Comparison between Manual Methods and Automated Analyzer iQ200® Iris Diagnostics): A Study for the Optimization of Urinalysis

José Ricardo Henneberg1, Railson Henneberg1, Aguinaldo José do Nascimento1, Gislene Kussen2, Fellype Carvalho Barreto3 and Aline Borsato Hauser1*

1Clinical Analysis Department, Universidade Federal do Paraná, Brazil
2Pharmaceutical Biochemistry of the Clinical Hospital, Universidade Federal do Paraná, Brazil
3Center for Health and Biological Sciences, Pontifícia Universidade Católica do Paraná, Brazil
Prof. Aline Borsato Hauser, Clinical Analysis Department, Sector of Health Sciences, Universidade Federal do Paraná, Rua Lothario Meissner, 632 Curitiba, PR 80210-170, Brazil,Tel.+55 41 33604084; E-mail: alinehauser@ufpr.br
30 July 2014; 30 October 2015; 01 November 2015
Henneberg JR, Henneberg R, Nascimento AJ, Kussen G, Barreto FC (2015) Comparison between Manual Methods and Automated Analyzer iQ200® Iris Diagnostics): A Study for the Optimization of Urinalysis. Int J Lab Med Res 1: 108. doi: http://dx.doi.org/10.15344/2455-4006/2015/108

Abstract

Aim: Urinalysis is an important laboratory exam that gives helpful information for the diagnosis of urinary tract disease.
Objective: The purpose of this study was to compare automated method results of test strips, microscopic analysis and counts with the manual methods.
Material and Methods: Urine samples (n=275) were analyzed by I-Chem® Velocity (iQ200®). White and red blood cells (WBC/RBC) counts by iQ200® were compared with manual methods (Neubauer and KCell). Leukocyte esterase and hemoglobin detection provided by the test strips were also analyzed.
Results: Leukocyte esterase was positive in 59.6% with 315,800±608,413 WBC/ml and negative in 40.4% with 24,838±34,152 WBC/ml (p<0.001). Hemoglobin was detected in 29.8% of the samples with 1,275,537±7,394,959 RBC/ml and negative in 70.2% with 26,316±46,424 RBC/ml (p=0.019). The iQ200® detected RBC in 159 samples, being 97.5% isomorphic cells and 2.5% unclassified iQ200® detected casts in 33 samples, classifying 36.4% as no inclusions and 63.6% with inclusions. Comparison of WBC/ RBC counts between Neubauer and K-cell® were equivalent (R2=0.87 and 0.75). Comparison of WBC/ RBC counts between iQ200 and manual methods showed higher values to the automation. Conclusion: Automated methods, such as iQ200® , seems to be a reliable tool for urinalysis in daily clinical practice, reducing the execution time and providing more accuracy. Distinction between isomorphic and dysmorphic RBC, casts and crystals identification still depend on microscopic examination. The use of automated device in urinalysis is a good practice, but it requires careful evaluation, taking into account the population, the volume of the analysis and the availability of skilled professional.


Summary

The purpose of this study was to compare automated method results of test strips, microscopic analysis and counts with the manual methods, using urine samples (n=275) analyzed by iQ200®. White and red blood cells (WBC/RBC) counts by iQ200® were compared with manual methods (Neubauer and KCell). Leukocyte esterase and hemoglobin detection provided by the test strips were also analyzed. We suggest that automated methods, such as iQ200®, seems to be a reliable tool for urinalysis in daily clinical practice, but the distinction between isomorphic and dysmorphic RBC, casts and crystals identification still depend on microscopic examination.

1. Introduction

Renal diseases are an important public health problem worldwide. Early recognition and management of these diseases are considered good strategies to reduce the progression to end-stage renal disease (ESRD) and, consequently, the number of chronic kidney disease (CKD) patients undergoing renal replacement therapy [1-5].

The Kidney Early Evaluation Program (KEEP®) recommends the use of urinalysis in daily clinical settings and underlines the importance of detecting hematuria and proteinuria as primary signs of kidney disease. Furthermore, the evaluation of urinary erythrocyte morphology has been considered crucial to differentiate a glomerular from a non-glomerular disease [6,7].

