma nodal metastases. J Clin Endocrinol Metab 2007;92:4278–81.10.1210/jc.2007-1075Search in Google Scholar PubMed

Received: 2014-8-14
Accepted: 2014-9-22
Published Online: 2014-10-30
Published in Print: 2015-8-1