In Vivo Measurement of Cervical Elasticity on Pregnant Women by Torsional Wave Technique: A Preliminary Study

1
Instituto de Investigación Biosanitaria, ibs.GRANADA, 18012 Granada, Spain
2
San Cecilio University Hospital, 18016 Granada, Spain
3
Department of Structural Mechanics, University of Granada, 18071 Granada, Spain
4
Excellence Research Unit, “Modelling Nature” (MNat), University of Granada, 18071 Granada, Spain
*
Author to whom correspondence should be addressed.
Sensors 2019, 19(15), 3249; https://doi.org/10.3390/s19153249
Submission received: 4 June 2019 / Revised: 18 July 2019 / Accepted: 22 July 2019 / Published: 24 July 2019
(This article belongs to the Special Issue Smart Sensors for Healthcare and Medical Applications)

Abstract

:
A torsional wave (TW) sensor prototype was employed to quantify stiffness of the cervix in pregnant women. A cross-sectional study in a total of 18 women between 16 weeks and 35 weeks + 5 days of gestation was performed. The potential of TW technique to assess cervical ripening was evaluated by the measurement of stiffness related to gestational age and cervical length. Statistically significant correlations were found between cervical stiffness and gestational age ( R 2 = 0.370 , p = 0.0074 , using 1 kHz waves and R 2 = 0.445 , p = 0.0250 , using 1.5 kHz waves). A uniform decrease in stiffness of the cervical tissue was confirmed to happen during the complete gestation. There was no significant correlation between stiffness and cervical length. A stronger association between gestational age and cervical stiffness was found compared to gestational age and cervical length correlation. As a conclusion, TW technique is a feasible approach to objectively quantify the decrease of cervical stiffness related to gestational age. Further research is required to evaluate the application of TW technique in obstetric evaluations, such as prediction of preterm delivery and labor induction failure.

1. Introduction

Approximately, 15 million babies are born preterm (before 37 weeks of gestation) per year, i.e., more than 1 in 10 newborns, and this number is rising in both developing countries and Europe [1,2]. Worldwide, complications of preterm birth are the main cause of child mortality under five years of age [1,3]. Prematurity often leads to long-term disabilities such as learning, visual and hearing problems. Clinical and social risk factors of preterm birth have been identified to develop feasible and cost-effective care measures to save children [1]. Nonetheless, a high proportion of spontaneous preterm birth remains unpredictable.
Current models based on cervical length, obstetric history, digital vaginal examination and echography of the cervix are not able to accurately predict a preterm birth with sufficient anticipation, and there is a lack of evidence on how to prevent preterm delivery [4,5]. Even though there is an agreement that cervical ripening plays a fundamental role during pregnancy, histological changes and biomechanical properties of the cervix are not entirely characterized. The current lack of a clinical tool for the quantitative evaluation of the biomechanic parameters of the cervix is probably a barrier to advance in preventing spontaneous preterm birth [6]. Since 2012, the WHO is encouraging to accelerate research into the causality of preterm birth, and to test effective approaches that would lead to save babies. Recently, elastography techniques are being put forward in the literature to assess quantitatively the stiffness of the cervix as a promising tool to estimate preterm birth risk, as well as to predict the success of labor induction [7,8,9,10,11,12].
Quasi-static elastography methods have been used to evaluate the cervical stiffness at different gestational ages. The result is a qualitative deformation gradient map, called elastogram. However, all methods have shown unclear results regarding reproducibility and associations between stiffness and gestational age [13,14,15,16]. The measurements are also liable to the sensor pressure applied by the clinician, which is not accounted for.
In contrast, dynamic elastography techniques have the strength to provide the absolute quantitative values of stiffness as an objective criterion to evaluate the process of cervical ripening [17,18]. This technology relies on shear ultrasonic waves that travel through the soft tissue. The measurement of the shear wave propagation speed allows characterizing shear stiffness. The commercially available dynamic Supersonic Shear Imaging (SSI) technique employs ultrasonics radiation force to generate shear waves. Shear wave speed in cervix was statistically significant lower in women delivered preterm and inpatients with preterm uterine dynamics compared to women delivered at term [12]. Peralta et al. [19] evaluated SSI elastography to quantify cervical stiffness in real time and its evolution in induced labor in ewes, concluding that stiffness decreases during maturation in induced labor. The Acoustic Radiation Force Imaging (ARFI) is based on displacements generated by an ultrasound beam using the same imaging probe. ARFI has already been proved to evaluate differences in mature versus immature cervical tissue ex vivo [20] and in vivo in pre- and post-labor induction [21]. Both studies agree that shear waves speeds are statistically significantly different in mature versus immature cervical tissue. Viscoelasticity maps of uterine corpus and cervix were assessed thought magnetic resonance elatography in nonpregnant women [22]. Results show a higher elasticity in uterine corpus, and similar viscosity compared with cervix.
The presented measurements data are taken using an alternative dynamic technique: torsional wave (TW) technique [23]. This is based on the propagation of shear waves through the tissue not only in depth but also radially, which makes the technique suitable for applications such as cervical tissue. Axis-symmetric waves allows the precise interrogation of soft tissue mechanical functionality in cylindrical geometries, which are challenged by current elastography approaches in small organs.
This work was aimed at evaluating the reliability and feasibility of TW technique to provide consistent data on the changes of the cervical stiffness during pregnancy. Eighteen singleton-pregnant women were recruited. The hypothesis were: (1) torsional wave technique has the capacity to quantify cervical stiffness defined by its elastic modulus; and (2) stiffness decreases along pregnancy. The second hypothesis stems from the fact that the cervical tissue behavior depends on changes in its multi-scale structure from a mechanical point of view. The cervical stroma microstructure is formed of cross-linked mesh of collagen immersed into viscous proteoglycan [24]. These biochemical compounds exist on different scales whose length is variable by several orders of magnitude. They provide a tractive and compressive strength to the cervical tissue. Despite the fact that cervical architecture changes during ripening [25], these modifications along pregnancy are still not well studied.