The symptoms caused by renal or lower urogenital tract pathologies have many similarities, which reinforce urinalysis is a valuable tool for the initial evaluation of the genitourinary system. In addition, despite of probably being one of the most ancient diagnostic tests used in medicine, urine examination continues to be an attractive option to obtain information on corporal homeostasis by means of a quick and easy collecting method. Urinalysis by manual method, using standard volume of urine and several types of chamber to count urinary elements, has been performed in many clinical. More recently, the burden of urinary specimens begs for automation and recent technological advances have turned the automated urine testing a reliable tool that has reduced the execution time for the analysis[8-10].

Whereas both the manual and automation methods have advantages and disadvantages, the present study sought to compare the results obtained by using automated urinalysis (iQ200®) and manual methods (Neubauer Chamber and K-Cell® Blades) in order to analyze counts as well as the ability to detect and identify urinary elements for the purpose of aid in the diagnosis of renal diseases.

2. Material and Methods

Two hundred seventy-five urine specimens obtained from patients (n= 275; age: 46.5±18.7 years; 215 (78.2%) females) at the Diagnosis Support Unit of the Hospital de Clinicas, Universidade Federal do Paraná (UAD-HC / UFPR) were analyzed in the period from April to June 2014. The samples were analyzed by equipment I-Chem® Velocity (IQ200®), manufactured by IRIS Diagnostics (Division of International Remote Imaging Systems, Inc. USA). IQ200® consists of two automated units, being one unit to the physical and biochemical parameters evaluation (test strips) and the other for the detection and counting of urinary elements.

Urine samples were analyzed within 2 hours after collection. After homogenization and aspiration of an aliquot of 1 ml, samples were placed between two sheets forming a fluid layer, which is under a microscope coupled to a video camera charged coupling device (CCD) that captured 500 frames per sample. The images were sent to the analytical unit of the computer. The particles were isolated and analyzed by Auto Particle Recognition (APR®), neural network software which classifies the particles prepared using size, texture, contrast and shape into 12 distinct categories: white blood cells (WBC), red blood cells (RBC), WBC aggregates, hyaline casts (without inclusions), casts not classified (with inclusions), epithelial cell sloughing, non-squamous epithelial cells, bacteria, fungi, mucus, sperm and crystals. The operator can edit the results by reviewing images. The quantification of particles was calculated from the number found in the volume analyzed [11,12]. The urinary elements were analyzed and confirmed in the screen of iQ200® by experienced professionals. Then, when the screen shows elements called "unclassified", the final classification was performed from the evaluation of the observed images. And, when necessary it was checked by optical microscopic analysis.

Selected samples were analyzed by the manual microscopic analysis using a Neubauer Chamber and K-Cell® Blade without centrifugation, as recommended by the European Urinalysis Guidelines (EUG) for studies comparing manual and automated methodologies in urinalysis [13].

The Neubauer Chamber has delimited areas used for cell count. The depth between the bottom surface of the cover glass that sits on elevated rails and the enclosed area is 0.1mm. The delimited area is a square of side 3 mm, divided into 9 large squares with each side of 1 mm. The four quadrants of WBC were used and the value obtained was multiplied by 2,500 to obtain the value of cell per ml. Counts were performed in duplicate and the average of the counts of each sample was used for statistical analysis. K-Cell® Blade has 10 wells and each well has two sets of nine circles. The volume of each circle corresponds to 0.011uL and the set of 9 circles corresponds to the total volume of 0.1ul. The result of cell count was multiplied by 10,000 to obtain the value of cell per ml. Counts were performed in duplicate and the average of the counts of each sample was used for statistical analysis, Figure 1 [14,15].

figure 1
Figure 1: A) Counting of WBC per ml in urine samples according to leukocyte esterase test, B) Counting of RBC per ml in urine samples according to the detection of hemoglobin by the test strip. Note: In legend A) p means positive and n means negative (p<0.001); B) p means positive and n means negative (p=0.002).

This study was submitted and approved by the Ethics Committee and Research Health Sector UFPR under number 15230113.5.0000.0102 (CAAE) and by the co-participant Institution (Hospital de Clinicas/ UFPR) issued the number 361 880.

