Article

A Rapid RT-LAMP Assay for SARS-CoV-2 with Colorimetric Detection Assisted by a Mobile Application

by
María Aurora Londono-Avendano
1,*,
Gerardo Libreros
1,
Lyda Osorio
2 and
Beatriz Parra
1
1
Departamento de Microbiología, Escuela de Ciencias Básicas, Facultad de Salud, Universidad del Valle, Calle 4B # 36-00, edificio 120, oficina 223/229, Cali 760043, Colombia
2
Escuela de Salud Pública, Facultad de Salud, Universidad del Valle, Calle 4B # 36-00, edificio 120, oficina 223/229, Cali 760043, Colombia
*
Author to whom correspondence should be addressed.
Diagnostics 2022, 12(4), 848; https://doi.org/10.3390/diagnostics12040848
Submission received: 20 December 2021 / Revised: 12 January 2022 / Accepted: 21 January 2022 / Published: 29 March 2022
(This article belongs to the Section Point-of-Care Diagnostics and Devices)

Abstract

:
Loop-mediated amplification has been promoted for SARS-CoV-2 screening, however, antigen tests are preferred in low-income countries and remote zones. Poor training in molecular biology, plus the need for RNA purification or reading instruments to overcome issues of sensitivity in colorimetric detection, are some of the reasons limiting the use of this technique. In this study, nasopharyngeal swabs, aspirates and saliva were amplified in an in-house LAMP assay and subject to colorimetric detection, achieved by the naked eye and by image analysis with a mobile application. Accuracy of detection by the naked eye ranged from 61–74% but improved to 75–86% when using the application. Sensitivity of the digital approach was 81% and specificity 83%, with poor positive predictive value, and acceptable negative predictive value. Additionally to the reported effect of some transport media’s pH, the presence of mucus and warming up of reagents while setting up the reaction critically affected performance. Accuracy per type of sample was 55, 70 and 80%, for swabs, aspirates and saliva, respectively, suggesting potential to improve the test in saliva. This assay, carried out in a closed tube, reduces contamination, has few pipetting steps and requires minimal equipment. Strategies to improve performance and implications of the use this sort of colorimetric LAMP for massive testing are discussed.

1. Introduction

Loop-mediated amplification with reverse transcription (RT-LAMP) has become a popular alternative to real-time polymerase chain reaction (qRT-PCR) in SARS-CoV-2 testing. One year after the World Health Organization declared a pandemic of COVID-19 [1], many protocols using this isothermal technique have been described in peer-reviewed journals. In many cases, detection limits of LAMP are comparable to PCR [2,3,4]. This, combined with a lower cost of amplification enzymes, has been the rationale behind the proliferation of these assays. LAMP is also one of the strategies for target enrichment in microfluidic sensors [5,6,7,8] and other devices being developed for point of care testing of this virus [9,10,11,12].
LAMP assays for SARS-CoV-2 surged in developed economies such as Japan, Korea, China, the United States and, to a lesser extent, the United Kingdom and the European community; among these, a few have emergency use approval from the Food and Drug Administration [13], and the European Centre for Disease Prevention and Control [14]. Early publications presented proof of principle [4,7,15,16,17,18,19,20,21,22,23,24], but those reporting accuracy in hundreds of samples increased during the second epidemic wave [6,25,26,27,28,29,30,31,32,33,34,35,36,37,38]. Although these authors indicate that RT-LAMP might constitute an affordable alternative for mass screening and point of care testing in areas where resources are limited, there are many reasons why this technique is not outcompeting antigen-based tests in middle- and low-income countries. Sensitivity of RT-LAMP in detecting SARS-CoV-2 ranges from 70–100%, depending on many factors [2,28,31,32,39,40]. At least two main categories of sample processing, RNA isolation and RNA release, plus three options for detection, need to be compared. Assays achieving agreement superior to 95% with qRT-PCR are those with RNA purification plus fluorometric [27,38,41,42], turbidimetric [38] or colorimetric detection [25,26,32,33,43,44]. RNA purification incurs a higher cost and longer process, pushing these approaches outside the scope of a rapid test. RNA release, although cheaper, faster and easier to implement, results in lower sensitivity [6,25,29,34,36,43,45,46,47,48].
A few LAMP protocols combining RNA release with colorimetric detection achieve accuracy superior to 80%. Among these, Dao Thi and co-workers [33] processed samples by boiling and detected the color turnout with microplate readers; Schellenberg and Yang, with their collaborators, relied on the naked eye but emphasize that performance is strongly influenced by pH in the sample, which depends on both the type of sample and the transport media [3,37,49]. Implementing LAMP assays by typical reactions in which pipetting of primers, enzyme and other ingredients is needed are time-consuming, at risk of contamination and discouraging for personnel not habituated to techniques in molecular biology; in spite of this disadvantage, LAMP is considered the easiest for rapid training in molecular identification of microbes [50,51]. It is noteworthy that enzymes for RT-LAMP are distributed globally by few providers, representing another limitation for low-income countries, not to mention the probabilities of a shortage in supplies [2,20]. To make colorimetric amplification deployable in primary care centers typical of rural Colombia, we evaluated an RT-LAMP assay that: (1) adopts fast sample processing and reduces pipetting steps, (2) uses closed tubes to minimize cross-contamination and (3) assists/replaces naked-eye detection with affordable software. Accuracy of the approach resembles that of commercial assays already introduced to low-income countries [30,35,52,53], but strategies to improve the performance of colorimetric LAMP and troubleshooting are discussed.

2. Materials and Methods

2.1. Origin of Samples

Nasopharyngeal swab (NPS), nasopharyngeal aspirate (NPA) or saliva (SA) stored in the repository of biological samples of the Laboratorio de Diagnóstico de Agentes Biológicos (LDAB) at the College of Health of the Universidad del Valle were used with authorization from the institutional human research ethics committee. LDAB facilities hold authorization from the Colombian National Institute of Health to perform SARS-CoV-2 testing by real-time PCR in NPSs and NPAs, which arrive in diverse transport media. Saliva was collected without transport media, under two research protocols approved by the institutional ethics committee.

2.2. Real-Time RT-PCR

Amplification was performed in RT-7500 or QuantStudio 5 thermal cyclers (Life Technologies, Carlsbad, CA, USA) following United States CDC recommendations [54]. Reactions contained a total final volume of 20 µL with 5 µL of template RNA, 500 nM of primer/probe mix and 1X SuperScript III One-Step RT-PCR System with Platinum Taq DNA polymerase (Life Technologies Corporation, Carlsbad, CA, USA). Amplification parameters were 50 °C-30 min; 95 °C-2 min; 45 cycles of: 95 °C-15 s, 55 °C-30 s. A cycle threshold (Ct) value of 42, or lower, for the N1 gene was considered to be positive for detection, while samples giving undetectable Ct for N1 but detectable N2 were considered undefined; samples giving RNAseP Ct lower than 42 and undetectable values for of N1 or N2 were considered negative.