2. Materials and Methods

2.1. Design of the Study

A cross-sectional study in healthy pregnant women was performed to assess stiffness modifications in cervix.

2.2. Healthcare Settings

The pre-pilot test study was carried out at San Cecilio University Hospital in Granada. The data were analyzed in the Ultrasonics Laboratory in the University of Granada.

2.3. Ethical Issues

The study met the principles of the Declaration of Helsinki. Approvals of the Ethical Committee in Human Research of University of Granada and Ethical Commission and Health Research of San Cecilio University Hospital in Granada were achieved.

2.4. Subjects

Eighteen healthy women were recruited from their routine medical visits during pregnancy, and TW technique explorations were performed in the Fetal Medicine Unit. The entire population of women in the study had pregnancies without any complication with a median of 26.4 (16 weeks to 35 weeks + 5 days) gestation weeks, and there was no twin pregnancy. A statistical power analysis was designed to estimate the size of the population. A multivariate continuous regression with a power of 80%, estimated significance in a two-tail distribution, and a recommended effect size ES = 0.30, yielded a sample size of 17 subjects. Exclusion criteria were multiple pregnancies, previous cervical surgeries and patients with information relative to malignant changes in the cervical tissue. All women enrolled in the evaluation provided agreement by signing a written consent and reading the information of the patient report.
For the exploration with TW technique, the participants emptied their bladder before the exploration and then were placed in the dorsal lithotomy position. The intravaginal device was allocated in contact with the cervical internal OS (see Figure 1). The measurements of cervical length were obtained by a transvaginal sonography probe, which was directed in the anterior fornix. A sagital view was obtained. Three TW technique and cervical length measurements per women were performed.

2.5. Torsional Wave Technique

Elastography quantification was achieved by the TW probe [26,27,28], which generated waves under safe threshold of energies. The device consisted in three parts: a torsional wave sensor (probe), an electronic system for generating and receiving the signal, and an interface software (Figure 2).
The probe was manufactured in 2017 and was composed of: (1) an electromechanical actuator which deleted electronic cross-talk [23]; (2) a receiver based on two polylactic acid rings where the piezoelectric elements were fitted; and (3) a case to contain the emitter and the receiver. The shear modulus was obtained assuming an elastic and incompressible medium by the following equation,
μ = ρ c s 2
where ρ is the density of the medium and c s is the torsional wave velocity, which is based on shear wave group velocity.
The excitation signal was a burst composed of a one-cycle frequency f ranging from 0.5 to 1.5 kHz with 10× averaging. The frequencies were chosen according to the results obtained in the work carried out by Callejas et al. [23].
An example of three different emitted and received signals is shown in Figure 3. The shear wave group velocity calculation algorithm was based on dividing the distance by the torsional wave time-of-flight. The signals were preprocessed by a low-pass filter close to the central frequency of the received signal. The time of flight was computed using three procedures: (1) searching the first time the signal raises 30% above zero; (2) subtracting a quarter of the period (inverse of the received signal central frequency) from the first peak; and (3) subtracting three quarters of the period (inverse of the received signal central frequency) from the second peak. All three methods provided similar estimates of the velocity, as shown in the results.

