Skip to content
Publicly Available Published by De Gruyter (O) November 16, 2022

Concept for the testing of automated functions in therapeutic medical devices

Konzept für die Prüfung von autonomen Funktionen von therapeutischen Medizingeräten

  • Sandra Henn

    Sandra Henn has held a B.Sc. degree since 2018 and a M.Sc. degree since 2020 in Medical Engineering Science from Universität zu Lübeck. Since 2020 she has been employed as research associate at the Institute for Electrical Engineering in Medicine at the Universität zu Lübeck.

    EMAIL logo
    , Bastian Kabuth

    Bastian Kabuth became a biological technical assistant in 2014. He has held a B.Sc. degree in Medical Engineering Science since 2018 and a M.Sc. degree in Entrepreneurship in Digital Technologies from University of Lübeck. Since 2021 he has been employed as research associate at the IME.

    , Franziska Schollemann

    Franziska Schollemann has held a M.Sc. degree in Electrical Engineering, Information Technology and Computer Engineering from RWTH Aachen University, since 2018. She has been employed as research associate at the Institute for Electrical Engineering in Medicine at the Universität zu Lübeck since 2021.

    , Carlotta Hennigs

    Carlotta Hennigs has held a B.Sc. degree since 2018 and a M.Sc. degree since 2020 in Medical Engineering Science from Universität zu Lübeck. Since 2020 she has been employed as research associate at the Institute for Electrical Engineering in Medicine at the Universität zu Lübeck.

    , Georg Männel

    Georg Männel purses the Ph. D. degree with the Institute for electrical engineering in medicine, University of Lübeck, Germany. His research focus lies with cyber physiological and safe automated systems in medical applications. Since 2020 he has been employed researcher at the Fraunhofer Research Institution for IMTE, Lübeck, Germany.

    , Michael Angern

    Michael Angern received both his B.Sc as well as his M.Sc from the University of Lübeck for his studies in medical engineering. Since 2020 he has worked as a research associate at the Fraunhofer Research Institution IMTE in the field of Systems Engineering in Medicine.

    and Philipp Rostalski

    Philipp Rostalski is a professor for Electrical Engineering in Medicine and head of the corresponding institute at the University of Lübeck. Since 2020 he has been also director at the Fraunhofer Research Institution for Individualized and Cell-based Medical Engineering (IMTE) in Lübeck. He received his Ph.D. degree from ETH Zurich, Switzerland and served as a Feodor Lynen Scholar at the Department of Mathematics and the Department of Mechanical Engineering, University of California Berkeley, USA.

Abstract

In this paper, a testing for highly automated function (HAF) is adapted from the automotive industry to therapeutic medical devices. It contains different steps to achieve a safety argumentation: First, scenarios of interest (SoI) based on a systematic generalization of failure mode and effect analysis (FMEA) are identified, then the concrete scenarios are generated using design of experiment (DoE). These scenarios are simulated virtually and physically and are then evaluated. The procedure is explained with the use of examples.

Zusammenfassung

In diesem Artikel wird ein Testkonzept für das Testen von hochautomatisierten Funktionen (HAF) aus der Automobilindustrie für medizinische Therapiegeräte adaptiert. Es beinhaltet mehrere Schritte, um eine Sicherheitsargumentation zu erreichen. Zunächst werden auf Basis einer systematischen Verallgemeinerung der Fehler-Möglichkeits- und Einfluss-Analyse (FMEA) relevante Szenarien (SoI) identifiziert. Dann werden mit Hilfe der Versuchsplanung (DoE) die konkreten Szenarien generiert. Diese werden virtuell und physikalisch simuliert und ausgewertet. Das Vorgehen wird beispielhaft erläutert.

1 Motivation

In order to combat the shortage of qualified staff in intensive care units [1] and to increase the quality and consistency of medical care, functions with different levels of automation, e.g. decision support systems and physiological closed loop controlled (PCLC) systems, are becoming increasingly important [2]. In particular highly automated functions (HAF) have the potential to reduce the workload on medical staff, optimise therapies and allow greater focus on individual patient care [2]. At the level of automation considered here, the clinician is not part of the control loop and can even leave the room. This is comparable to the level of automation in the automotive industry where the car passenger is not part of the driving process. However, significant risks can also arise from the use of HAF (e.g. inadequate administered drugs, excessive pressure resulting in ventilator induced lung injury and many more). Due to the risk of patient harm when implementing a HAF, paired with the increased response time of clinicians, it must be ensured that these systems operate reliably and can handle all situations within the intended use or at least trigger a timely warning. This includes, most importantly, ensuring that the system can respond adequately even to rare exceptional cases or inform the clinician about hazardous situations. Lal et al. [3] found in a study that one reason for discontinuation of automated therapy was an inadequate response of the HAF to patient behaviour. Here, the device could not respond appropriately to some unusual situations. It is therefore necessary to thoroughly test and analyze HAF for rare cases and potentially hazardous situations. Therapeutic medical devices incorporating the HAF considered here likely have the highest risk class according to the European Medical Device Regulation (MDR) [4] and may be subject to clinical investigations. Naively testing all theoretical possible situations would be very expensive and, in most cases, practically impossible. Likewise, a high number of animal experiments or potentially safety critical studies on humans would not be ethically justifiable. A possible approach to achieve sufficient test coverage without testing all possible variants, is to make an adequate preselection of representative test cases.