Statistical analysis was performed using Statistica10® (Statsoft) as analysis of variance (ANOVA) randomized model, Z test for two proportions to mutually exclusive categories and analysis of regression and correlation "Pearson".

3. Results

The WBC count per ml by iQ200® device was compared with the presence of the leukocyte esterase test strip. Positive esterase test was observed in 59.6% (n=164) and the results obtained by IQ200® in these samples presented 315,800±608,413 WBC per ml. Negative esterase was observed in 40.4% (n=111) and presented 24,838±34,152 WBC per ml (p<0.001), Figure 2.

figure 2
Figure 2: A) Frequency of RBC morphology in urine samples after microscopic analyzer, B) Dysmorphic RBC in urine sample, microscopy optical, fresh exam (increase 400x).

Regarding the presence of hemoglobin in the test strip, 82 samples showed positive reaction (29.8%) and 193 showed negative reaction (70.2%). When comparing the presence of hemoglobin in the test strip in relation to the RBC count by IQ200®, the negative samples presented 26,316±46,424 RBC per ml and positive samples presented 1,275,537±7,394,959 RBC per ml (p=0.019), Figure 3.

figure 3
Figure 3: Frequency of different types of casts after microscopic analysis.

Regarding RBC morphology, the automated method detected 159 samples with RBC, being 155 (97.5%) with isomorphic morphological pattern and 4 (2.5%) considered "unclassified". The images provided by IQ200® were not sharp enough to reclassify. Thus, "unclassified samples" were analyzed by optical microscopy. RBC from these samples was dysmorphic, Figure 4.

figure 4
Figure 4: Comparisons of RBC counts by three different methods.

The iQ200® equipment identified the presence of 33 samples with urinary casts which were classified as hyaline or no inclusions 36.4% (n=12) or other casts with inclusions 63.6% (n=21). The casts were reclassified based on the images provided by iQ200®: hyaline casts were present in 21 samples (64%), granular casts in 6 (18%) and hyaline and granular casts in 6 (18%). A similar result was obtained by optical microscopy analyses: hyaline casts were observed in 20 samples (59%), granular casts in 07 (21%) and hyaline and granular casts in 06 (18%). Importantly, hyaline and RBC casts (3%) was detected in only one sample by optical microscopy and not by the iQ200®, Figure 5.

figure 5
Figure 5: Comparisons of RBC counts by three different methods.

Table 1 show the WBC and RBC counts by three methods used in the study. Despite the significant difference in maximum scores, the confidence interval (CI) was the same in both manual methods except by equipment which many more cells are counted.

table 1
Table 1: WBC and RBC counts by iQ 200®, K-cell® Blade and Neubauer Chamber (n=275).

The comparison of WBC count between Neubauer Chamber and K-cell® Blade presented a random error of 13.11% and a systematic error of 0.16% (R2=0.87). The comparison of WBC count between KCell® and iQ200® showed random error of 51.89% and a systematic error of 9.3% (R2=0.48). This comparison showed that the values found in iQ200® were on average 1.6 times higher and the correlation between the two methodologies decreases when the values were above 1500/μl. Comparison of WBC count between Neubauer and IQ200® showed random error of 39.47% and a systematic error of 6.1% (R2=0.60). The iQ200® counts were on average 1.6 times higher and the correlation between the two methodologies decreases when the values were above 1000/μl.

The comparison of RBC count between the two methodologies Neubauer and K-Cell® demonstrated that both are equivalent, presenting a casual error of 25% and systematic error of 11% (R2=0.7548). The comparison of the RBC count between K-Cell®and iQ200® showed a random error of 21% and a systematic error of 2% (R2=0.7897). This comparison showed values on average 1.16 times higher in iQ200®. The comparison of the RBC count between Neubauer and iQ200® presented a random error of 43% and a systematic error of 18% (R2=0.5736). This comparison showed values on average 2.75 times higher in iQ200®.

4. Discussion

This study compared three different methods of urine analysis in an attempt to verify the advantages and disadvantages of each method. Regarding automated test strip analysis, it was compared the results in the leukocyte esterase test. Comparing the positive results and counts by equipment, it was possible to demonstrate the low frequency of false-positive results in the test strip. The main causes of falsepositive results include oxidizing agents, reaction with the nuclei of epithelial cells and the use of drugs such as imipenem, meropenem and clavulanate [16,17].