2.3. Primers and Initial Conditions for LAMP

Two sets of primers (“Primers set A” [23] and “Primers set B” [55]) were initially evaluated. LAMP reactions were performed in 0.2 mL tubes containing molecular grade water, 12.5 µL of WarmStart master mix (New England Biolabs, Ipswich, MA, USA), plus one of three different preparations of each primer set, then incubated at 65 °C for 30 and 40 min and inactivated at 85 °C. Results were detected by the naked eye with three possible outcomes: negative (sample remains pink), faint (hard to report a change, reaction turns orangish) and positive (sample turns yellow); correspondence between color changes and amplification was evaluated by gel electrophoresis with four possible outcomes: positive (bands with the expected size pattern), negative (only primer bands detectable), inefficient amplification (smear within the expected sizes) and reduced specificity (bands with an abnormal size pattern). Under these conditions, “Primers set A” yielded reduced specificity (data not shown) and therefore findings correspond to “Primers set B” from here on.
Primers set B: nts 13,434–13,636, ORF1ab
BF3: TGCTTCAGTCAGCTGATG
BB3: TTAAATTGTCATCTTCGTCCTT
BFIP: TCAGTACTAGTGCCTGTGCCCACAATCGTTTTTAAACGGGT
BBIP: TCGTATACAGGGCTTTTGACATCTATCTTGGAAGCGACAACAA
BLoopF: CTGCACTTACACCGCAA
BLoopB: GTAGCTGGTTTTGCTAAATTCC

2.4. Mobile-Assisted Visualization

To solve cases where the naked eye finds undefined results, the mobile version of SpotXel (GmbH, Germersheim, Germany) was used [56]. Reaction tubes were photographed before and after incubation at 65 °C; for this, tubes were accommodated in horizontal racks in a set up that positions the cell phone at a predefined height and illumination. Photographs were later used to obtain values of color intensity, using the microarray option of SpotXel (referred as “yellow intensity” from here on) with the grid adjusted to one square per tube; recorded values were both exported in tables and stored as screen shots. Samples yielding ≥ 1.0-fold change in the value of yellow intensity were considered positive.

2.5. Optimization to Avoid RNA Purification

Four strategies for sample processing were tested prior to evaluating the sensitivity and specificity of a final protocol with “Primers set B”. RNA purification was performed with the AccuPrep® viral RNA extraction kit (Bioneer Corporation, Daejeon, Korea) following the manufacturer’s instructions. Fast processing was carried out by mixing 50 µL of the sample 1:1 in PCRopsis Reagent RVD (Entopsis, Miami, FL, USA) and boiling at 95 °C for five minutes, as indicated by the manufacturers. Inexpensive processing was carried out by diluting 20 µL of the sample 1:1 in PBS or DEPC water and boiling at 95 °C for five minutes. The outcome was evaluated as the percentage of positive and negative agreement by gel electrophoresis, visual detection and mobile application as described above.

2.6. Rapid LAMP Strategy

“Primers set B” were prepared to final concentrations and stored in stocks of 24 reactions at −20 °C; parallel stocks of 24 tubes containing 20 µL of “lysis buffer” (DEPC water, Sigma-Aldrich, Darmstadt, Germany) were stored at room temperature. For sample processing, one tube of primers and one tube of lysis buffer per sample, plus positive and negative controls, were used in the next steps: (i) in the sample manipulation area, using a microbiological cabinet, 20 µL of the sample was added to “lysis buffer”, then heated at 95 °C for five minutes in a thermal block; (ii) in the clean area, 12.5 µL of WarmStart enzymes was added to the prepared primers; (iii) back in the sample manipulation area, 2.0 µL of boiled sample was added to the primers + enzyme mix, mixed thoroughly and incubated for 40 min at 65 °C in a thermal block. Results of visual and application-assisted detection were recorded as described above.

2.7. Detection Limit of the Rapid Assay

Detection limits were established by spiking NSP, NPA and saliva. Briefly, 50 µL of total RNA purified from a positive sample carrying 871.63 × 106 copies/µL of SARS-CoV-2 genetic material (measured with a curve of known amounts of synthetic positive control) was added to 450 µL of NSP, NPA or saliva previously established to have undetectable amounts of viral RNA. Serial dilutions were performed and aliquots from eight preps were subject to RNA purification and qRT-PCR to obtain a copy number, and to a rapid protocol to identify the range of detection. The outcomes were established by the visual and application-assisted methodologies described above.

2.8. Cross-Reactivity

Nasopharyngeal aspirates positive for eight of the most frequent respiratory viruses in Colombia (adenovirus, influenza A and B, respiratory syncytial virus, metapneumovirus and parainfluenza viruses 1, 2 and 3) and a nasopharyngeal swab containing human coronavirus OC43 were tested with the rapid LAMP procedure. Results were recorded by visual detection and application-assisted methodologies, as previously described.

2.9. Performance of the Rapid LAMP Assay

A total of 208 samples comprising 86 NPS, 62 NPA and 60 saliva samples were tested. Tubes were photographed before and after incubation at 65 °C for 40 min and results of visual observation and values of image analysis were tabulated. Agreement with qRT-PCR, sensitivity and other statistics of performance of the visual and the mobile-assisted detection approaches were calculated with SPP, and plotted in R (v.3.5.1, R Foundation for Statistical Computing, Vienna, Austria) using ggplot2. RT-qPCR experiments were analyzed using QuantStudio Design and Analysis Software (v.2.3.3, ThermoFisher, Waltham, MA, USA), or exported for analysis in R software.

3. Results

3.1. Setting Up the Conditions

The colorimetric assay was first tested under the ideal conditions, i.e., with purified RNA, which gave positive outcomes observable down to Ct from 30–33 (Figure 1). Those Ct values correspond to ~621–97 copies of SARS-CoV-2 genome/µL in the real-time protocol used. At this point, the RT-LAMP assay did not generate faint results and the colorimetric detection coincided with electrophoresis in all repeats. Visual detection was feasible with RVD in those samples with Ct lower than 23, reaching an agreement of 62.5% in the three repeats. Pre-processing with PBS and direct addition of the sample gave the lowest percentages of agreement, which were also inconsistent between gel, color turnout and the application; for these treatments, the electrophoresis indicated null or inefficient amplification even in samples with high viral loads. The use of DEPC water allowed amplification for samples with Ct up to 23 that matched electrophoresis, visual outcome and fold change in yellow intensity; if inefficient amplification is counted as positive by contrast to the lack of a smear in negative samples, the accuracy of this treatment in electrophoretic detection is 87.5%. DEPC processed samples with Ct lower than 23 had negative visual outcomes but the changes in color intensity were enough to be classified as positive by the application, however, fold change values in these samples lacked a clear match to viral load, probably because samples consisted of randomly distributed 60% NPS and 40% NPA. Overall, boiling the sample in water gave closer results to RNA purification, with a total agreement of 75% in the visual detection and 100% when using SpotXel. It is worth noting that with the use of purified RNA or pre-processing in RVD, the reaction starts at a higher pH compared to the other strategies; this is a factor against colorimetric detection that will be addressed below.

3.2. Detection Limit of a Selected Strategy

The rapid, mobile-assisted protocol correctly detected positives when testing samples with 22,594, 112,814 and 88,682 copies/µL of SARS-CoV-2 RNA in NPSs, NPAs and saliva, respectively. These values correspond to Cts from 21–25, in agreement with the preliminary findings presented above for H20 + boiling. Detection was still obtained in other dilutions but repeatability dropped for both visual and application-based detection (Table 1).

3.3. Cross-Reactivity

The assay did not produce positive results when used with samples known to have other respiratory viruses (Table 2). Most samples, however, were nasopharyngeal aspirates and this finding may not translate to other types of samples.

