External Evaluation of the Quality of the Four Emergency Biochemical Parameters in Antananarivo

Article Information

Miora Koloina Ranaivosoa1*, Zafindrasoa Domoina Rakotovao-Ravahatra2, Rakotomalala Toky Randriamahazo3, Rasoamialy-Soa Razanakolona4, Olivat Rakoto Alson5, Andry Rasamindrakotroka6

1Laboratory of Biochemistry, Joseph Ravoahangy Andrianavalona University Hospital Antananarivo, Madagascar

2Laboratory of Biochemistry, Joseph Raseta Befelatanana University Hospital Antananarivo, Madagascar

3Laboratory of Immunology, Joseph Ravoahangy Andrianavalona University Hospital Antananarivo, Madagascar

4Department of Parasitology, HIV/AIDS Reference Laboratory, Madagascar

5Department of Biological Haematology, Medical Biology Department of the Faculty of Medicine Antananarivo, Madagascar

6Department of Immunology, Laboratory of Training and Research in Medical Biology, University of Antananarivo, Madagascar

*Corresponding Author: Miora Koloina Ranaivosoa, Laboratory of Biochemistry of Joseph Ravoahangy Andrianavalona University Hospital Antananarivo, Madagascar

Received: 25 January 2020; Accepted: 13 March 2020; Published: 20 March 2020

Citation:

Miora Koloina Ranaivosoa, Zafindrasoa Domoina Rakotovao-Ravahatra, Rakotomalala Toky Randriamahazo, Rasoamialy-Soa Razanakolona, Olivat Rakoto Alson, Andry Rasamindrakotroka. External Evaluation of the Quality of the Four Emergency Biochemical Parameters in Antananarivo. Journal of Biotechnology and Biomedicine 2 (2020): 029-038.

View / Download Pdf Share at Facebook

Abstract

The External Evaluation of the Quality is an important procedure for assessing the performance of a laboratory. The aim of the present study is to assess the performance of the the biochemistry laboratory of Joseph Ravoahangy Andrianavalona by the dosage of the most prescribed 4 emergency biochemical parameters. It is a retrospective evaluative study of the performance of the biochemistry laboratory of Joseph Ravoahangy Andrianavalona University Hospital Center on a period of 4 months from January 2019 to April 2019 involving 8 samples of External Evaluation of the Quality provided by the Quality Insurance Association of the clinical Biology Laboratories in France. Serum creatinine, blood glucose, uremia and serum electrolytes were dosed on the 8 samples which carried out 32 dosages. During a period of 4 months the accuracy was 94% for the 32 carried out dosages. The precision had been 97% if the Variation Coefficient Ratio (VCR) was below 1.50. The monthly assessment of the 3 studied parameters: serum creatinine, blood glucose and uremia were always good during the 4 months. The imprecision (VCR >1.50) concerned the disease control of the sodium levels in February. The University Hospital Center Joseph Ravoahangy Andrianavalona Biochemistry laboratory was efficient during the External Evaluation of the Quality. Therefore, these results were reliable and contribute to the improvement of the diagnosis, the monitoring and the aftercare of the patients. Every laboratory for medical analysis ought to organize a periodical External Evaluation of the Quality to monitor and improve biological analysis results.