The automotive industry faces similar challenges in the development of driver assistance systems and automated driving. For example, for the highway pilot a wrong decision of the system can as well lead to fatal consequences such as a deadly accident [5]. Maurer et al. [6] determined that it would be necessary to test a highway pilot 6.62 billion test kilometers to prove its’ safety. To address this issue in the automotive industry test-systems are developed in order to ensure safety as the key goal. An ontology-based testing approach was selected to create simulations that identify critical scenarios [5]. These critical scenarios are in turn physically evaluated on a test bench (lung simulator) and thus significantly reduce the number of required physical tests.

This article follows a cross-innovation approach by applying the established automotive test concept to a similar test challenge in medical technology. Due to the cross-innovation strategy, terms from the automotive industry [5] are used for the same aspects in medical technology. The proposed test concept starts with an ontology-based method to be able to identify and test critical and rare scenarios that could lead to patient harm. This test method, as in the automotive industry, should enable an informed decision on the acceptability of the residual risk with a reduced number of test cases but on par with almost complete test coverage.

2 Proposed test concept

The test concept presented is designed for automated functions at the system level, taking into account the patient’s physiology and the clinical context for the overall systems behavior, e.g. in physiological closed loop controlled (PCLC) systems. The challenge thereby is the selection of potentially critical scenarios to be tested, which might result in a hazardous situation to the patient, e.g. the risk of injury, and also are representative for all other scenarios not tested explicitly. That implies, that any controller of the PCLC system closing the loop has to ensure, that the modelled scenarios are handled without posing an incapable risk. Figure 1 shows the test concept for HAF based on the test method from the automotive sector [5]. This test concept includes a database for the representation of expert knowledge and possible parameters. Starting from the intended use of the HAF, SoI can be derived from the database, containing a set of therapeutic device and patient parameters. The SoIs also indicate which patients and therapeutic device models are needed to mathematically describe the physiology and function adequately. Based on the selected parameters, DoE can be used to create an experimental plan for the simulation of the interaction of the patient and the therapeutic device using the models. From the simulations, characteristic values can be extracted to evaluate the therapeutic device. On the basis of these characteristic values and patient-specific limits, it is possible to assess whether a critical case exists and could be a hazardous situation for the patient. Note, that the selection process is based on the intended use and expert knowledge and does not consider the HAF as such. The goal is to test limitations of the HAF by defining a set of critical scenarios in which the HAF has to make the right decisions for ensuring the patients safety. Therefore, the critical scenarios are physically simulated with the to-be-examined HAF on a test bench. The last step is to derive a safety argumentation for the function of the therapeutic medical device from this in compliance with the corresponding standards (e.g. DIN EN 60601-1-10 [7]). The process shown in Figure 1 is explained in more detail in the following.

Figure 1: 
Proposed test concept for HAFs of medical therapeutic devices adapted from [5].
Figure 1:

Proposed test concept for HAFs of medical therapeutic devices adapted from [5].

2.1 Database

In this ontology-based testing approach a database containing qualitative expert knowledge forms the basis for the selection of the test cases of interest. In order to manage the complexity of modelling a wide range of medical information a systematic generalization of failure mode and effect analysis (FMEA) is used which is usually applied in risk management [8]. Since complex structures and large amounts of data are common in risk management, an adaption of the system design from FMEA as a basic structure of the acquired expert knowledge base is deemed appropriate. Following the FMEA structure of cause-effects chains, this knowledge base consists of four main elements: system components, functions, failures and resulting harms (see legend in Figure 2). Those main elements are structured hierarchically. With these components and the hierarchical structure, two systems are described: the human body and the therapeutic device. The human system can be described on the one hand by literature sources and the therapeutic device on the other hand by technical documentation. The two systems are described by dividing each into hierarchical components to which individual functions are assigned. Both expert knowledge and research data can be used to describe the interaction of these systems and the interdependencies. With the help of the dependencies, cause and effect can be determined. If a cause leads to a hazardous situation (a function cannot be performed properly) for the patient, a fault is triggered in the database. Because of the hierarchical structure a cascade of cause-and-effect faults can be mapped.

Figure 2: 
Shown is the hierarchical database of components, functions and failures of both the patient and the ventilator. A harm for the patient arises from a particular combination of patient characteristics and device settings.
Figure 2:

Shown is the hierarchical database of components, functions and failures of both the patient and the ventilator. A harm for the patient arises from a particular combination of patient characteristics and device settings.