Comparing the negative results of leukocyte esterase and the WBC count by equipment, the negative samples had a mean of 24,838 WBC per mL iQ200®. This result could show low sensitivity to values considered significant (> 10,000/ml). The sensitivity of the test described by the equipment manufacturer provides information of a detection threshold for the presence of the esterase from counts above 25,000 WBC per ml [11]. Therefore, the equipment detects within the proposed linearity, but above the levels considered significant for WBC. The test is based in the presence of the enzyme esterase in the urine, which occurs after cell lyses not only due to infectious processes, but also due to low urinary density or alkaline pH. According to Skitek (2004), a false negative result may occur in poorly preserved samples or when analyzed after 2 hours of collection [18]. The main factors that may lead to a false negative results are: high levels of glucose and proteins, yellow pigment, crenation due to high density of WBC that prevents the release of esterases or drugs and chemicals such as ascorbic acid, oxalic acid, cephalexin, cephalothin, gentamicin, and tetracycline [19,20].

Some authors have demonstrated considerable heterogeneity in the accuracy of test strip [9,10,21,22]. A negative test strip is not necessarily indicative of WBC counts below the reference value, as confirmed by our data. As so, the analysis of WBC should be performed by microscopic or by automated methods even in samples that tested negative for the presence of leukocyte esterase. Automated systems have been developed to screen urines in order to eliminate those in which a culture is highly unlikely to yield clinically relevant results. These systems have recently added the detection of a specific parameter “small particles”, which has been added to improve the sensitive of the assay. However, the addition of this parameter did not add any benefit to results obtained by traditional measures, especially WBCs, in identifying urines for which a culture should be performed [10,23]. Then, considering also that test strip has low sensitivity in samples with few WBCs count (<25,000/ml), we do not recommend the isolated use of just one of the methods commented as a screening for urine culture.

Regarding the detection of hemoglobin by the test strip, the difference found can be explained by the methodology used by manufacturers of test strips, which is based on the pseudoperoxidase activity of the heme radical present in hemoglobin [24]. If there is no free hemoglobin molecule, the chemical reaction may be negative, even when RBCs are present in the urine sample. Whereas RBC counts above 5,000 cells per ml are clinically significant, our results showed that it is not possible to rely only on the test strip. However, the isolated use of microscopy shows the limitation of not detecting “ghosts” (when only the RBC membrane is faintly visible due to the hemolysis) arising from samples with free hemoglobin produced by hemolytic disorders or by lysis of RBC in the urinary tract. Our study highlights the importance of concomitant use of chemical and microscopic analysis to evaluate the presence of blood in the urine. The chemical test strip determines the presence of free hemoglobin and the microscopic method confirms the presence of the RBC that do not reacted in the test strip since they are intact and allows the identification of dysmorphic RBC.

In devices using flow cytometer, the RBC can be confused with yeast, since this method uses laser beams that identify the cell as the size or shape of the generated pulse [25]. Altekin and colleagues (2010) reported impaired RBC quantification due to the influence of particles as yeast, oxalate crystals or bacteria [26]. In our study, there was no false-positive result for RBC when yeasts were present, suggesting the correct identification of these elements by the imaging system.

Our study seeks to contribute to assess the use of automation in urinalysis as a diagnostic aid in glomerular diseases. Currently, urinalysis flow cytometer supplies through impedance histograms with data on the volume of RBC, allowing prior information of likely morphological changes arising from the passage of RBC by glomerular capillaries [27,28,29]. The iQ200® usually recognizes and classifies the isomorphic RBC, but does not classify dysmorphic RBC, only relates to “unclassified” [30,12]. Boven et al. (2012) compared the use of iQ200®with phase contrast microscopy (n=207) in patients with suspected dysmorphism. The iQ200® showed enough images for classification of D1 and D2 cells, but was not sufficient to classify D3 cell forms (D1/D2 are acanthocyts with lesser or greater loss of cytoplasmic staining and D3 represent all other forms of cell deformity), emphasizing the importance of using the microscope for a more detailed study [31].