3.4. Performance of the Visual Approach

This approach gave an accuracy of 68%, with an elevated number of false negatives (summarized in Table 3). Global agreement with qRT-PCR, calculated as global kappa index, is fair (0.34), however, after analyzing data by type of sample, better agreement is observed with saliva (0.44, 0.24–0.66), followed by NPA (0.03, −0.20–0.26), and NPS (0.02, 0.16–0.20) (see Supplementary Table S1). The high number of faint results is associated with mucus in the sample and the viral load (see Figure 2). Those samples classified as mucus resulted more often in false positives than those with any other type of appearance; there are false negatives in all categories of sample quality but they concentrate in groups with Ct values above 23 (Figure 2), which again represents the characteristic sensitivity identified during optimization. Interestingly, bloody samples and those collected in transport media containing pH indicator (fuchsia), which remain reddish after boiling, did not tend to give false negatives.

3.5. Performance of Assay with Mobile Application

The accuracy of this approach is nearly 81%, which is below the 85% threshold that the Colombian Institute of Health has recommended for SARS-CoV-2 diagnostics (Table 3). There was a moderate agreement with the qRT-PCR protocol (kappa index = 0.55) which, as discussed below, is related not only to detection limits but to technical issues and a considerable number of false positives. Performance was slightly better in saliva (79.5% of accuracy) compared to NPS and NPA (~60% and 70% accuracy, respectively), which suggests a potential to improve detection in saliva after further optimizations of the assays. Fold change has clear differences in the average between true positives and true negatives (1.20 ± 0.16 vs. 0.89 ± 0.07, respectively) (Table 3). This represents changes of 22–32% in yellow intensity, which, as illustrated in Figure 3, has a relation to viral load, but there is no linearity. Variation in fold change was high in negatives, but also in intermediate samples with Ct from 16–28; this also occurred in data from the limit of detection (data no shown), and translates into more false positives in these groups compared to those of Ct > 28.
Figure 4 shows the performance of the imaging approach in terms of viral load and its potential use for SARS-CoV-2 screening. In both symptomatic and asymptomatic patients, accuracy of the assay is higher than 85% for those samples with Cts below 23, however, those with lower viral loads gave less accurate results, especially in symptomatic patients. As indicated by the standard deviations, performance varied among types of samples, with a higher reduction in the case of NPS, in which performance is from 60–50% above Ct 23; in NPA, accuracy diminished below 85% above Ct 29, whereas in saliva accuracy dropped below 50% in Ct groups 23–28 and 29–36 but increased again in Ct group 37–42 (Supplementary Figure S1).

4. Discussion

The detection limit with purified RNA was at some point between 621 and 97 copies of viral RNA/µL (Figure 1), but between ~112,000 and ~22,000 copies with the application. Similar to other RT-LAMP assays for SARS-CoV-2, boiling the samples in water was suitable to avoid RNA purification [3,37,49], however, when used in the visual detection mode, this assay is inaccurate; with the mobile application, the assay’s accuracy reaches up to 81%, resembling that of some commercial tests with reduced performance already introduced to low-income countries [30,35,52,53]. There is a sort of recovery in performance in samples with Ct above 29 in NPS and saliva samples, in both the visual (data not shown) and application approach (Supplementary Figure S1); the finding of a similar trend in the detection limit assays suggests a behavior of the analyte, which would explain the lack of linearity in values of fold change described in Figure 3; perhaps other mathematical approaches to transform data would adjust better to this behavior.
In addition to those previously reported, other factors that affect the performance of colorimetric LAMPs were identified. Those samples collected in sodium azide were found unfit for LAMP (data not shown), whereas those containing excessive mucus are very likely to give false positives. Other authors have shown that saliva samples have a decreased sensitivity in qRT-PCR [57] and LAMP [58,59] due to variable pH, which affects the RNA purification protocol and likely amplification, resulting in quantification that deviated from other types of samples. Mucus samples also represent difficulties for qRT-PCR which are solved by diluting and/or modifying the RNA purification methodology; here, the LAMP assays for mucus samples were carried out with material already diluted in UTM for qRT-PCR, therefore other strategies are needed to overcome this issue.
Initial intensity of pink as a factor influencing performance of both visual and application-assisted detection was left undocumented, but it is worthy of discussion. Reactions processed with purified RNA and RVD started with the deepest pink color if compared to those where the sample was processed by other strategies. This, although related to the pH of the respective buffers, seems to also depend on the sensitivity of phenol red to environmental conditions. As noticed by Dao Thi and co-workers [33], commercial reagents for colorimetric LAMP based on phenol red cannot be prepared in advance because they acidify with long exposure to air. We found that keeping the reaction mix on ice and tubes closed while adding other samples helps to maintain pink uniformity; this was incorporated as a routine procedure after optimization.
Accuracy of the mobile-assisted approach was similar to other tests of colorimetric RT-LAMP plus boiling in nasopharyngeal swabs, detecting either the N gene [33] or the E gene [49]. These findings suggests that abundance and secondary structures of the target, in this case the ORF1ab overlap, have low influence on performance. More interestingly, saliva performed better than other types of samples and has potential to be improved. Yang and collaborators [3] optimized colorimetric RT-LAMP in saliva by using a pH stabilizer with NaOH and primers for the E gene. Based on the standard deviation of fold change (Table 4), it seems that aspirates and swabs exhibit a less flexible relation of color turnout to viral load compared to saliva, likely because they are often collected in buffered solutions. This is supported by the homogeneity of fold change among positive samples when using RVD as a pH stabilizer (Figure 1). Although stabilizing saliva is acceptable for qualitative detection, it might tamper with a semi-quantitative approach because the relation of yellow intensity to viral load appears to be more linear in non-buffered samples.
Double photographing of the tubes to obtain the fold change is the main disadvantage of the application-based approach. It is time-consuming, requires uniformity of light and the same equipment to acquire both images; after identifying other factors that influence the behavior of the analyte, the inclusion of a curve in each experiment might be enough to sort out negatives and positives; setting up cutoff values for yellow intensity after incubation seems viable with a cutoff of 100, although the correlation between yellow intensity after incubation and Ct is less direct than Ct vs. fold change (Supplementary Figure S2). Although reading plates are more likely to be cross-contaminated than closed tubes in the scenarios imagined for this assay, the approach cannot be discarded for low-income countries. While regular polystyrene plates do not resist incubation at 65 °C, this assay, if transferred to ELISA reading systems in sealed polypropylene plates, might allow not only a more uniform environment with respect to light source, but also the ability to test more samples simultaneously; if the software has a similar relationship between yellow intensity and viral load in plates with a transparent seal, the assay could be performed without a need to purchase equipment, but issues in reading uniformity might still need local solutions.