The result is a structured knowledge base, that can be easily extended with further knowledge in various directions, enabling a modular growth and input from different sources. The database can be used to extract a preselection of the relevant failures to find test SoI for the intended use. Based on the SoI and the intended use, patient and therapeutic device models with the necessary input parameters can be derived. The defined dependencies of the database are also used to calculate patient-dependent limit values for therapeutic device parameters.

2.2 Selection of concrete scenarios

In this step, the SoI, which contain the parameter ranges to be investigated, are converted into concrete test scenarios. In order to provide sufficient data for an informed risk assessment, the selection of the test scenarios is a critical challenge. A concrete scenario is thereby the selection of a specific value and its change over time for all parameters modeled in the database, e.g. oxygen saturation of the patient during ventilation. The problem of selecting the critical scenarios is manifold, among others due to the fact that the underlying distribution of the parameters, their dependencies and changes over time are generally unknown. This problem is amplified, by the challenge of gathering adequate patient data. Therefore, we propose a design of experiment (DoE) approach to select parameter sets similar to the concept of the automotive industry [5].

With methods of DoE, the smallest possible number of test cases can be found with simultaneously high test coverage through mathematical and statistical approaches. There are different methods to design matrix experiments with DoE e.g. D-Optimal Designs, Response Surface Designs or Quasi-Random Designs. The methods must always be balanced between the highest possible resolution and the lowest possible number of experiments. Therefore, there is no generally applicable method for a specific application. A suitable DoE method must always be selected iteratively. Starting with basic methods such as orthogonal fields to reduce the experimental effort is recommended. Those orthogonal fields are tables that contain parameter settings [9]. These are combined with each other in such a way that one can obtain results with an approximate statistical accuracy. The parameter settings result from the parameter ranges determined with the database. The single settings of the parameters are referred to as factors. A specific set of factors is one scenario to be evaluated. The number of tested factor combinations in relation to all possible scenarios is called a fraction. For a full fraction N parameters with F factors in the parameter space require N F scenarios. To reduce the amount of the scenarios and the associated runtime, the quantity of factors and fractions can be systematically reduced.

Furthermore, the influence of a single parameter can be determined by using interaction effects matrix plots [9]. This information can be used in further iterations to optimise the results. In addition, determined non-linear and dynamic interactions of parameters can be taken into account.

2.3 Simulation

For the simulation, the patient and therapeutic device models determined from the database and the parameter settings of the individual trials of the DoE are used. The therapeutic device model uses the therapeutic device parameters to generate the input signal for the patient model, with which the various patients are represented. The input signal corresponds to theoretically possible behaviours of the HAF. In the simulations the different trials of the DoE are then performed virtually.

The patient and therapeutic device models can be iteratively extended and adapted on the basis of the database and the intended use. Other patient characteristics such as diseases or different therapy modes can be added. Further dependencies between patient and therapeutic device can also be included in the database. This can then be taken into account by extending the models for the simulations. The simulations result in a data set that includes the response of the individual patients to the different behaviours of the therapeutic device. The behaviour of the therapeutic device corresponds to the possible behaviour of a HAF. With this method, not only the desired behaviour is depicted, but also possible over and under reactions are simulated.

2.4 Risk analysis

In the risk analysis, the simulation results are evaluated. For this purpose, values for characteristic ventilation parameters are first determined from the physiological models and simulation curves (e.g. CO2 concentration, pressure, volume). Characteristic parameters are used as a basis for examining the assumed behaviour of a potential HAF with regard to possible patient harms. The desired ranges of the characteristic parameters of an HAF are defined individually for each patient. For this purpose, the dependencies described in the database can be used. Now the characteristic parameters can be used to determine whether the theoretically possible behaviours of the HAF are within the desired range or whether limit values have been exceeded. If limit values are exceeded, the risk for the patient is determined. The risk results from the severity of harm caused by not complying with the limit values and the probability that this scenario will occur in practice. This risk evaluation can be done in accordance with ISO 14971 [10] using a risk acceptance matrix. With this a decision can be made whether the risk of a scenario is acceptable. Scenarios in which an insufficient function of an HAF must be excluded because the risk for a patient would be too high are examined in physical simulations.

2.5 Bench test

In order to ensure the correct function of the HAF, system level tests are carried out on a test bench, capable of emulating the physiological system at the interface to the device. This requires a highly flexible simulator design, in order to represent the large number of potential scenarios. Such a simulator could eventually also be used, for a replacement of animal trials in pre-clinical evaluation and research [11].

2.6 Safety argumentation

The aim of the concept shown for testing HAF is to develop a safety argumentation. For a comprehensive statement about the safety of the functions, a wide range of parameter setting, linear and non-linear behaviour and interactions of parameters need to be taken into account within the database. The more properties and dependencies are modelled and analysed, the more precise statements can be made. According to DIN 60601-1 [12] however, risks resulting from data gaps or incorrect data must be taken into account.

For the DoE step it is important that every theoretically possible behaviour of the automated function is assumed in order to be able to make an informed and reliable assessment of the safety. When DoE is applied, it must also be considered that an adequately high resolution is chosen for sufficient test coverage. Insufficient test coverage can weaken the safety argumentation.