Safety Considerations

A new medical diagnostic equipment needs to follow the specifications described in the Food and Drug Administration (FDA) guidelines [29] for the application in clinical practice. It is necessary that the Torsional Wave technique be safe for humans. There are three parameters that should be evaluated according to the acoustic output in the use of Fetal Imaging and Other (FDA): the mechanical index ( M I < 1.9 ), the spatial peak pulse average intensity ( I S P P A < 190 W / cm 2 ), and the spatial peak temporal average intensity ( I S P T A < 94 mW / cm 2 ). The calculation of these parameters was made as follows:
M I = P R P / F c
where PRP is the peak rarefractional pressure of the torsional wave in (MPa) and Fc is the center frequency (MHz).
I S P P A = P 0 2 / ( 2 * ρ * c )
where P 0 is the maximal acoustic pressure generated by the electromechanical actuator, ρ is the density of the medium, and c is the sound speed in the medium.
I S P T A = I S P P A * Δ t / 1
where Δ t is the excitation pulse duration.
The three parameters were experimentally estimated. The excitation signal used was a low-frequency ultrasonic sine-burst at a central frequency of 1 kHz, consisting of one cycle of 1 ms and 16 Vpp amplitude. This excitation signal was generated by a wave generator (Agilent 33220A, Santa Clara, CA, USA). The response signal was registered using a decibel sensor (YH-610 Environment Multimeter). The signal traveled through a water layer before arriving to the decibel sensor and different distances from 5 cm to 0 cm. To convert the pressure recorded by the decibel sensor into water acoustic pressure, the equation that relates the impedances of the two media (air–water) was used:
T = 2 * Z a i r ( Z a i r + Z w a t e r ) 2
where T is the transmission coefficient and Z a i r and Z w a t e r are the acoustic impedance of the air and water, respectively.

2.6. Statistic Analysis

The evolution of cervical stiffness tissue during pregnancy was quantified. Normal distribution of the data was checked for each velocity calculation algorithm by the normal quantile–quantile plot (Q–Q plot) and the Shapiro–Wilk test. The mean values for each velocity calculation procedure were compared to the normal distribution of these values. The coefficient of determination ( R 2 ) for linear regression analysis was calculated to provide the correlations: (a) between gestational age and cervical velocity ( c s ) using 0.5, 1 and 1.5 kHz torsional waves, for the three velocity calculation algorithms; (b) between gestational age and cervical length; and (c) between stiffness and cervical length. Data were analyzed using the MATLAB (Release 2014b Mathworks, Natick, MA, USA). T-test was calculated to estimate p-values. A statistically significance for p < 0.05 was assumed.

3. Results

The experimental results obtained to evaluate the three security parameters according to the Food and Drug Administration guidelines were as follows.
The maximum pressure registered after converting the pressure recorded by the decibel sensor into water acoustic pressure was 3.99 × 10 5 MPa. The maximal acoustic pressure and the peak rarefractional pressure of the torsional wave in water was P 0 = 3.99 × 10 4 bars. The three previous parameters were obtained with the cited experimental conditions:
M I = 0.0013 < 1.9
I S P P A = P 0 2 / ( 2 * ρ * c ) = 5.3 W / cm 2 < 190 W / cm 2
considering the density of the medium ρ = 1000 kg / m 3 , and the sound speed in the medium 1500 m/s.
I S P T A = I S P P A * Δ t / 1 = 5.3 mW / cm 2 < 94 mW / cm 2
A normal distribution for the three velocity calculation procedures was found through Q–Q test (Figure 4) and Shapiro–Wilk test. The obstetric characteristics of the population in the study are shown in Table 1.
Three measurements of TW stiffness and cervical length per subject were determined from all women. Box plots of these data observed in each patient were calculated and a linear regression was fitted with 80% confidence intervals (see Figure 5). In this work, the three frequencies were used to study the effect of attenuation on the cervical tissue. The selected frequency configuration was 1 kHz, which was the optimal measure to yield the highest amplitude signals, the best shear wave speed reconstructions and a significant correlation with gestational age. In some measurements, frequencies 1.5 kHz, yielded amplitudes of signal similar to the amplitude of noise probably due to attenuation, and consequently anomalous values of the cervical stiffness were obtained. In contrast, the noise masked the amplitude of the signal in some data with frequencies 0.5 kHz. All missing data were due to signal noise.
The decrease of the stiffness was computed from the data in Figure 5 using Equation (1). The error bars are estimated from the three velocity estimation algorithms described in the methods. The three overlapping regressions (continuous and dashed lines in Figure 5, Figure 6 and Figure 7) correspond to each velocity estimation algorithms.
Similar correlations are shown in Figure 6 and Figure 7 for 1.5 kHz and 0.5 kHz, respectively, where some of the measurements were rejected due to noise in the signal. A stronger association between gestational age and cervical stiffness was found ( R 2 = 0.370, p = 0.0074 , Figure 5) compared to gestational age and cervical length correlation ( R 2 = 0.025, p = 0.6043 , Figure 8).
No high associations ( R 2 < 0.5 for all cases) and no significant correlation (p > 0.05) were obtained between stiffness and cervical length (Figure 9).