5. Conclusions

This assay did not reach national standards to be used as a diagnostic test for SARS-CoV-2 in Colombia, but was able to identify symptomatic and asymptomatic persons with high viral loads in all kinds of samples tested. In comparison, commercial antigen-based tests, as reviewed by Dinnes et al. (2021) [60], have a sensitivity of 72–79.0% in symptomatic and 51–74% in asymptomatic patients, with major limitations in those patients sampled more than ten days after having acquired the virus. Although viral loads do not match symptomatology by being influenced by timing and expression of sub-genomic RNAs, high viral loads correlate with infectivity [61,62] and have been linked to index cases and super-spreading events [37,63]. Although viral loads are influenced by timing, they do not match symptomatology and are associated with expression of sub-genomic RNAs, high viral load correlates with infectivity [61,62] and have been linked to index cases and super-spreading events [37,63]. The Abbott ID Now assay and other low-sensitivity but widely marketed SARS-CoV-2 tests avoid the use of microbiological cabinets by using azides or NaOH during sample collection. This LAMP assay is still subject to laboratory infrastructure, but, similar to them, can easily spot individuals with high viral loads, the ones more likely to transmit the virus. Low-sensitivity tests, initially intended for point of care diagnosis, are being widely used to facilitate the return of employees who have entered quarantine and for routinely testing personnel in hospitals and schools. Implementing sample self-collection and inactivation protocols might allow the use of this assay in laboratory-free facilities, which is more likely to be achieved if sampling saliva.
The set of primers used in this study was designed to amplify the overlap between ORF1a and ORF1b genes, implying that the true positives identified are patients who actually harbor full genomes of SARS-CoV-2. In consequence, this strategy serves for detecting potential super-spreaders and its use in massive screening can have a direct impact on reducing transmission. Further studies to correlate day of contagion and symptom onset with performance of the test might reveal its suitability for early diagnostics and decision making in clinical or epidemiological scenarios.
Thus far, in-house RT-LAMP for detection of SARS-CoV2 in Latin America has been validated in Peru [64,65]. This is the first attempt to develop a local protocol and achieve efficient detection with minimal cost in Colombia. Although this assay does not meet all the affordable, specific, sensitive, user-friendly, rapid, equipment-free and deliverable to the user (ASSURED) guidelines for low-resource settings proposed by the World Health Organization [60], it has advantages in the context of the COVID-19 pandemic. A test with reduced sensitivity generates false negatives which can be compensated by re-testing, whereas if specificity is reduced, the frequent finding of false positives does not affect transmissibility but impacts the economy. Based on our results, an assay for saliva is more likely to meet the ASSURED criteria or to be used combined with other inexpensive strategies because it has the highest sensitivity and its sampling has the lower cost.

Supplementary Materials

The following supporting information can be downloaded at: https://www.mdpi.com/article/10.3390/diagnostics12040848/s1, Table S1: Performance of visual detection per type of sample. Figure S1: Performance of the mobile-assisted detection approach by type of sample. Figure S2: Relation between yellow intensity and viral load after incubation.

Author Contributions

All authors participated in designing the study and reviewing the whole manuscript. M.A.L.-A. conducted LAMP assays, B.P. conducted qRT-PCR assays, G.L. managed biological, administrative, economic and human resources, L.O. supervised epidemiological concerns during the study. All authors have read and agreed to the published version of the manuscript.

Funding

This work was funded by Universidad del Valle (grant 1851), and Sistema General de Regalias (grant 2020000100109).

Institutional Review Board Statement

The study was conducted according to the guidelines of the Declaration of Helsinki, with usage of stored samples approved by the Institutional Committee for the Review of Human Ethics of the College of Health, Universidad del Valle.

Informed Consent Statement

Patient consent for nasopharyngeal swabs and aspirates was waived based in the authorization that LDAB facilities hold to keep a repository of diagnostics samples; posterior or parallel usage of samples is done with authorization of the institutional ethics committee. Saliva samples were obtained with written informed consent, following the procedures authorized by the institutional ethics committee.