The simulations are based on patient models. With regard to the safety conclusion, it should be mentioned here that these models are a replica and do not exactly represent the behaviour of a real patient. Credible patient models and lung simulators would have the advantage of increasing the validity of the safety argumentation [11]. The safety argumentation is highly dependent on the accuracy of the models and simulators, and may not be reliable if real-world conditions are not adequately represented. This dependency must be taken into account in the safety argumentation according to ISO 14971 [10]. If the HAF does not show any undesired behaviour in the bench tests and the risks of the single steps of the proposed test concept are acceptable in accordance to ISO 14971 [10], then a conclusion can be made that the relevant cases tested are very likely to be safe. Pre-clinical tests could thus be reduced.

3 Example implementation

For exemplification of the proposed test setup, the regulation of minute volume (MV) in mechanical ventilation as HAF is chosen. It is selected from the control hierarchy for automated mechanical ventilation as outlined by Männel et al. [13], because of its simplicity in control parameters and complexity in physiological dependencies. An extension to all aspects of the described control hierarchy is easily possible. With the proposed test concept a possible test procedure for the evaluation of this HAF is presented. For a simplified but illustrative example, we use Body Mass Index (BMI) as a parameter to represent different types of patients. The BMI is defined by height and weight. This dependencies of parameters can be described using appropriate formulas in the patient models.

3.1 Example database

As a first iteration, a small model was set up using the FMEA tool Plato e1ns (PLATO AG, Lübeck, Germany), containing all relevant system components from the respiratory system and the ventilator system, as well as their functions and failures. This model was used to investigate cause and effect relationships. The model is represented in a tree structure conceptually depicted in Figure 2. From the dependencies described in this model, it can e.g. be concluded that a person’s BMI in combination with certain ventilation parameters can be a cause of various harms. It may be further analyzed that the pathophysiological changes of the respiratory system contribute significantly to an increased risk of complications in patients with a high BMI during ventilation.

Obesity is considered a widespread risk factor for diseases and roughly 67% of men and 53% of women are overweight in Germany [14]. This results in a very high relevance for everyday clinical practice. The probability of obesity in ventilated patients is therefore high. It follows that for the SoI, the patient parameters of height and weight are relevant.

The respiratory rate (RR) and ΔP (difference between plateau pressure and positive end exspiratory pressure) can be set by the ventilator. A scenario consists of a chosen combination of values of these parameters and additional constant parameters within the mathematical model representing the patient and his therapy setting. The range of possible values of the single parameters is stored in the database.

3.2 Example selection of concrete scenarios

In order to illustrate the selection of concrete scenarios, the parameters weight, height, RR and ΔP selected with the database are used. The chosen value ranges of the variable parameters are listed in Table 1. On the patient side, value ranges were chosen that lead to a BMI for an overweight person. For the ventilator parameters, the range that can occur in practice was used. With these four parameters and for example ten factors, 410 scenarios would result. This corresponds to a full fraction with a high resolution because of the ten factors. To reduce the simulation effort an iterative approach is used to find a suitable sub-fraction and a representative number of factors [9]. For this first iteration the chosen method is the Box–Behnken design. Each row in the generated output matrix contains the settings for all listed parameters, either −1 (minimum), 0 (mean) or 1 (maximum) [15].

Table 1:

Variable parameters for the matrix experiments.

Parameter Range
Height in cm 160–200
Weight in kg 100–160
RR in breaths per minute (BPM) 10–30
ΔP in cm H2O 5–20

For four parameters the Box–Behnken design yields an experimental design that contains 27 trials. The individual trials of the experiment are centered between the worst-case combination of extreme cases of the enclosed space. These worst case combinations are omitted from the experiments, so that this does not result in extreme cases, which only occur with a negligible probability, but in boundary cases, which are more likely to be possible. An extract of the first seven trials of the generated output DoE-matrix is shown in Table 2.

Table 2:

Variable parameters (weight in kg, height in cm, respiratory rate (RR) in breaths per minute, ΔP in cm H2O) with the corresponding Box–Behnken elements in brackets.

Trial no. Weight Height RR ΔP
1 100 (−1) 160 (−1) 20 (0) 12.5 (0)
2 100 (−1) 200 (1) 20 (0) 12.5 (0)
3 160 (1) 160 (−1) 20 (0) 12.5 (0)
4 160 (1) 200 (1) 20 (0) 12.5 (0)
5 130 (0) 180 (0) 10 (−1) 5 (−1)
6 130 (0) 180 (0) 10 (−1) 20 (1)
7 130 (0) 180 (0) 30 (1) 5 (−1)

3.3 Example simulation

For the simulation of the concrete scenarios, a one-compartment model as shown in Figure 3 used as a simple example for patient modelling [16]. The model can be extended to represent other aspects of patient behaviour. The one-compartment model, displayed in Figure 3, contains a pressure source (P aw), a resistance (R) and compliance (C), which are connected to

(1) P aw = R V ̇ ( t ) + 1 C V ( t ) P mus .