In our study, it was detected 159 samples with RBC, 155 with isomorphic morphological pattern and 4 considered "unclassified". The presence of dysmorphic RBC was confirmed in the screen, but the distinction between the dysmorphic forms was not possible. The samples "unclassified" were confirmed by optical microscopy according to the technique described by Barros Silva et al. [32]. This finding reaffirmed the importance of reviewing the urine sample by optical microscopy when the "unclassified" message is released.

While the iQ200® recognizes casts through images, flow cytometer system regards the laser deviation as the particle size, each output signal is converted by the equipment software and plotted on a graph as small particles (RBC, WBC and others) and larger particles (casts and epithelial cells). But also does not differentiate between different casts [25,7]. In our study, the results for automated and microscopic analyses were similar. The presence of hyaline casts detected by iQ200® was confirmed by microscopy. Even though the equipment image evaluation had been provided, it was not possible to identify the morphological characteristics and the microscopic evaluation was necessary whenever “unclassified” casts were detected. Another important situation was the finding of RBC casts by microscopy not reported by iQ200® images. This occurred only in one sample, a fact perhaps explained by heterogeneity of urine samples that constitute a true colloidal solution, so it may be that the volume aspirated by iQ200 not presents such casts. This can also occur in the manual microscopic analysis by centrifugation errors, as reported by Ito et al. [33].

Considering that urinary casts are important findings in renal diseases, the correct identification of RBC, WBC, waxy or fatty casts, are important to assist in the diagnosis of glomerulonephritis, pyelonephritis, nephrotic syndrome or even to indicate the severity of a kidney disease [34-38]. So, the distinction between the righteous isomorphic and dysmorphic RBC, casts and crystals identification still remain dependent on microscopic examination, as well as the presence of rare or unusual elements such as Trichomonas vaginalis, Pthirus pubis, Enterobius vermicularis eggs and others. In this context, training and professional experience are crucial for the quality of the analysis.

The comparison between WBC/RBC counts by Neubauer Chamber and Kcell® Blade showed similar results between the two methods. The comparison between manual analysis and iQ200®, showed higher counts by the equipment, a fact that occurs due to the increased number of cells counted by automated systems. Other studies also reported higher results in automated systems such as iQ200® [39,30].

Although the amount of WBC in urine is not directly related to the intensity of infectious processes, values above 10,000 per ml are considered significant and when massive WBC per ml is found, it may be related to pyelonephritis and WBC casts are common findings. Furthermore, the correct quantification of RBC in urine is important because is directly proportional to the intensity lesion in pathologies that affect the urinary system, especially to check the progression of lesions [40].

Although the laboratory results should only be compared if used the same method as manual (Neubauer, KCell or others) or as automated (optical or flow cytometers), in our study the cell counts showed an acceptable variation in the methods used that does not exclude the reliability of the result. The iQ200® proved to be a useful tool in urinalysis providing results quickly and securely, to improve the flow of samples in the routine and not depend on previous sample preparation. However, the identification of certain urinary elements requires microscopic review. KCell® Blade seems to be a good alternative to manual urinalysis, as demonstrated by the good correlation with Neubauer chamber, a higher velocity analysis and provides a quantitative result per ml. It has been recognized that the manual counts are less accurate because of low number of cells is counted and of beeing operator dependent. Otherwise, it has been reported automated system inability to correctly identify all the elements present in urine sediment such as dismorphic RBC, yeast, oval crystals of calcium oxalate, degenerated cells and mucus cells grouped [41,26,39].

Urinalysis is an important tool in screening and detecting urinary elements found in the urogenital system diseases that must be performed judiciously and with standard techniques. We suggest that the implementation of automated devices, such as iQ200®, in urinalysis should be carefully evaluated taking into account the population to be served, the volume of the analysis performed, the availability of skilled professionals and the relation cost/benefit, without losing focus on the quality of the laboratory result.

Competing Interests

The authors declare that they have no competing interests.

Author Contributions

Nicola Daniele, Fulvia Fraticelli and Silvia Franceschilli, contributed in draſting the manuscript. Francesco Zinno, contributed in draſting and reviewing the manuscript.