Acknowledgments

We are especially thankful for students, professionals and other staff participating in “Equipo COVID 2020–2021”. We are also grateful to Adolfo Contreras, Universidad del Valle, for sharing aliquots of saliva samples summitted to qRT-PCR in LDAB facilities.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Cucinotta, D.; Vanelli, M. WHO Declares COVID-19 a Pandemic. Acta Biomed. 2020, 91, 157–160. [Google Scholar] [CrossRef] [PubMed]
  2. Thompson, D.; Lei, Y. 2020. Mini Review: Recent Progress in RT-LAMP Enabled COVID-19 Detection. Sens. Actuators Rep. 2020, 2, 100017. [Google Scholar] [CrossRef] [PubMed]
  3. Yang, Q.; Meyerson, N.R.; Clark, S.K.; Paige, C.L.; Fattor, W.T.; Gilchrist, A.R.; Barbachano-Guerrero, A.; Healy, B.G.; Worden-Sapper, E.R.; Wu, S.S.; et al. Saliva TwoStep for Rapid Detection of Asymptomatic SARS-CoV-2 Carriers. Elife 2021, 10, e65113. [Google Scholar] [CrossRef]
  4. Mautner, L.; Baillie, C.K.; Herold, H.M.; Volkwein, W.; Guertler, P.; Eberle, U.; Ackermann, N.; Sing, A.; Pavlovic, M.; Goerlich, O.; et al. Rapid Point-of-Care Detection of SARS-CoV-2 Using Reverse Transcription Loop-Mediated Isothermal Amplification (RT-LAMP). Virol. J. 2020, 17, 160. [Google Scholar] [CrossRef]
  5. Ali, Z.; Aman, R.; Mahas, A.; Rao, G.S.; Tehseen, M.; Marsic, T.; Mahfouz, M.M. iSCAN: An RT-LAMP-Coupled CRISPR-Cas12 Module for Rapid, Sensitive Detection of SARS-CoV-2. Virus Res. 2020, 288. [Google Scholar] [CrossRef]
  6. Schermer, B.; Fabretti, F.; Damagnez, M.; Di Cristanziano, V.; Heger, E.; Arjune, S.; Tanner, N.A.; Imhof, T.; Koch, M.; Ladha, A.; et al. Rapid SARS-CoV-2 Testing in Primary Material Based on a Novel Multiplex RT-LAMP Assay. PLoS ONE 2020, 15, e0238612. [Google Scholar] [CrossRef]
  7. Zhu, X.; Wang, X.; Han, L.; Chen, T.; Wang, L.; Li, H.; Li, S.; He, L.; Fu, X.; Chen, S.; et al. Multiplex Reverse Transcription Loop-Mediated Isothermal Amplification Combined with Nanoparticle-Based Lateral Flow Biosensor for the Diagnosis of COVID-19. Biosens. Bioelectron. 2020, 166, 112437. [Google Scholar] [CrossRef]
  8. Huang, W.E.; Lim, B.; Hsu, C.C.; Xiong, D.; Wu, W.; Yu, Y.; Jia, H.; Wang, Y.; Zeng, Y.; Ji, M.; et al. RT-LAMP for Rapid Diagnosis of Coronavirus SARS-CoV-2. Microb. Biotechnol. 2020, 13, 950–961. [Google Scholar] [CrossRef] [Green Version]
  9. Brandsma, E.; Verhagen, H.; van de Laar, T.; Claas, E.; Cornelissen, M.; van den Akker, E. Rapid, Sensitive and Specific SARS Coronavirus-2 Detection: A Multi-Center Comparison between Standard qRT-PCR and CRISPR Based DETECTR. J. Infect. Dis. 2020, 2, 206–213. [Google Scholar] [CrossRef]
  10. Wang, R.; Qian, C.; Pang, Y.; Li, M.; Yang, Y.; Ma, H.; Zhao, M.; Qian, F.; Yu, H.; Liu, Z.; et al. opvCRISPR: One-Pot Visual RT-LAMP-CRISPR Platform for SARS-cov-2 Detection. Biosens. Bioelectron. 2020, 172, 112766. [Google Scholar] [CrossRef]
  11. Hou, T.; Zeng, W.; Yang, M.; Chen, W.; Ren, L.; Ai, J.; Wu, J.; Liao, Y.; Gou, X.; Li, Y.; et al. Development and Evaluation of a Rapid CRISPR-Based Diagnostic for COVID-19. PLoS Pathog. 2020, 16, e1008705. [Google Scholar] [CrossRef] [PubMed]
  12. Broughton, J.P.; Deng, X.; Yu, G.; Fasching, C.L.; Singh, J.; Streithorst, J.; Granados, A.; Sotomayor-Gonzalez, A.; Zorn, K.; Gopez, A.; et al. Rapid Detection of 2019 Novel Coronavirus SARS-CoV-2 Using a CRISPR-Based DETECTR Lateral Flow Assay. medRxiv 2020. [Google Scholar] [CrossRef]
  13. Food and Drug Administration. In Vitro Diagnostics EUAs—Molecular Diagnostic Tests for SARS-CoV-2. Available online: https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-authorizations-medical-devices/vitro-diagnostics-euas#individual-molecular (accessed on 31 October 2021).
  14. European Center for Disease Prevention and Control. Diagnostic Testing and Screening for SARS-CoV-2. Available online: https://www.ecdc.europa.eu/en/novel-coronavirus/laboratory-support (accessed on 31 October 2021).
  15. Lalli, M.A.; Langmade, J.S.; Chen, X.; Fronick, C.C.; Sawyer, C.S.; Burcea, L.C.; Wilkinso, M.N.; Fulton, R.S.; Heinz, M.; Buchser, W.J.; et al. Rapid and Extraction-Free Detection of SARS-CoV-2 from Saliva by Colorimetric Reverse-Transcription Loop-Mediated Isothermal Amplification. Clin. Chem. 2021, 67, 415–424. [Google Scholar] [CrossRef]
  16. Österdahl, M.F.; Lee, K.A.; Lochlainn, M.N.; Wilson, S.; Douthwaite, S.; Horsfall, R.; Sheedy, A.; Goldenberg, S.D.; Stanley, C.J.; Spector, T.D.; et al. Detecting SARS-CoV-2 at Point of Care: Preliminary Data Comparing Loop-Mediated Isothermal Amplification (LAMP) to Polymerase Chain Reaction (PCR). BMC Infect. Dis. 2020, 20, 783. [Google Scholar] [CrossRef] [PubMed]
  17. Ganguli, A.; Mostafa, A.; Berger, J.; Aydin, M.Y.; Sun, F.; Stewart de Ramirez, S.A.; Bashir, R. Rapid Isothermal Amplification and Portable Detection System for SARS-CoV-2. Proc. Nat. Acad. Sci. USA 2020, 117, 22727–22735. [Google Scholar] [CrossRef] [PubMed]
  18. Kim, J.H.; Kang, M.; Park, E.; Chung, D.R.; Kim, J.; Hwang, E.S. A Simple and Multiplex Loop-Mediated Isothermal Amplification (LAMP) Assay for Rapid Detection of SARS-CoV. Biochip J. 2019, 13, 341–351. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  19. Zhang, Y.; Ren, G.; Buss, J.; Barry, A.J.; Patton, G.C.; Tanner, N.A. Enhancing Colorimetric Loop-Mediated Isothermal Amplification Speed and Sensitivity with Guanidine Chloride. BioTechniques 2020, 69, 178–185. [Google Scholar] [CrossRef]
  20. Alekseenko, A.; Barrett, D.; Pareja-Sanchez, Y.; Howard, R.J.; Strandback, E.; Ampah-Korsah, H.; Rovšnik, U.; Zuniga-Veliz, S.; Klenov, A.; Malloo, J.; et al. Direct Detection of SARS-CoV-2 Using Non-Commercial RT-LAMP Reagents on Heat-Inactivated Samples. Sci. Rep. 2021, 11, 1820. [Google Scholar] [CrossRef]
  21. Varona, M.; Eitzmann, D.R.; Anderson, J.L. Sequence-Specific Detection of ORF1a, BRAF, and ompW DNA Sequences with Loop Mediated Isothermal Amplification on Lateral Flow Immunoassay Strips Enabled by Molecular Beacons. Anal. Chem. 2021, 93, 4149–4153. [Google Scholar] [CrossRef]
  22. Kitagawa, Y.; Orihara, Y.; Kawamura, R.; Imai, K.; Sakai, J.; Tarumoto, N.; Maeda, T. Evaluation of Rapid Diagnosis of Novel Coronavirus Disease (COVID-19) Using Loop-Mediated Isothermal Amplification. J. Clin. Virol. 2020, 129, 104446. [Google Scholar] [CrossRef]
  23. Lamb, L.E.; Bartolone, S.N.; Ward, E.; Chancellor, M.B. Rapid Detection of Novel Coronavirus/Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) by Reverse Transcription-Loop-Mediated Isothermal Amplification. PLoS ONE 2020, 15, e0234682. [Google Scholar] [CrossRef]