4. Discussion

This study focused on assessing the feasibility of torsional wave technique to quantify the changes in cervical stiffness during pregnancy, which were measured by shear wave speed. The presented results show, for the first time in vivo, the viability of torsional waves to objectively measure cervical elasticity in pregnant women. The observed data therefore support Hypothesis 1 that torsional wave technique has the capacity to quantify cervical stiffness defined by its elastic modulus.
The presented observations also support Hypothesis 2 that shear stiffness decreases during pregnancy. Cervical stiffness was shown to significantly decrease with gestational age, which is compatible with observations by former researchers that assessed cervical ripening by different techniques [8,30,31,32]. A gradual reduction from about 40 kPa at the beginning of pregnancy to close to zero at delivery was obtained in the study carried out by Peralta et al. [30]. A correction due to the difference of range of shear wave frequencies of ARFI was considered, about a higher order of magnitude, which affect the apparent stiffness given the viscoelastic behaviour of cervical tissue. Thus, cervical ripening is directly related to the time to delivery. Correlation between cervical stiffness and gestational age assessed by TW technique showed a higher correlation to gestational aged compared to quantification through shear wave speed (SSI) ( R 2 = 0.37 vs. R 2 = 0.29 ) [12].
A weaker correlation was found between cervical stiffness and cervical length than with gestational age, which is compatible with previous studies [11,30], using dynamic and quasi-static elastography, respectively, but contrary to observations by Hernandez-Andrade et al. [16], who found that associations between cervical tissue strain and cervical length was higher than with gestational age. This inconsistency feeds a debate, which could be at least partially explained by the inherent limitations of the commercially available quasi-static elastography technologies [14,15,16,17], as this technique provides a qualitative estimation of the cervical stiffness through an indirect measurement.
The experiment results support that TW technique is safe to be used in pregnant women. All the values obtained were far below the thresholds according to the Food and Drug Administration (FDA) guidelines reference parameters in Fetal Imaging and Other. The mechanical index ( M I ) was 0.0013 (<1.9), the spatial peak pulse average intensity ( I S P P A ) was 5.3 W / cm 2 (< 190 W / cm 2 ), and the spatial peak temporal average intensity ( I S P T A ) was 5.3 mW / cm 2 (< 94 mW / cm 2 ).
The limitations of this research are linked to the nature of propagation of torsional wave in cervical tissue as well as its complex microarchitecture. Some mechanical hypotheses have been raised in the literature about the hystologic features of the cervix to estimate the shear stiffness elasticity, assuming homogeneous, non viscous, isotropic and semi-infinite medium [20,33,34,35,36]. The equation employed in this study to estimate the cervix stiffness is only based on shear wave group velocity. However, the behavior of cervical tissue is dispersive, that is, the higher are the shear wave frequencies, the higher are the shear waves speeds and, therefore, phase–velocity-based techniques would lead to a direct calculation of shear modulus. The time-of-tflight technique measured the shear wave group velocity, which is dependent on the envelope of the propagating elastic wave.
Finally, due to the exploratory nature of this study about the feasibility of torsional wave technique to assess cervical maturation, a small population of patients was recruited. To extend the validity and reliability of the proposed technology, larger complementary studies are needed. The protocol of measurements by TW technique will be enhanced by applying the optimal contact conditions between the probe and the cervix [37]. We are positive that torsional waves are a tool with potential to objectively diagnose early cervical ripening disorders and preterm birth.