References

  1. Dirks JH, Robinson SW (2005) The global perspective of the International Society of Nephrology: a decade of experience with COMGAN. Kidney Int 68: 1395-1410. View
  2. Dirks JH, de Zeeuw D, Agarwal SK, Atkins RC, Correa-Rotter R, et al. (2005) Prevention of chronic kidney and vascular disease: toward global health equity--the Bellagio 2004 Declaration. Kidney Int Suppl 98: S1-S6. View
  3. Foley RN, Parfrey PS, Sarnak MJ (1998) Clinical epidemiology of cardiovascular disease in chronic renal disease. Am J Kidney Dis 32: S112-119. View
  4. Foley RN, Parfrey PS, Sarnak MJ (1998) Epidemiology of cardiovascular disease in chronic renal disease. J Am Soc Nephrol 9: S16-23. View
  5. Hauser AB, Stinghen AE, Kato S, Bucharles S, Aita C, et al. (2008) Characteristics and causes of immune dysfunction related to uremia and dialysis. Perit Dial Int 28 Suppl 3: S183-187. View
  6. Whaley-Connell AT, Vassalotti JA, Collins AJ, Chen SC, McCullough PA (2012) National Kidney Foundation's Kidney Early Evaluation Program (KEEP) annual data report 2011: executive summary. Am J Kidney Dis 59: S1-S4. View
  7. Block DR, Lieske JC (2012) Automated urinalysis in the clinical lab. MLO Med Lab Obs 44: 8-10, 12. View
  8. Hauss O (2008) Bringing Urinalysis into 21st Century: From Uroscopy to Automated Flow Citometry. Sysmex J Int 18: 38-43.
  9. Okada H, Horie S, Inoue J, Kawashima Y (2007) The Basic Performance of Bacteria Couting for Diagnosis or Urinary Tract Infection use the Fully Automated Urine Particle Analyser UF-1000i. Sysmex J Int 17: 95-111.
  10. Parta M, Hudson BY, Le TP, Ittmann M, Musher DM, et al. (2013) IRIS iQ200 workstation as a screen for performing urine culture. Diagn Microbiol Infect Dis 75: 5-8. View
  11. Diagnostics I. IQ200® Operators Manual. Equipment I-Chem® Velocity (IQ200®). USA, 2004. Epub Division of International Remote Imaging Systems, Inc. USA.
  12. Linko S, Kouri TT, Toivonen E, Ranta PH, Chapoulaud E, et al. (2006) Analytical performance of the Iris iQ200 automated urine microscopy analyzer. Clinica chimica acta; international journal of clinical chemistry 372: 54-64. View
  13. European Confederation of Laboratory Medicine (2000) European Urinalysis Guidelines. Scand J Clin Lab Invest Suppl 231: 1-86. View
  14. Bastidas O. Nota Técnica 2014. View
  15. Vacutest Vacutest Kima Urine 2012.
  16. Kouri T, Fogazzi G, Gant V, Hallander H, Hofmann W, et al. (2000) European urinalysis guidelines. Scand J Clin Lab Invest 60: 1-96.
  17. Laudicina R, Fenn JP, Freeman V, McCoy C, McLane MA, et al. (2011) Research in clinical laboratory science: professionals' involvement. Clin Lab Sci 24: 235-242. View
  18. Skitek M (2004) Imprecision and instability of common analytes in urine analysed by semi- automated dipstick urinalysis. Accred Qual Assur 9: 700- 703. View
  19. Strasinger SK, Di Lorenzo MS (2009) Urinalysis and Body Fluids (6th edn.), F a Davis Co, Philadelphia, USA. View
  20. Mundt LA, Shanahan K (2011) Graff´s textbook of routine urinalysis and body fluids (2nd ed.) Lippincott Williams & Wilkins, Philadelphia, USA. View
  21. Karakukcu C, Kayman T, Ozturk A, Torun YA (2012) Analytic performance of bacteriuria and leukocyturia obtained by UriSed in culture positive urinary tract infections. Clin Lab 58: 107-111. View
  22. Okada H, Sakai Y, Miyazaki S, Arakawa S, Hamaguchi Y, et al. (2000) Detection of significant bacteriuria by automated urinalysis using flow cytometry. J Clin Microbiol 38: 2870-2872. View