  24. Lu, R.; Wu, X.; Wan, Z.; Li, Y.; Jin, X.; Zhang, C. A Novel Reverse Transcription Loop-Mediated Isothermal Amplification Method for Rapid Detection of Sars-Cov-2. Int. J. Mol. Sci. 2020, 21, 2826. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  25. Baek, Y.H.; Um, J.; Antigua, K.; Park, J.H.; Kim, Y.; Oh, S.; Kim, Y.I.; Choi, W.S.; Kim, S.G.; Jeong, J.H.; et al. Development of a Reverse Transcription-Loop-Mediated Isothermal Amplification as a Rapid Early-Detection Method for Novel SARS-CoV-2. Emerg. Microb. Infect. 2020, 9, 998–1007. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  26. Rohaim, M.A.; Clayton, E.; Sahin, I.; Vilela, J.; Khalifa, M.E.; Al-Natour, M.Q.; Munir, M. Artificial Intelligence-Assisted Loop Mediated Isothermal Amplification (AI-LAMP) for Rapid Detection of SARS-CoV-2. Viruses 2020, 12, 972. [Google Scholar] [CrossRef] [PubMed]
  27. Mohon, A.N.; Oberding, L.; Hundt, J.; van Marle, G.; Pabbaraju, K.; Berenger, B.M.; Lisboa, L.; Griener, T.; Czub, M.; Doolan, C.; et al. Optimization and Clinical Validation of Dual-Target RT-LAMP for SARS-CoV-2. J. Virol. Methods 2020, 286, 113972. [Google Scholar] [CrossRef]
  28. Rödel, J.; Egerer, R.; Suleyman, A.; Sommer-Schmid, B.; Baier, M.; Henke, A.; Löffler, B. Use of the variplextm SARS-CoV-2 RT-LAMP as a Rapid Molecular Assay to Complement RT-PCR for COVID-19 Diagnosis. J. Clin. Virol. 2020, 132, 104616. [Google Scholar] [CrossRef]
  29. Hu, X.; Deng, Q.; Li, J.; Chen, J.; Wang, Z.; Zhang, X.; Hou, T. Development and Clinical Application of a Rapid and Sensitive Loop-Mediated Isothermal Amplification Test for SARS-CoV-2 Infection. MSphere. 2020, 5, e00808–e00820. [Google Scholar] [CrossRef]
  30. Nagura-Ikeda, M.; Imai, K.; Tabata, S.; Miyoshi, K.; Murahara, N.; Mizuno, T.; Kato, Y. Clinical Evaluation of Self-Collected Saliva by Quantitative Reverse Transcription-PCR (RT-qPCR), Direct RT-qPCR, Reverse Transcription-Loop-Mediated Isothermal Amplification, and a Rapid Antigen Test to Diagnose COVID-19. J. Clin. Microbiol. 2020, 58, e01438-20. [Google Scholar] [CrossRef]
  31. Lee, J.Y.H.; Best, N.; McAuley, J.; Porter, J.L.; Seemann, T.; Schultz, M.B.; Stinear, T.P. Validation of a Single-Step, Single-Tube Reverse Transcription Loop-Mediated Isothermal Amplification Assay for Rapid Detection of SARS-CoV-2 RNA. J. Med. Microbiol. 2020, 69, 1169–1178. [Google Scholar] [CrossRef]
  32. Chow, F.W.N.; Chan, T.T.Y.; Tam, A.R.; Zhao, S.; Yao, W.; Fung, J.; Lau, S.K.P. A Rapid, Simple, Inexpensive, and Mobile Colorimetric Assay Covid-19-Lamp for Mass on-Site Screening of Covid-19. Int. J. Mol. Sci. 2020, 21, 5380. [Google Scholar] [CrossRef]
  33. Dao Thi, V.L.; Herbst, K.; Boerner, K.; Meurer, M.; Kremer, L.P.M.; Kirrmaier, D.; Anders, S. A Colorimetric RT-LAMP Assay and LAMP-Sequencing for Detecting SARS-CoV-2 RNA in Clinical Samples. Sci. Transl. Med. 2020, 556, eabc7075. [Google Scholar] [CrossRef] [PubMed]
  34. Parker, C.W.; Singh, N.; Tighe, S.; Blachowicz, A.; Wood, J.M.; Seuylemezian, A.; Vaishampayan, P.; Urbaniak, C.; Hendrickson, R.; Laaguiby, P.; et al. End-to-End Protocol for the Detection of SARS-CoV-2 from Built Environments. mSystems 2020, 5, e00771-20. [Google Scholar] [CrossRef] [PubMed]
  35. Eckel, F.; Küsters, F.; Drossel, B.; Konert, M.; Mattes, H.; Schopf, S. Variplex™ Test System Fails to Reliably Detect SARS-CoV-2 Directly from Respiratory Samples without RNA Extraction. Eur. J. Clin. Microbiol. Infect. Dis. 2020, 39, 2373–2377. [Google Scholar] [CrossRef] [PubMed]
  36. Yu, L.; Wu, S.; Hao, X.; Dong, X.; Mao, L.; Pelechano, V.; Chen, W.H.; Yin, X. Rapid Detection of COVID-19 Coronavirus Using a Reverse Transcriptional Loop-Mediated Isothermal Amplification (RT-LAMP) Diagnostic Platform. Clin. Chem. 2020, 66, 975–977. [Google Scholar] [CrossRef] [PubMed]
  37. Yang, Q.; Saldi, T.K.; Gonzales, P.K.; Lasda, E.; Decker, C.J.; Tat, K.L.; Fink, M.R.; Hager, C.R.; Davis, J.C.; Ozeroff, C.D.; et al. Just 2% of SARS-CoV-2-Positive Individuals Carry 90% of the Virus Circulating in Communities. Proc. Natl. Acad. Sci. USA 2021, 118, e2104547118. [Google Scholar] [CrossRef]
  38. Yan, C.; Cui, J.; Huang, L.; Du, B.; Chen, L.; Xue, G.; Yuan, J. Rapid and Visual Detection of 2019 Novel Coronavirus (SARS-CoV-2) by a Reverse Transcription Loop-Mediated Isothermal Amplification Assay. Clin. Microb. Infect. 2020, 26, 773–779. [Google Scholar] [CrossRef]
  39. Dong, Y.; Wu, X.; Li, S.; Lu, R.; Li, Y.; Wan, Z.; Qin, J.; Yu, G.; Jin, X.; Zhang, C. Comparative Evaluation of 19 Reverse Transcription Loop-Mediated Isothermal Amplification Assays for Detection of SARS-CoV-2. Sci. Rep. 2021, 11, 2936. [Google Scholar] [CrossRef]
  40. Janíková, M.; Hodosy, J.; Boor, P.; Klempa, B.; Celec, P. Loop-Mediated Isothermal Amplification for the Detection of SARS-CoV-2 in Saliva. Microb. Biotechnol. 2021, 14, 307–316. [Google Scholar] [CrossRef]
  41. Pham, J.; Meyer, S.; Nguyen, C.; Williams, A.; Hunsicker, M.; McHardy, I.; Gendlina, I.; Goldstein, D.Y.; Fox, A.S.; Hudson, A.; et al. Performance Characteristics of a High-Throughput Automated Transcription-Mediated Amplification Test for SARS-CoV-2 Detection. J. Clin. Microbiol. 2020, 58, e01669-20. [Google Scholar] [CrossRef]
  42. Howson, E.L.A.; Kidd, S.P.; Armson, B.; Goring, A.; Sawyer, J.; Cassar, C.; Cross, D.; Lewis, T.; Hockey, J.; Rivers, S.; et al. Preliminary Optimisation of a Simplified Sample Preparation Method to Permit Direct Detection of SARS-CoV-2 within Saliva Samples Using Reverse-Transcription Loop-Mediated Isothermal Amplification (RT-LAMP). J. Virol. Methods 2021, 289, 114048. [Google Scholar] [CrossRef]
  43. Jiang, M.; Pan, W.; Arasthfer, A.; Fang, W.; Ling, L.; Fang, H.; Daneshnia, F.; Yu, J.; Liao, W.; Pei, H.; et al. Development and Validation of a Rapid, Single-Step Reverse Transcriptase Loop-Mediated Isothermal Amplification (RT-LAMP) System Potentially to Be Used for Reliable and High-Throughput Screening of COVID-19. Front. Cell. Infect. Microbiol. 2020, 10, 331. [Google Scholar] [CrossRef] [PubMed]
  44. Nawattanapaiboon, K.; Pasomsub, E.; Prombun, P.; Wongbunmak, A.; Jenjitwanich, A.; Mahasupachai, P.; Vetcho, P.; Chayrach, C.; Manatjaroenlap, N.; Samphaongern, C.; et al. Colorimetric Reverse Transcription Loop-Mediated Isothermal Amplification (RT-LAMP) as a Visual Diagnostic Platform for the Detection of the Emerging Coronavirus SARS-CoV-2. Analyst 2021, 146, 471–477. [Google Scholar] [CrossRef] [PubMed]