5. Conclusions

The presented experimental observations prove that, firstly, cervical stiffness was a valuable predictor variable of gestational age at the moment of evaluation. Secondly, TW technique is a tool that allows quantifying cervical shear stiffness during pregnancy. Finally, this technique is safe to be used in pregnant women.
TW technique might provide clinically relevant data on the cervical ripening in addition to that obtained from digital exploration and standard sonography. Further research is required to assess the TW technique feasibility in obstetric evaluations, such as probabilistic inverse problems based on viscoelastic models for the prediction of preterm delivery and labor induction failure.

Author Contributions

P.M., A.C., J.M. and G.R. conceived and designed the study. P.M., F.S.M. and G.R. performed the measurements. P.M., A.C., J.M. and G.R. analyzed the data and wrote the paper.

Funding

This research was funded by the Ministry of Education grant numbers DPI2017-83859-R, DPI2014-51870-R and UNGR15-CE-3664, Ministry of Health grant numbers DTS15/00093 and PI16/00339, and Junta de Andalucía grant numbers, PI-0107-2017 and PIN-0030-2017.

Acknowledgments

We thank the Department of Electronics and Computer Technology for assistance with the electronic system of the torsional wave sensor.

Conflicts of Interest

Massó has nothing to disclose. Molina, Rus, Melchor and Callejas are partners of INNITIUS, a company to develop ultrasonic sensors; they have not receive funding from the company for the submitted work.

Abbreviations

The following abbreviations are used in this manuscript:
TWTorsional Wave
WHOWorld Health Organization
ARFIAcoustic Radiation Force Imaging
SSISupersonic Shear Imaging
PLAPolylactic Acid
Internal OSInternal Orifice of the Cervix
ESEffect size