  23. St John A, Boyd JC, Lowes AJ, Price CP (2006) The use of urinary dipstick tests to exclude urinary tract infection: a systematic review of the literature. Am J Clin Pathol 126: 428-436. View
  24. Fogazzi GB, Verdesca S, Garigali G (2008) Urinalysis: core curriculum 2008. Am J Kidney Dis 51: 1052-1067. View
  25. Gai M, Piccoli GB, Segoloni GP, Lanfranco G (2003) Microscopic urinalysis and automated flow cytometry in a nephrology laboratory. Clin Chem 49: 1559-1560. View
  26. Altekin E, Kadiçesme O, Akan P, Kume T, Vupa O, et al. (2010) New generation IQ-200 automated urine microscopy analyzer compared with KOVA cell chamber. J Clin Lab Anal 24: 67-71. View
  27. Fogazzi GB, Cantù M, Garigali G (2005) Anisocytes and poikilocytes in the urine. Nephrol Dial Transplant 20: 840-841. View
  28. Kubota H, Yamabe H, Ozawa K, Fukushi K, Ohsawa H, et al. (1988) Mechanism of urinary erythrocyte deformity in patients with glomerular disease. Nephron 48: 338-339. View
  29. Köhler H, Wandel E, Brunck B (1991) Acanthocyturia--a characteristic marker for glomerular bleeding. Kidney Int 40: 115-120. View
  30. Lamchiagdhase P, Preechaborisutkul K, Lomsomboon P, Srisuchart P, Tantiniti P, et al. (2005) Urine sediment examination: a comparison between the manual method and the iQ200 automated urine microscopy analyzer. Clinica chimica acta; international journal of clinical chemistry 358:167-174. View
  31. Boven LA, Kemperman H, Demir A (2012) A comparative analysis of the Iris iQ200 with manual microscopy as a diagnostic tool for dysmorphic erythrocytes in urine. Clin Chem Lab Med 50: 751-753. View
  32. Barros Silva GE, Costa RS, Ravinal RC, Saraiva e Silva J, Dantas M, et al. (2010) Evaluation of erythrocyte dysmorphism by light microscopy with lowering of the condenser lens: A simple and efficient method. Nephrology (Carlton) 15: 171-177. View
  33. Ito CA, Pecoits-Filho R, Bail L, Wosiack MA, Afinovicz D, et al. (2011) Comparative analysis of two methodologies for the identification of urinary red blood cell casts. J Bras Nefrol 33: 402-407. View
  34. Spinelli D, Consonni D, Garigali G, Fogazzi GB (2013) Waxy casts in the urinary sediment of patients with different types of glomerular diseases: results of a prospective study. Clinica chimica acta 424: 47-52. View
  35. Fogazzi GB, Leong SO (1996) The erythrocyte cast. Nephrol Dial Transplant 11: 1649-1652. View
  36. Fogazzi GB, Cameron JS (1996) Urinary microscopy from the seventeenth century to the present day. Kidney Int 50: 1058-1068. View
  37. Egido J, Rojas-Rivera J (2011) Protocolo diagnóstico de leucocituria y cilindruria. Medicine - Programa de Formación Médica Continuada Acreditado 10: 5590-5591. View
  38. Martin-Moreno PL (2011) Interpretación del urianálisis. Medicine - Programa de Formación Médica Continuada acreditado 10: 5383-5385.
  39. Wah DT, Wises PK, Butch AW (2005) Analytic performance of the iQ200 automated urine microscopy analyzer and comparison with manual counts using Fuchs-Rosenthal cell chambers. Am J Clin Pathol 123: 290-296. View
  40. Leonardo de Souza V, Maria Goretti Moreira Guimarães P, Pedro Guatimosim V (2005) The importance of the dysmorphic erythrocyte for investigation of the source of hematuria: literature review. Jornal Brasileiro de Patologia e Medicina Laboratorial 41: 83-94. View
  41. Cappelletti Pea (2013) Is Automated Intelligent Microscopy (AIM) an answer to the medical needs of the urinalysis? iQ200 Iris in urine sediment testing. Diagnostic and workflow improvements 9: 61-78.