  45. Lau, Y.L.; Ismail, I.B.; Izati Binti Mustapa, N.; Lai, M.Y.; Tuan Soh, T.S.; Hassan, A.H.; Peariasamy, K.M.; Lee, Y.L.; Goh, P.P. A Sensitive Reverse Transcription Loop-Mediated Isothermal Amplification Assay for Direct Visual Detection of SARS-CoV-2. Am. J. Trop. Med. Hyg. 2020, 103, 2350–2352. [Google Scholar] [CrossRef]
  46. Ben-Assa, N.; Naddaf, R.; Gefen, T.; Capucha, T.; Hajjo, H.; Mandelbaum, N.; Elbaum, L.; Rogov, P.; King, D.A.; Kaplan, S.; et al. Direct on-the-Spot Detection of SARS-CoV-2 in Patients. Exp. Biol. Med. (Maywood N. J.) 2020, 245, 1187–1193. [Google Scholar] [CrossRef] [PubMed]
  47. Lau, Y.L.; Ismail, I.; Mustapa, N.I.; Lai, M.Y.; Tuan Soh, T.S.; Hassan, A.; Peariasamy, K.M.; Lee, Y.L.; Chong, Y.M.; Sam, I.C.; et al. Real-Time Reverse Transcription Loop-Mediated Isothermal Amplification for Rapid Detection of SARS-CoV-2. Peer J. 2020, 8, e9278. [Google Scholar] [CrossRef]
  48. Fowler, V.L.; Armson, B.; Gonzales, J.L.; Wise, E.L.; Howson, E.L.A.; Vincent-Mistiaen, Z.; Fouch, S.; Maltby, C.J.; Grippon, S.; Munro, S.; et al. A Highly Effective Reverse-Transcription Loop-Mediated Isothermal Amplification (RT-LAMP) Assay for the Rapid Detection of SARS-CoV-2 Infection. J. Infect. 2021, 82, 117–125. [Google Scholar] [CrossRef] [PubMed]
  49. Schellenberg, J.J.; Ormond, M.; Keynan, Y. Extraction-Free RT-LAMP to Detect SARS-CoV-2 Is Less Sensitive but Highly Specific Compared to Standard RT-PCR in 101 Samples. J. Clin. Virol. 2021, 136, 104764. [Google Scholar] [CrossRef]
  50. Hamburger, J.; Abbasi, I.; Kariuki, C.; Wanjala, A.; Mzungu, E.; Mungai, P.; Muchiri, E.; King, H.K. Evaluation of Loop-Mediated Isothermal Amplification Suitable for Molecular Monitoring of Schistosome-Infected Snails in Field Laboratories. Am. J. Trop. Med. Hyg. 2013, 88, 344–351. [Google Scholar] [CrossRef]
  51. Nguyen, A.V.; Orlofsky, A.; Pubill, K.; Tawde, M.; Li, G.; Mata, D.; Bermudes, O.; Fernandez, M.; Santana, J.; Kim, W.; et al. Loop-Mediated Isothermal Amplification (LAMP) as a Rapid, Affordable and Effective Tool to Involve Students in Undergraduate Research. Front. Microbiol. 2020, 11, 603381. [Google Scholar] [CrossRef]
  52. Freire-Paspuel, B.; Garcia-Bereguiain, M.A. Low Clinical Performance of the Isopollo COVID-19 Detection Kit (M Monitor, South Korea) for RT-LAMP SARS-CoV-2 Diagnosis: A Call for Action against Low Quality Products for Developing Countries. Int. J. Infect. Dis. 2021, 104, 303–305. [Google Scholar] [CrossRef]
  53. Wilson-Davies, E.S.W.; Mahanama, A.I.K.; Samaraweera, B.; Ahmed, N.; Friar, S.; Pelosi, E. Concerns Regarding the Sensitivity of the OptiGene Direct SARS-CoV-2 LAMP Assay and Its Suitability for Use in at-Risk Groups and Hospital Staff. J. Infect. 2021, 82, 282–327. [Google Scholar] [CrossRef] [PubMed]
  54. Centers for Disease Control and Prevention. Research Use Only Real-Time RT-PCR Protocol for Identification of 2019-nCoV. 2020. Available online: https://www.cdc.gov/coronavirus/2019-ncov/lab/rt-pcr-detection-instructions.html (accessed on 29 March 2021).
  55. El-Tholoth, M.; Bau, H.H.; Song, Y. A Single and Two-Stage, Closed-Tube, Molecular Test for the 2019 Novel Coronavirus (COVID-19) at Home, Clinic, and Points of Entry. ChemRxiv 2020. [Google Scholar] [CrossRef]
  56. Sicasys Software GmbH; Spotxel® Reader : Germersheim, Germany, 2017.
  57. Medeiros da Silva, R.C.; Nogueira Marinho, L.C.; de Araújo Silva, D.N.; Costa de Lima, K.; Pirih, F.Q.; Luz de Aquino Martins, A.R. Saliva as a Possible Tool for the SARS-CoV-2 Detection: A Review. Travel Med. Infect. Dis. 2020, 38, 101920. [Google Scholar] [CrossRef] [PubMed]
  58. Rabe, B.A.; Cepko, C. SARS-CoV-2 Detection Using Isothermal Amplification and a Rapid, Inexpensive Protocol for Sample Inactivation and Purification. Proc. Nat. Acad. Sci. USA 2020, 117, 24450–24458. [Google Scholar] [CrossRef]
  59. Meyerson, N.R.; Yang, Q.; Clark, S.K.; Paige, C.L.; Fattor, W.T.; Gilchrist, A.R.; Barbachano-Guerrero, A.; Sawyer, S.L. A Community-Deployable SARS-CoV-2 Screening Test Using Raw Saliva with 45 Minutes Sample-to-Results Turnaround. Medrxiv 2020. [Google Scholar] [CrossRef]
  60. Dinnes, J.; Deeks, J.J.; Adriano, A.; Berhane, S.; Davenport, C.; Dittrich, S.; Emperador, D.; Takwoingi, Y.; Cunningham, J.; Beese, S.; et al. Cochrane COVID-19 Diagnostic Test Accuracy Group. Rapid, Point-of-Care Antigen and Molecular-Based Tests for Diagnosis of SARS-CoV-2 Infection. Cochrane Database Syst. Rev. 2020, 8, CD013705. [Google Scholar] [CrossRef] [PubMed]
  61. Jones, T.C.; Biele, G.; Mühlemann, B.; Veith, T.; Schneider, J.; Beheim-Schwarzbach, J.; Bleicker, T.; Tesch, J.; Schmidt, M.L.; Sander, L.E.; et al. Estimating Infectiousness throughout SARS-CoV-2 Infection Course. Science 2021, 373, eabi5273. [Google Scholar] [CrossRef]
  62. van Kampen, J.J.A.; van de Vijver, D.A.M.C.; Fraaij, P.L.A.; Haagmans, B.L.; Lamers, M.M.; Okba, N.; van den Akker, J.P.C.; Endeman, H.; Gommers, D.A.M.P.J.; Cornelissen, J.J.; et al. Duration and Key Determinants of Infectious Virus Shedding in Hospitalized Patients with Coronavirus Disease-2019 (COVID-19). Nat. Commun. 2021, 12, 267. [Google Scholar] [CrossRef]
  63. Kawasuji, H.; Takegoshi, Y.; Kaneda, M.; Ueno, A.; Miyajima, Y.; Kawago, K.; Fukui, Y.; Yoshida, Y.; Kimura, M.; Yamada, H.; et al. Transmissibility of COVID-19 Depends on the Viral Load around Onset in Adult and Symptomatic Patients. PLoS ONE 2020, 15, e0243597. [Google Scholar] [CrossRef]
  64. Juscamayta-López, E.; Valdivia, F.; Horna, H.; Tarazona, D.; Linares, L.; Rojas, N.; Huaringa, M. A Multiplex and Colorimetric Reverse Transcription Loop-Mediated Isothermal Amplification Assay for Sensitive and Rapid Detection of Novel SARS-CoV-2. Front. Cell Infect. Microbiol. 2021, 11, 653616. [Google Scholar] [CrossRef]
  65. Escalante-Maldonado, O.; Vidal-Anzardo, M.; Donaires, F.; Solis-Sanchez, G.; Gallesi, I.; Pampa-Espinoza, L.; Huaringa, M.; Rojas-Serrano, N.; García, C.; Angles-Yanqui, E.; et al. Standardization and Validation of an in House RT-LAMP Molecular Test for the Diagnosis of SARS-CoV-2. Rev. Peru Med. Exp. Salud. Publica 2021, 38, 7–16. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Agreement of sample pre-processing strategies. Percentages in each strategy are values of total agreement, calculated based on results for eight NPSs and four NPAs with Ct from 11–33 plus two negatives of each type of sample (labeled un); for esthetic reasons, only half of the pictures are shown. For all gels: smears were counted as positives, M = 100 pb ladder, numbers 11, 18, 23, 26, 30, 33 = Ct of the sample. For tubes for visual detection: “orangish” was taken as negative, numbers 11, 18, 23, 26, 30, 33 = Ct of the sample. For the application (App): order is for tubes shown for visual detection (Ct 11, 18, 23, 26, 30, 33, un, un), Fc (fold change) = after/before; if Fc > 1.0, the sample was classified as positive.