Figure 3: 
RC-model with airway pressure (P
aw), respiratory muscle pressure (P
mus), respiratory resistance (R), flow 



(



V

̇


)



$(\dot{V})$



, lung compliance (C) based on [17].
Figure 3:

RC-model with airway pressure (P aw), respiratory muscle pressure (P mus), respiratory resistance (R), flow ( V ̇ ) , lung compliance (C) based on [17].

In general the airway resistance is defined as the pressure difference between the mouth (atmospheric pressure) and the alveoli of the lungs, divided by the airflow. At high body weight, abdominal fat accumulation affects lung mechanics, thus a BMI dependency exists. The airways can also be narrowed by fat deposits [18]. Therefore, the resistance of the airway (R) is increased and the compliance of the lung (C) is decreased [19]. Pelosi et al. [19] descried a simple heuristic relationship between BMI and airway resistance (R) as follows

R ( B M I ) = 2.55 e 0.03 B M I .

The airway resistance (R) expressed in cm H2O  s L−1 describes the flow resistance of the entire airway system. It depends on the dimensions of the upper airways, the pulmonary airway tree and the acini with its alveoli [16]. A high BMI, i.e. obesity increases breathing resistance. This indicates that the caliber of the airways are reduced during the entire tidal breathing cycle [20]. In addition, Pelosi et al. [19] described a simple heuristic relationship between BMI and lung compliance (C) as follows

C ( B M I ) = 233.3 e 0.086 B M I + 40 .

Pelosi et al. [19] demonstrated that lung compliance (C) expressed in mL  cm H2O−1 decreased with increased BMI, and decrease was observed even with a small increase in body mass. Using the respiratory rate (RR) expressed in min−1, the time for one breathing cycle t tot can be calculated according to

t tot = 1 R R

and is expressed in s as well as t i and t e . The inspiratory expiratory ratio (I:E ratio) is set to 1:1.5 [21]. By inserting t e = 1.5 · t i into the equation

t i = t tot t e

and solving for t i , we obtain

t i = 2 5 t tot .

The factor 2/5 is given by the I:E ratio. The airway pressure (P aw) is assumed to be a square wave and depends on the parameters PEEP, ΔP, t i and t e . The P aw curve was smoothed to be closer to physiological conditions (see Figure 4). The respiratory muscle pressure (P mus) is set to zero as a fully sedated state is considered with no muscle activity. The PEEP, which is not considered a relevant parameter for the function under consideration in this simple example, is fixed to the lowest suitable value of 5 cm H2O, cf [21]. In Figure 5 the plots of the first seven trials can be seen. The differential equations (see Eq. (1)) need to be solved to plot the volume curves (see Figure 5) of the trials of the DoE. Based on the simulation, the minima and maxima of the respective curves were determined. The tidal volume (V T ) expressed in L indicates the amount of air that enters the patient during normal inspiration [22]. To calculate the tidal volume (V T ), the respective minima is subtracted from the corresponding maxima. The minute volume

M V = V T R R

expressed in L min−1 is the product of the respiratory rate (RR) and the tidal volume (V T ) [22].

Figure 4: 
Plots of the first seven exemplary P
aw square waves for the DoE’s respective trials. The P
aw square waves are smoothed. The plots relating to trials 1 to 4 are superimposed. The plots are shown over the time section between 13 and 21 s. The plots of the experiments are differentiated by color and line structure.
Figure 4:

Plots of the first seven exemplary P aw square waves for the DoE’s respective trials. The P aw square waves are smoothed. The plots relating to trials 1 to 4 are superimposed. The plots are shown over the time section between 13 and 21 s. The plots of the experiments are differentiated by color and line structure.

Figure 5: 
Plots of the first seven exemplary trials of the DoE. The plots of trials 1 to 4 are superimposed. The plots are shown over the time section between 13 and 21 s. The plots of the experiments are differentiated by color and line structure.
Figure 5:

Plots of the first seven exemplary trials of the DoE. The plots of trials 1 to 4 are superimposed. The plots are shown over the time section between 13 and 21 s. The plots of the experiments are differentiated by color and line structure.

3.4 Example risk analysis

Threshold values calculated from the parameter settings are determined for this example. For this purpose, the ideal body weight (IBW) is calculated for each given patient based on height and gender. For the experiments, a male patient was assumed. The ideal body weight for men (IBW M ) is expressed in kg and is calculated with

I B W M = 50 + 0.91 ( h e i g h t 152.4 ) .

The height and the value 152.4 are expressed in cm. The factor 0.91 is expressed in kg cm−1 and the value 50 is expressed in kg. For the ventilated patients, a V T of 0.006–0.008 L kg−1 standard body weight is set [21]. Multiplying IBW by 0.008 L kg−1 results in

V T max = I B W M 0.008

and multiplying IBW by 0.006 L kg−1 one obtains

V T min = I B W M 0.006 .