Keywords

Accreditation, Quality control, Performance

Accreditation articles Accreditation Research articles Accreditation review articles Accreditation PubMed articles Accreditation PubMed Central articles Accreditation 2023 articles Accreditation 2024 articles Accreditation Scopus articles Accreditation impact factor journals Accreditation Scopus journals Accreditation PubMed journals Accreditation medical journals Accreditation free journals Accreditation best journals Accreditation top journals Accreditation free medical journals Accreditation famous journals Accreditation Google Scholar indexed journals Quality control articles Quality control Research articles Quality control review articles Quality control PubMed articles Quality control PubMed Central articles Quality control 2023 articles Quality control 2024 articles Quality control Scopus articles Quality control impact factor journals Quality control Scopus journals Quality control PubMed journals Quality control medical journals Quality control free journals Quality control best journals Quality control top journals Quality control free medical journals Quality control famous journals Quality control Google Scholar indexed journals Performance articles Performance Research articles Performance review articles Performance PubMed articles Performance PubMed Central articles Performance 2023 articles Performance 2024 articles Performance Scopus articles Performance impact factor journals Performance Scopus journals Performance PubMed journals Performance medical journals Performance free journals Performance best journals Performance top journals Performance free medical journals Performance famous journals Performance Google Scholar indexed journals Biochemical Parameters articles Biochemical Parameters Research articles Biochemical Parameters review articles Biochemical Parameters PubMed articles Biochemical Parameters PubMed Central articles Biochemical Parameters 2023 articles Biochemical Parameters 2024 articles Biochemical Parameters Scopus articles Biochemical Parameters impact factor journals Biochemical Parameters Scopus journals Biochemical Parameters PubMed journals Biochemical Parameters medical journals Biochemical Parameters free journals Biochemical Parameters best journals Biochemical Parameters top journals Biochemical Parameters free medical journals Biochemical Parameters famous journals Biochemical Parameters Google Scholar indexed journals Antananarivo articles Antananarivo Research articles Antananarivo review articles Antananarivo PubMed articles Antananarivo PubMed Central articles Antananarivo 2023 articles Antananarivo 2024 articles Antananarivo Scopus articles Antananarivo impact factor journals Antananarivo Scopus journals Antananarivo PubMed journals Antananarivo medical journals Antananarivo free journals Antananarivo best journals Antananarivo top journals Antananarivo free medical journals Antananarivo famous journals Antananarivo Google Scholar indexed journals Biochemistry articles Biochemistry Research articles Biochemistry review articles Biochemistry PubMed articles Biochemistry PubMed Central articles Biochemistry 2023 articles Biochemistry 2024 articles Biochemistry Scopus articles Biochemistry impact factor journals Biochemistry Scopus journals Biochemistry PubMed journals Biochemistry medical journals Biochemistry free journals Biochemistry best journals Biochemistry top journals Biochemistry free medical journals Biochemistry famous journals Biochemistry Google Scholar indexed journals blood glucose articles blood glucose Research articles blood glucose review articles blood glucose PubMed articles blood glucose PubMed Central articles blood glucose 2023 articles blood glucose 2024 articles blood glucose Scopus articles blood glucose impact factor journals blood glucose Scopus journals blood glucose PubMed journals blood glucose medical journals blood glucose free journals blood glucose best journals blood glucose top journals blood glucose free medical journals blood glucose famous journals blood glucose Google Scholar indexed journals emergency articles emergency Research articles emergency review articles emergency PubMed articles emergency PubMed Central articles emergency 2023 articles emergency 2024 articles emergency Scopus articles emergency impact factor journals emergency Scopus journals emergency PubMed journals emergency medical journals emergency free journals emergency best journals emergency top journals emergency free medical journals emergency famous journals emergency Google Scholar indexed journals Variation Coefficient articles Variation Coefficient Research articles Variation Coefficient review articles Variation Coefficient PubMed articles Variation Coefficient PubMed Central articles Variation Coefficient 2023 articles Variation Coefficient 2024 articles Variation Coefficient Scopus articles Variation Coefficient impact factor journals Variation Coefficient Scopus journals Variation Coefficient PubMed journals Variation Coefficient medical journals Variation Coefficient free journals Variation Coefficient best journals Variation Coefficient top journals Variation Coefficient free medical journals Variation Coefficient famous journals Variation Coefficient Google Scholar indexed journals evaluation articles evaluation Research articles evaluation review articles evaluation PubMed articles evaluation PubMed Central articles evaluation 2023 articles evaluation 2024 articles evaluation Scopus articles evaluation impact factor journals evaluation Scopus journals evaluation PubMed journals evaluation medical journals evaluation free journals evaluation best journals evaluation top journals evaluation free medical journals evaluation famous journals evaluation Google Scholar indexed journals

Article Details

1. Introduction

Medical analysis laboratories had a very important part to play in the disease diagnosis, monitoring and management. Biological analysis results should be reliable for patients well-being. Therefore, inter-laboratories comparison was necessary in order to control the results and the eventual errors which might happen. The External Evaluation of the Quality (EEQ) was a procedure for perfomances assessment of a laboratory through an inter-laboratories comparison realised by an organiser who substantially respected the requirement of ISO 43-11 (cf. § 5, 6, 4) and the current regulation with a sample from unknown controls. The realisation of an EEQ was a part of one of the six preliminary works to establish the aims of analytic performance [1, 2]. The biochemistry laboratory of Joseph Ravoahangy Andrianavalona University Hospital Center (CHUJRA) was yearly registered at the EEQ program of the Quality Insurance Association of France Clinical Laboratories (ASQUALAB). Therefore, the aim of this study was to assess the performance of the CHUJRA biochemistry laboratory through the dosage of four emergency biochemical parameters which were the most prescribed (serum creatinine, uremia, blood glucose and serum electrolytes) on the automaton Mindray BS 300®.

2. Materials and Methods

It was a retrospective evaluative study of CHUJRA biochemistry laboratory performance during a period of four months, from January 2019 to April 2019 through the 8 samples from EEQ BIOLABO®, with unknown values, at two levels: a normal level and a pathologic level, provided by the association ASQUALAB in France. These 8 samples were tested on biochemical automaton Mindray BS 300® by dosing the 4 most prescribed emergency biochemical parameters (serum creatinine, blood glucose, uremia and serum electrolytes). The EEQ BIOLABO® samples were multi-parameter lyophilized blood serum at two levels presented as special boxes intended for the EEQ of clinical biochemical parameters provided by the association ASQUALAB in France. The results of all EEQ samples were included in this study and consisted of the 4 routinely most prescribed parameters.

The EEQ BIOLAB® samples which were used had different number for each month:

  • In January 2019 : BS 1901 and BS 1902
  • In February 2019 : BS 1903 and BS 1904
  • In March 2019 : BS 1905 and BS 1906
  • In April 2019 : BS 1907 and 1908.

Before each step, EEQ BIOLABO samples were reconstituted by 5ml distilled water as it was recommended by the supplier. The dosage was done once only for each level and for each parameter. The results were collected on results sheets. Then they were sent on ASQUALAB website to be analysed. ASQUALAB sent the EEQ report one month after, once the results were analysed. The results were reported in a tabular summary form and a bar graph for each control level.

Concerning the data analysis method, the parameters for the evaluation of the precision and the accuracy were measured. The parameters for the evaluation of the method dispersion (precision) consisted of the average, the Variation Coefficient (VC) that is to say the dispersion measurement of calculated results, the Standard Error (SE) and the Variation Coefficient Ratio (VCR) [3]. The study of these parameters allowed us to identify random errors. VCR was the ratio between the laboratory VC and the peer group VC. Concerning its interpretation, VCR<1 means the laboratory precision is better than that of the peer group; VCR>1 means imprecision; VCR>1.5 needs the search of the imprecision causes; VCR>2 means maintenance and compulsory corrective action [3].

The parameters for the evaluation of the measurement accuracy were formed by calculating the bias and the Z-score or SEI (Standard Error Index) [4]. Z-score gave the deviation between the measured value and the target value. These latter allowed the search for possible factual errors. Regarding the interpretation, Z-score S means satisfactory accuracy; Z-score D means questionable accuracy requiring attention; Z-score I means unsatisfactory accuracy requiring action [4]. Each of the participants results were assessed compared to the general average of the results provided by the other participating laboratories or to the general average from those who used the same device (peer group).If the results showed anomalies in relation to the defined acceptability criteria, the participants were alerted [5]. The results should be first of all interpreted with data comparison for all the techniques [4] then with the peer group the laboratories of which were ranked according to the used method, depending on the methods principles, on instruments and devices, on reagents and calibration [6]. These criteria were mentioned on the results report for the EEQ results interpretation. The study was realised after the validation of the plant manager and the Merieux Foundation.

