The Accuracy and Clinical Relevance of the Multi-echo Dixon Technique for Evaluating Changes to Hepatic Steatosis in Patients with Non-alcoholic Fatty Liver Disease Treated with Formulated Food
Ying YangKuanghui XuXiaofei ChenJianping DingJunping ShiJie Li
Author information
JOURNAL OPEN ACCESS

2023 Volume 22 Issue 2 Pages 263-271

Details
Abstract

Purpose: The Multi-echo Dixon (ME-Dixon) is a non-invasive quantitative MRI technique to diagnose non-alcoholic fatty liver disease (NAFLD). In this study, the hydrogen proton MR spectroscopy (1H-MRS) was used as a reference to explore the accuracy of the ME-Dixon technique in evaluating hepatic steatosis in NAFLD patients after ingesting formulated food and its correlation with changes in clinical indicators.

Methods: Twenty-seven patients with NAFLD were enrolled. Fifteen patients completed 12 weeks of treatment with prebiotics and dietary fiber. In addition, abdominal MRI scans and blood tests were performed before and after treatment. The MRI-proton density fat fraction (MRI-PDFF) and MRS-PDFF were measured using the ME-Dixon and 1H-MRS techniques. The Bland–Altman method and Pearson correlation analysis were used to test the consistency of the two techniques for measuring the liver fat content and the changed values. Besides, correlation analysis was conducted between the MRI-PDFF value and metabolic indicators.

Results: In the PDFF quantification of 42 person-times and the monitoring of the PDFF change in 15 patients under treatment, there was a good consistency and a correlation between MRI and MRS. At baseline, MRI-PDFF was positively correlated with insulin resistance index (HOMA-IR), fatty liver index (FLI), and liver enzymes. After treatment, the changes in MRI-PDFF were positively correlated with the recovery degree of FLI and liver enzymes.

Conclusion: ME-Dixon has a good consistency and a correlation with MRS in quantifying the liver fat content and monitoring the treatment effect, which may be used as an accurate indicator for clinical monitoring of changes in the liver fat content.