In general the tidal volume V T is expressed in L.

To determine the maximum minute volume

M V max = V T max R R

one uses V Tmax and for the minimum minute volume

M V min = V T min R R

one uses V Tmin. The individually calculated parameters MV and V T are compared with the respective determined threshold values. The result for V T is display in Figure 6. The evaluation of MV is carried out analogously. Based on these results, a risk assessment is carried out. The risk results from the combination of the severity of the harm and the probability of occurrence (see Figure 7). The severity of the harm is assessed by how much it deviates from the desired area. The probability of occurrence of harms is composed of the probability of occurrence of obesity in combination with unfavourable ventilation parameters. In this example the intended use of the HAF is the ventilation of patients with a high BMI. The occurrence of obesity is therefore always present. The time of use of the HAF during therapy can be used to evaluate the probability of occurrence of faulty ventilation parameters. Since the function is used permanently, there is a continuous possibility of selecting incorrect parameters for ventilation.

Figure 6: 
The calculated tidal volumes for all 27 experiments performed. For visualization, the different V

T
 are normalized to the patient-specific limits (V

Tmin, V

Tmax).
Figure 6:

The calculated tidal volumes for all 27 experiments performed. For visualization, the different V T are normalized to the patient-specific limits (V Tmin, V Tmax).

Figure 7: 
Shown is the risk that arises for the simulated patients depending on the probability and severity of harm.
Figure 7:

Shown is the risk that arises for the simulated patients depending on the probability and severity of harm.

If a concrete scenario (see Figure 7) results in an unacceptable risk, the scenario is critical and must be evaluated in a physical simulation, e.g. on a test bench. The manufacturer addresses the identified risks as part of the risk management process. This implies, that depending on the type of risk, the intended use must be restricted or a redesign is necessary.

3.5 Example bench test

With the bench tests the critical scenarios are simulated and it is tested whether the HAF can handle those situations. This includes also the option of alarming. These tests can be performed on lung simulators which generate the simulated physical conditions on the interface of the actual therapy device running the HAF. A lung simulator is being developed for this bench test. The concept of the lung simulator by Bautsch et al. [23] includes a system that can represent a wide range of scenarios. Other commercially available lung simulators maybe used as well depending on their functionality and the HAF under test.

3.6 Example safety argumentation

Under the consideration of two conditions the safety of the HAF may be assessed and potential weak spots and faults be addressed. The first aspect investigates that the HAF can handle the critical scenarios in the physical simulation. The other aspect corresponds to compliance with ISO 14971 [10] throughout the entire testing process. This includes taking into account the risks that arise from the individual test steps as described in Section 2.6.

For this example a conclusion about which behaviour of the automated function could lead to damage can only be made for the defined range of overweight, e.g. of a BMI of 25–50 kg m−2, as defined as the SoI within the database. This would therefore also correspond to the scope of application for this example.

The chosen Box–Behnken design operates with maxima, minima and mean values and therefore gives a good overview of the borders of the parameter space. However, with the Box–Behnken design only the borders and the centre can be examined. Intermediate states or local critical states cannot be investigated with this method. To close this data gap for compliance with DIN EN 60601-1 [12] further iterations of experimental designs are necessary, in which other parameters and different types of dependencies of these parameters can also be taken into account. For example, an increased BMI tends to have a positive effect in combination with chronic obstructive pulmonary disease (COPD).

The one-compartment model used in this example implementation corresponds to an approximation of patient behaviour. The risks that this simplification entails must be taken into account. This also applies to test benches that are only an approximate representation of human physiology.

4 Discussion

This article presented a concept for testing HAF and provided a simplified example for implementing the different steps of the test concept. Therapeutic medical devices incorporating HAF will likely have the highest risk class according to the European Medical Device Regulation (MDR) [4]. According to the current status, a pre-clinical and a clinical test phase might thus be necessary to test these medical devices. As already described, with automated functions it is hardly possible to check all possible scenarios with these tests and in particular complicated and safety-critical test cases are difficult to realize. This would always result in a very incomplete picture of the safety of the automated function. The systematic approach of the proposed test concept offers the advantage that it can be ensured that only the relevant cases are tested in the preclinical tests. This increases the significance and reduces the testing effort on preclinical trials.

5 Conclusion and future work

A concept for testing automated functions in medical technology was derived from a previously proposed test concept in the automotive industry and illustrated on a simplified medical use case. The individual steps were explained conceptually using a selected example. It could be shown that with the test structure an assessment of a HAF is a promising alternative also for the medical domain. These tests can be used to define the intended use of the device. Also risks for patients within the intended use can be identified and can lead to a potential redesign of the device. Several possibilities for extending the individual test steps were presented. The database needs of course to be supplemented with further risk factors in ventilation and diseases as well as interactions with e.g. medication. In the case of a large database, supporting tools are needed for evaluation in order to identify risks.

To avoid relevant test gaps, further methods of DoE will be used in the future. Here, the trade-off between the lowest possible number of simulations and a high test coverage is a key challenge. For simulation, patient models need to be refined and extended to replicate patient characteristics more precisely and comprehensively.