Figure 1. Agreement of sample pre-processing strategies. Percentages in each strategy are values of total agreement, calculated based on results for eight NPSs and four NPAs with Ct from 11–33 plus two negatives of each type of sample (labeled un); for esthetic reasons, only half of the pictures are shown. For all gels: smears were counted as positives, M = 100 pb ladder, numbers 11, 18, 23, 26, 30, 33 = Ct of the sample. For tubes for visual detection: “orangish” was taken as negative, numbers 11, 18, 23, 26, 30, 33 = Ct of the sample. For the application (App): order is for tubes shown for visual detection (Ct 11, 18, 23, 26, 30, 33, un, un), Fc (fold change) = after/before; if Fc > 1.0, the sample was classified as positive.
Diagnostics 12 00848 g001
Figure 2. Effect of sample quality on visual detection. Categories correspond to: tp = true positives, tn = true negatives, fp = false positives, qRT-PCR negatives that turned yellow in LAMP, fn = false negatives, qRT-PCR positives that gave faint or negative visual outcomes. Percentage of agreement with qRT-PCR is 57.1, 62.4, 75.0, 50.0 and 62.0, for bloody, clear, fuchsia, mucus and precipitated conditions, respectively.
Figure 2. Effect of sample quality on visual detection. Categories correspond to: tp = true positives, tn = true negatives, fp = false positives, qRT-PCR negatives that turned yellow in LAMP, fn = false negatives, qRT-PCR positives that gave faint or negative visual outcomes. Percentage of agreement with qRT-PCR is 57.1, 62.4, 75.0, 50.0 and 62.0, for bloody, clear, fuchsia, mucus and precipitated conditions, respectively.
Diagnostics 12 00848 g002
Figure 3. Relation between fold change and viral load. Y-axis indicates fold change values for a total of 208 samples, with the horizontal line corresponding to the threshold established. Boxes contain average and standard deviation of the fold change per Ct group, where Neg corresponds to the group with undetectable RNA of SARS-CoV-2 by qRT-PCR. The red line marks a fold change = 1.0.
Figure 3. Relation between fold change and viral load. Y-axis indicates fold change values for a total of 208 samples, with the horizontal line corresponding to the threshold established. Boxes contain average and standard deviation of the fold change per Ct group, where Neg corresponds to the group with undetectable RNA of SARS-CoV-2 by qRT-PCR. The red line marks a fold change = 1.0.
Diagnostics 12 00848 g003
Figure 4. Performance of the application-mediated detection among symptomatic and asymptomatic patients. Accuracy = tp + tn/total of symptomatic or asymptomatic in each Ct group; standard deviation was established based in the calculated accuracy per type of sample in each group. The red line marks a 85% of accuracy, required in Colombia for SARS-CoV-2 diagnostic tests.
Figure 4. Performance of the application-mediated detection among symptomatic and asymptomatic patients. Accuracy = tp + tn/total of symptomatic or asymptomatic in each Ct group; standard deviation was established based in the calculated accuracy per type of sample in each group. The red line marks a 85% of accuracy, required in Colombia for SARS-CoV-2 diagnostic tests.
Diagnostics 12 00848 g004
Table 1. Detection limit of the rapid LAMP assay.
Table 1. Detection limit of the rapid LAMP assay.
NPSNPASAL
Estimated No CopiesPositives
Visual
Positives
SpotXel
Estimated No CopiesPositives
Visual
Positives
SpotXel
Estimated No CopiesPositives
Visual
Positives
SpotXel
564,8494/44/4558,8014/44/4582,7304/44/4
48,8104/44/4112,8144/44/488,6823/44/4
22,5944/44/422,5632/43/423,3090/42/4
70011/42/410,8272/42/479061/42/4
6212/42/413813/43/49330/41/4
1811/41/41793/42/41861/42/4
360/42/4361/43/4370/40/4
0.00/40/00.00/40/40.00/40/4
Table 2. Cross-reactivity.
Table 2. Cross-reactivity.
VirusNo. Samples Tested *Positives,
Visual
Positives,
SpotXel
AdV40/40/4
InfA30/30/3
InfB20/20/2
hMPNV30/30/3
Parainf (1–3)60/60/6
RSV40/40/4
HCoV-OC4310/10/1
* Each sample was processed twice in the fast assay, but HCoV-OC43 was also evaluated with purified RNA.
Table 3. Estimated performance of visual detection.
Table 3. Estimated performance of visual detection.
RT-LAMP Visual
PositivesFaintNegativesTotal
qRT-PCR CDCPositives623534131
Faint0000
Negatives4145977
Total664993208
Accuracy *67.8 (61.03–74.1)
Kappa index *0.34 (0.26–0.42)
* Faint results were assigned to false positives or false negatives when calculating.
Table 4. Performance of the assay with application-assisted detection.
Table 4. Performance of the assay with application-assisted detection.
Global
qPCR (+) = 131
qPCR (−) = 77
NPS
qPCR (+) = 61
qPCR (−) = 25
NPA
qPCR (+) = 37
qPCR (−) = 25
Saliva
qPCR (+) = 33
qPCR (−) = 27
Accuracy80.6
(74.6–85.8)
54.8
(43.7–65.6)
69.8
(56.8–80.8)
79.5
(67.1–88.8)
Sensitivity74.8
(66.5–82.0)
42.62
(30.0–55.9)
46.0
(29.5–63.1)
57.6
(39.2–74.5)
Specificity83.1
(72.9–90.7)
60.0
(38.7–78.9)
80.0
(59.3–93.2)
88.9(
70.8–97.7)
Positive predictive value65.51
(53.4–75.9)
31.4
(20.7–44.5)
49.6
(29.4–69.9)
68.9
(42.4–87.0)
Negative predictive value88.5
(84.9–91.3)
70.9
(62.4–78.2)
77.5
(70.8–83.1)
83.0
(76.3–88.1)
Fold change (average ± sd)tp = 1.20 ± 0.16tp = 1.17 ± 0.14tp = 1.22 ± 0.17tp = 1.25 ± 0.15
tn = 0.89 ± 0.07tn = 0.91 ± 0.04tn = 0.91 ± 0.05tn = 0.85 ± 0.09
fp = 1.13 ± 0.07fp = 1.10 ± 0.07fp = 1.01 ± 0.01fp = 1.16 ± 0.05
fn = 0.92 ± 0.06fn = 0.94 ± 0.04fn = 0.90 ± 0.04fn = 0.90 ± 0.08
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Londono-Avendano, M.A.; Libreros, G.; Osorio, L.; Parra, B. A Rapid RT-LAMP Assay for SARS-CoV-2 with Colorimetric Detection Assisted by a Mobile Application. Diagnostics 2022, 12, 848. https://doi.org/10.3390/diagnostics12040848

AMA Style

Londono-Avendano MA, Libreros G, Osorio L, Parra B. A Rapid RT-LAMP Assay for SARS-CoV-2 with Colorimetric Detection Assisted by a Mobile Application. Diagnostics. 2022; 12(4):848. https://doi.org/10.3390/diagnostics12040848

Chicago/Turabian Style

Londono-Avendano, María Aurora, Gerardo Libreros, Lyda Osorio, and Beatriz Parra. 2022. "A Rapid RT-LAMP Assay for SARS-CoV-2 with Colorimetric Detection Assisted by a Mobile Application" Diagnostics 12, no. 4: 848. https://doi.org/10.3390/diagnostics12040848

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop