Open Access

Investigations of Thermocouple Drift Irregularity Impact on Error of their Inhomogeneity Correction

 and    | Mar 06, 2014

Cite

The article examines: (i) the reasons of error due to thermoelectric inhomogeneity of electrodes of thermocouples acquired during prolonged use; (ii) the neural network method of error correction based on a generalization of verification results in several temperature fields; (iii) the method of investigating the impact of changing the speed of the conversion characteristic drift of thermocouple on error correction; (iv) results of this investigation. It is shown that residual error for type K thermocouples at the 5 % level of significance does not exceed μ±0.46 oС and one at the 10 % level of significance does not exceed ±0.25 °С

eISSN:
1335-8871
Language:
English
Publication timeframe:
6 times per year
Journal Subjects:
Engineering, Electrical Engineering, Control Engineering, Metrology and Testing