As a key step, the risk analysis needs to be extended to check further characteristics in order to be able to make a more precise assessment.


Corresponding author: Sandra Henn, Institute for Electrical Engineering in Medicine, Universität zu Lübeck, Lübeck, Germany, E-mail:

About the authors

Sandra Henn

Sandra Henn has held a B.Sc. degree since 2018 and a M.Sc. degree since 2020 in Medical Engineering Science from Universität zu Lübeck. Since 2020 she has been employed as research associate at the Institute for Electrical Engineering in Medicine at the Universität zu Lübeck.

Bastian Kabuth

Bastian Kabuth became a biological technical assistant in 2014. He has held a B.Sc. degree in Medical Engineering Science since 2018 and a M.Sc. degree in Entrepreneurship in Digital Technologies from University of Lübeck. Since 2021 he has been employed as research associate at the IME.

Franziska Schollemann

Franziska Schollemann has held a M.Sc. degree in Electrical Engineering, Information Technology and Computer Engineering from RWTH Aachen University, since 2018. She has been employed as research associate at the Institute for Electrical Engineering in Medicine at the Universität zu Lübeck since 2021.

Carlotta Hennigs

Carlotta Hennigs has held a B.Sc. degree since 2018 and a M.Sc. degree since 2020 in Medical Engineering Science from Universität zu Lübeck. Since 2020 she has been employed as research associate at the Institute for Electrical Engineering in Medicine at the Universität zu Lübeck.

Georg Männel

Georg Männel purses the Ph. D. degree with the Institute for electrical engineering in medicine, University of Lübeck, Germany. His research focus lies with cyber physiological and safe automated systems in medical applications. Since 2020 he has been employed researcher at the Fraunhofer Research Institution for IMTE, Lübeck, Germany.

Michael Angern

Michael Angern received both his B.Sc as well as his M.Sc from the University of Lübeck for his studies in medical engineering. Since 2020 he has worked as a research associate at the Fraunhofer Research Institution IMTE in the field of Systems Engineering in Medicine.

Philipp Rostalski

Philipp Rostalski is a professor for Electrical Engineering in Medicine and head of the corresponding institute at the University of Lübeck. Since 2020 he has been also director at the Fraunhofer Research Institution for Individualized and Cell-based Medical Engineering (IMTE) in Lübeck. He received his Ph.D. degree from ETH Zurich, Switzerland and served as a Feodor Lynen Scholar at the Department of Mathematics and the Department of Mechanical Engineering, University of California Berkeley, USA.

  1. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: S. Henn and F. Schollemann were supported by the “Cross-Innovation-Center–TANDEM Phase III (TANDEM III-CIC)” LPW-E/1.1.1/1521. G. Männel and M. Angern were supported by the European Union–European Regional Development Fund (ERDF), the Federal Government and Land Schleswig Holstein, Project No. 12420002. B. Kabuth was supported by “Medical Cause and Effects Analysis” within the AI funding policy of Schleswig Holstein (“KI-Förderrichtlinie”, Project No. 220 21 011). We thank Charlott Danielson (Fraunhofer Research Institution for Individualized and Cell-Based Medical Engineering) for her critical review of regarding regulatory aspects of this project.

  3. Conflict of interest statement: The authors declare no conflicts of interest regarding this article.

References

[1] H. Lüttmann, Fachkräftemangel in der Intensivpflege, GRIN Verlag, 2017 [Online], Available at: https://www.grin.com/document/378874.Search in Google Scholar

[2] M. J. Khodaei, N. Candelino, A. Mehrvarz, and N. Jalili, “Physiological closed-loop control (PCLC) systems: review of a modern frontier in automation,” IEEE Access, vol. 8, p. 1, 2020. https://doi.org/10.1109/ACCESS.2020.2968440.Search in Google Scholar

[3] R. A. Lal, M. Basina, D. M. Maahs, K. Hood, B. Buckingham, and D. M. Wilson, “One year clinical experience of the first commercial hybrid closed-loop system,” Diabetes Care, vol. 42, no. 12, pp. 2190–2196, 2019. https://doi.org/10.2337/dc19-0855.Search in Google Scholar PubMed PubMed Central

[4] Europäisches Parlament und Rat der europäischen Union, “Verordnung (EU) 2017/745 über Medizinprodukte, zur Änderung der Richtlinie 2001/83/EG, der Verordnung (EG) Nr. 178/2002 und der Verordnung (EG) Nr. 1223/2009 und zur Aufhebung der Richtlinien 90/385/EWG und 93/42/EWG des Rates,” Amtsblatt Der Europäischen Union L, vols. 117/1, 2017.Search in Google Scholar

[5] E. Böde, M. Bücker, W. Damm, et al.., Identifikation und Quantifizierung von Automationsrisiken für hochautomatisierte Fahrfunktionen, Oldenburg, OFFIS e.V., 2019.Search in Google Scholar