3. Results

For the results of the EEQ, the accuracy of the analysis results was 94% with an inaccuracy of kaliemia in March 2019. This evaluation of the accuracy of each parameter assessed by the Z-score is represented in Tables 1, 2, 3, 4 for the months of January, February, March and April. The last column represents the evaluation of the biochemical analysis. For each analysis, the comment takes into account the results obtained for each of the two samples evaluated in relation to the general average and to the bias. The assessment was good if the calculated bias is within the range of the acceptability limits specific to each analysis. The systematic error (SE) is displayed if the results of the 2 samples are affected by an error of the same sign. This comment is accompanied by a "+" sign if the results are above the high acceptability limit or by a "-" sign if the results concerned are below the low acceptability limit. Concerning the precision, it was 97% with imprecise natremia level in February 2019 according to Table 5. Table 5 represents the summaries of the variation coefficient ratio (VCR) of the 6 parameters analyzed compared to the peer group. Random error (RE) has been observed if one of the 2 results deviates from the acceptability limits or both, but the biases are not of the same sign.

fortune-biomass-feedstock

Table 1: External Evaluation of the Quality (EEQ) results for samples BIO1901-BIO1902 in January 2019.

fortune-biomass-feedstock

Table 2: External Evaluation of the Quality (EEQ)  results for samples BIO1903-BIO1904 in February 2019.

fortune-biomass-feedstock

Table 3: External Evaluation of the Quality (EEQ) results for samples BIO1905-BIO1906 in March 2019.

fortune-biomass-feedstock

Table 4: External Evaluation of the Quality (EEQ) results for samples BIO1907-BIO1908 in April 2019.

fortune-biomass-feedstock

Table 5: Summary of Variation Coefficient Ratio (VCR) for the 6 parameters analyzed in relation to peer groups.

4. Discussion

The realisation rhythm of EEQ program was a periodical one. It is admitted that from twice to 4 times yearly participation corresponded to a reasonable rhythm [7, 8]. The CHUJRA chemistry laboratory was able to realise once a month the passage of EEQ solutions according to the recommendations of the ASQUALAB assessment program and others too [6, 9]. In this present study, 4 parameters were chosen; they were mostly prescribed in emergency. Nevertheless in our daily practice, 29 routine biochemical parameters were monthly assessed thanks to ASQUALAB such as transaminases and serum calcium levels. Several EEQ organisations used the Z-scores in the participants return reports. But the analysis of VCR was also important [10]. Concerning this present study results, the accuracy was appreciated by Z-score. It was 94% in this study. An inaccuracy was reported for kaliemia in March 2019. It was a systematic error or a bias one. It could be related to the deterioration of automaton lamp, the progressive accumulation of remnants into the tubing or on the electrodes, the reagents aging, the progressive deterioration of the control components, or the optical fibre integrity. Similarly, the lamp failure, its sudden variation or the change of reagents package could be the responsible of this error [11].

After the accuracy, the precision was estimated by the VCR. This precision was 97% in the present study. However an imprecision was reported for natremia in February 2019. According to the literature, imprecision could be linked to an incorrect operation of the analytical process concerning the reagents, the equipment, the staff, the calibration procedures and the internal control [12]. VCR analysis was not very informative and not very sensitive. It was a second line security indicator [6]. The EEQ results encouraged to establish approaches for curative and / or preventive actions [5, 9]. But late reception (1 month after the input of results) of EEQ report results did not allow to develop a preventive action or an immediate correction but rather a healing action. Therefore it was recommended to keep a volume increment of each EEQ for moving again samples at the report return in order to make verification [10].