[6] M. Maurer, J. C. Gerdes, B. Lenz, and H. Winner, Autonomous Driving, Heidelberg, Springer Berlin Heidelberg, 2016.10.1007/978-3-662-48847-8Search in Google Scholar

[7] DIN EN 60601-1-10:2021-11 – Medizinische elektrische Geräte – Teil 1-10: Allgemeine Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale – Ergänzungsnorm: Anforderungen an die Entwicklung von physiologischen geschlossenen Regelkreisen (IEC 60601-1-10:2007 + A1:2013 + A2:2020), Berlin, Beuth Verlag, 2021.Search in Google Scholar

[8] P. A. García Aguirre, L. Perez-Dominugez, D. Luviano-Cruz, R. Romero-López, and E. Leon-Castro, “PFDA-FMEA-VSM application for outsourcing risk analysis,” Preprints, 2022. https://doi.org/10.20944/preprints202201.0002.v1.Search in Google Scholar

[9] A. Richter, Versuchsplanung – Design of Experiments, Heidelberg, Springer Vieweg Berlin, 2014, pp. 12–15.Search in Google Scholar

[10] ISO 14971: 2019 – Medical Devices Application of Risk Management to Medical Devices, Berlin, Beuth Verlag, 2019.Search in Google Scholar

[11] Food and Drug Administration, “Draft guidance for industry and food and drug administration staff,” in Assessing the Credibility of Computational Modeling and Simulation in Medical Device Submissions, Rockville, Publisher, 2021.Search in Google Scholar

[12] DIN EN 60601-1: 2013-12 – Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance, Berlin, Beuth Verlag, 2013.Search in Google Scholar

[13] G. Männel, C. Hoffmann, and P. Rostalski, “A robust model predictive control approach to intelligent respiratory support,” in 2018 IEEE Conference on Control Technology and Applications (CCTA), IEEE, 2018, pp. 12–17.10.1109/CCTA.2018.8511363Search in Google Scholar

[14] G. Mensink, Übergewicht und Adipositas in Deutschland: Werden wir immer dicker? Berlin, Robert Koch-Institut, 2012.Search in Google Scholar

[15] MATLAB, Box–Behnken Design – MATLAB BBDesign – MathWorks, Deutschland, [Online], Available at: https://de.mathworks.com/help/stats/bbdesign.html (visited on 01/13/2022).Search in Google Scholar

[16] J. H. T. Bates, Lung Mechanics – An Inverse Modeling Approach, Cambridge, Cambridge University Press, 2009.10.1017/CBO9780511627156Search in Google Scholar

[17] A. Ben-Tal, “Simplified models for gas exchange in the human lungs,” J. Theor. Biol., vol. 238, no. 2, pp. 474–495, 2006. https://doi.org/10.1016/j.jtbi.2005.06.005.Search in Google Scholar PubMed

[18] J. G. Elliot, G. M. Donovan, K. C. Wang, F. H. Green, A. L. James, and P. B. Noble, “Fatty airways: implications for obstructive disease,” Eur. Respir. J., vol. 54, no. 6, pp. 1–10, 2019. https://doi.org/10.1183/13993003.00857-2019.Search in Google Scholar PubMed

[19] P. Pelosi, M. Croci, I. Ravagnan, et al.., “The effects of body Mass on lung volumes, respiratory mechanics, and gas exchange during general anesthesia,” Anesthesia & Analgesia, vol. 87, no. 3, pp. 654–660, 1998.10.1213/00000539-199809000-00031Search in Google Scholar

[20] C. M. Salome, G. G. King, and N. Berend, “Physiology of obesity and effects on lung function,” J. Appl. Physiol., vol. 108, no. 1, pp. 206–211, 2010. https://doi.org/10.1152/japplphysiol.00694.2009.Search in Google Scholar PubMed

[21] DGAI, “Webinar zur S3-Leitlinie Invasive Beatmung und Einsatz extrakorporaler Verfahren bei akuter respiratorischer Insuffizienz,” Der Chirurg., vol. 92, no. 9, pp. 851–852, 2021. https://doi.org/10.1007/s00104-021-01458-x.Search in Google Scholar PubMed

[22] H. Lang, Beatmung für Einsteiger, Berlin, Heidelberg, Springer Berlin Heidelberg, 2020 [Online], Available at: http://link.springer.com/10.1007/978-3-662-59294-6.10.1007/978-3-662-59294-6Search in Google Scholar

[23] F. Bautsch, G. Männel, and P. Rostalski, “Development of a novel low-cost lung function simulator,” in Current Directions in Biomedical Engineering, vol. 5, Berlin, Walter de Gruyter GmbH, 2019.10.1515/cdbme-2019-0140Search in Google Scholar

Received: 2022-01-31
Accepted: 2022-10-04
Published Online: 2022-11-16
Published in Print: 2022-11-25

© 2022 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 11.5.2024 from https://www.degruyter.com/document/doi/10.1515/auto-2022-0010/html
Scroll to top button