The present study allowed us to check up the reliability of the biochemical analysis results in CHUJRA. We found that the analysis performance was good in the majority of the cases. Therefore biologists and technicians in this laboratory were assured that the validated analysis results corresponded to the patients state of health and these latter will be properly supported by clinicians. Similarly, diseases were well diagnosed and their monitoring was improved. The present study highlighted the importance of EEQ which should be performed by all medical analysis laboratories. It was true that ASQUALAB EEQ cost rather expensive (300 € per year). Nevertheless, when lacking adequate amount of financial resources, medical analysis laboratories could perform inter-laboratories comparisons of their biological tests and could correct one another. The main thing was to be able to check if the daily performed biological tests were in standards. In short, as healthcare worker, our aim was always to look for ways to improve patient care. So, the patient life expectancy would be extended.

5. Conclusion

EEQ was an assessment procedure of a laboratory performance. It allowed the laboratory to know if the results of biological tests were reliable and also allowed to perform corrective actions in order to improve the reliability of these results. CHUJRA biochemical laboratory was efficient during the External Evaluation of Quality. So, these results of biochemical tests were reliable and contribute to the diagnosis improvement, to the monitoring and to the aftercare of the patients. Every laboratory of medical analysis ought to organise a periodical External Evaluation of Quality results in order to monitor and to improve the results of the biological tests.

Acknowledgements

We would like to thank all the staff of the laboratory of University Hospital of Ravoahangy Andrianavalona and all the laboratory technicians. Similarly, we would like to express our gratitude to the director of establishment for authorizing us to carry out this study.

Conflict of Interest

None

References

  1. Technical accreditation guide for the evaluation of computer systems in medical biology. SH GTA 02. Revision 00. COFRAC (2013).
  2. Klee GG. Establishment of outcome-related analytic performance goals. Clin Chem 56 (2010): 714-722.
  3. NF EN ISO 15189: 2007-Laboratoire d’analyses de biologie médicale-Exigences particulières concernant la qualité et la compétence. Paris: AFNOR (2007).
  4. Giannoli JM, Szymanowicz A. Propositions de recommandations pour l’utilisation pratique des contrôles internes de qualité (CIQ) dans un laboratoire de biologie médicale. Ann Biol Clin 64 (2011): 489-498.
  5. Yadav R, Bhartiya JP, Verma SK, et al. External quality assessment scheme (EQAS): Our experience as a participating laborator. IJRRMS 3 (2013).
  6. Berge Kristensen GB, Meijer P. Interpretation of EQA results and EQA-based trouble shooting. Biochem Medica 27 (2017): 2749-2762.
  7. Technical accreditation guide: quality control in medical biology. SH GTA 06. Revision 00. COFRAC.
  8. Vassault A, Annaix V, Houlbert C, et al. Recommendations for mastering the pre-analytical and analytical phases of delocalized medical biology exams. Ann Biol Clin 70 (2012): 233-238.
  9. Geethanjalai FS, Fleming J, Swaminathan S, et al. External quality assurance-role of ACBI/CMC scheme. Indian J Clin Biochem 21(2006): 211-212.
  10. Szymanowicz A. Optimized interpretation and relevant use of external quality assessment (EQA). Feuillets de Biologie VOL LI N° 294 - MAI (2010).
  11. Technical guide for verification accreditation (scope a)/validation (scope b) of methods in medical biology. SH GTA 04. Révision 00-Avril. COFRAC (2011).
  12. Dumontet M. Implementation, use and exploitation of quality control in order to ensure analytical validation, metrological control of analysis instruments and determination of measurement uncertainty. Spectra Biologie 26 (2007): 35-37.

Journal Statistics

Impact Factor: * 5.3

CiteScore: 2.9

Acceptance Rate: 11.01%

Time to first decision: 10.4 days

Time from article received to acceptance: 2-3 weeks

Discover More: Recent Articles

Grant Support Articles

© 2016-2024, Copyrights Fortune Journals. All Rights Reserved!