References

  1. WHO. March of Dimes, The Partnership for Maternal, Newborn & Child Health, and Save the Children 2012; Born too Soon: The global Action Report on Preterm, Birth; Magn. Reson. Med.; Howson, C.P., Kinney, M.V., Lawn, J.E., Eds.; WHO: Geneva, Switzerland, 2012. [Google Scholar]
  2. Zeitlin, J.; Szamotulska, K.; Drewniak, N.; Mohangoo, A.D.; Chalmers, J.; Sakkeus, L.; Irgens, L.; Gatt, M.; Gissler, M.; Blondel, B.; et al. Wideband MR elastography for viscoelasticity model identification. BJOG Int. J. Obstet. Gynecol. 2013, 120, 1356–1365. [Google Scholar] [CrossRef] [PubMed]
  3. WHO. Fact Sheet—Preterm Birth; Magn. Reson. Med.; WHO: Geneva, Switzerland, 2016. [Google Scholar]
  4. Sananes, N.; Langer, B.; Gaudineau, A.; Kutnahorsky, R.; Aissi, G.; Fritz, G.; Boudier, E.; Viville, B.; Nisand, I.; Favre, R. Prediction of spontaneous preterm delivery in singleton pregnancies: where are we and where are we going? A review of literature. J. Obstet. Gynaecol. 2014, 34, 457–461. [Google Scholar] [CrossRef] [PubMed]
  5. Ross, M.G.; Beall, M.H. Prediction of preterm birth: Nonsonographic cervical methods. Semin. Perinatol. 2009, 33, 312–316. [Google Scholar] [CrossRef] [PubMed]
  6. Feltovich, H.; Hall, T.J.; Berghella, V. Beyond cervical length: Emerging technologies for assessing the pregnant cervix. Am. J. Obstet. Gynecol. 2012, 207, 345–354. [Google Scholar] [CrossRef]
  7. Khalil, M.R.; Thorsen, P.; Uldbjerg, N. Cervical ultrasound elastography may hold potential to predict risk of preterm birth. Ultrasound Obstet. Gynecol. 2013, 60, A4570. [Google Scholar]
  8. Kobbing, K.; Fruscalzo, A.; Hammer, K.; Mollers, M.; Falkenberg, M.; Kwiecien, R.; Klockenbusch, W.; Schmitz, R. Quantitative elastography of the uterine cervix as a predictor of preterm delivery. J. Perinatol. 2014, 34, 774. [Google Scholar] [CrossRef]
  9. Swiatkowska-Freund, M.; Preis, K. Elastography of the uterine cervix: Implications for success of induction of labor. Ultrasound Obstet. Gynecol. 2011, 38, 52–56. [Google Scholar] [CrossRef]
  10. Pereira, S.; Frick, A.P.; Poon, L.C.; Zamprakou, A.; Nicolaides, K.H. Successful induction of labor: Prediction by pre-induction cervical length, angle of progression and cervical elastography. Ultrasound Obstet. Gynecol. 2014, 44, 468–475. [Google Scholar] [CrossRef]
  11. Wozniak, S.; Czuczwar, P.; Szkodziak, P.; Milart, P.; Wozniakowska, E.; Paszkowski, T. Elastography in predicting preterm delivery in asymptomatic, low-risk women: A prospective observational study. BMC Pregnancy Childbirth 2014, 14, 238. [Google Scholar] [CrossRef]
  12. Muller, M.; Aït-Belkacem, D.; Hessabi, M.; Gennisson, J.L.; Grangé, G.; Goffinet, F.; Lecarpentier, E.; Cabrol, D.; Tanter, M.; Tsatsaris, V. Assessment of the cervix in pregnant women using shear wave elastography: A feasibility study. Ultrasound Med. Biol. 2015, 41, 2789–2797. [Google Scholar] [CrossRef]
  13. Thomas, A.; Kummel, S.; Gemeinhardt, O.; Fischer, T. Real-time sonoelastography of the cervix: Tissue elasticity of the normal and abnormal cervix. Acad. Radiol. 2007, 14, 193–200. [Google Scholar] [CrossRef] [PubMed]
  14. Molina, F.S.; Gómez, L.F.; Florido, J.; Padilla, M.C.; Nicolaides, K.H. Quantification of cervical elastography: A reproducibility study. Ultrasound Obstet. Gynecol. 2012, 39, 685–689. [Google Scholar] [CrossRef] [PubMed]
  15. Fruscalzo, A.; Schmitz, R.; Klockenbusch, W.; Steinhard, J. Reliability of cervix elastography in the late first and second trimester of pregnancy. Ultraschall Med. Eur. J. Ultrasound 2012, 33, E101–E107. [Google Scholar] [CrossRef] [PubMed]
  16. Hernandez-Andrade, E.; Hassan, S.S.; Ahn, H.; Korzeniewski, S.J.; Yeo, L.; Chaiworapongsa, T.; Romero, R. Evaluation of cervical stiffness during pregnancy using semiquantitative ultrasound elastography. Ultrasound Obstet. Gynecol. 2013, 41, 152–161. [Google Scholar] [CrossRef] [Green Version]
  17. Feltovich, H.; Hall, T.J. Quantitative imaging of the cervix: Setting the bar. Ultrasound Obstet. Gynecol. 2013, 41, 121–128. [Google Scholar] [CrossRef] [PubMed]
  18. Bamber, J.; Cosgrove, D.; Dietrich, C.F.; Fromageau, J.; Bojunga, J.; Calliada, F. EFSUMB guidelines and recommendations on the clinical use of ultrasound elastography. Part I: Basic principles and technology. Ultraschall Med. 2013, 34, 169–184. [Google Scholar] [CrossRef] [PubMed]
  19. Peralta, L.; Mourier, E.; Richard, C.; Chavette-Palmer, P.; Muller, M.; Tanter, M.; Rus, G. 117 in vivo evaluation of the cervical stiffness evolution during indiced labor in ewes using elastohraphy. Reprod. Fertil. Dev. 2015, 27, 150–151. [Google Scholar] [CrossRef]
  20. Carlson, L.C.; Feltovich, H.; Palmeri, M.L.; Dahl, J.J.; Muñoz del Rio, A.; Hall, T.J. Estimation of shear wave speed in the human uterine cervix. Ultrasound Obstet. Gynecol. 2014, 43, 452–458. [Google Scholar] [CrossRef]
  21. Carlson, L.C.; Romero, S.T.; Palmeri, M.L.; Muñoz del Rio, A.; Esplin, S.M.; Rotemberg, V.M.; Feltovich, H. Changes in shear wave speed pre and post induction of labor: A feasibility study. Ultrasound Obstet. Gynecol. 2014. [Google Scholar] [CrossRef]
  22. Jiang, X.; Asbach, P.; Streitberger, K.J.; Thomas, A.; Hamm, B.; Braun, J.; Sack, I.; Guo, J. In vivo high-resolution magnetic resonance elastography of the uterine corpus and cervix. Eur. Radiol. 2014, 24, 3025–3033. [Google Scholar] [CrossRef]
  23. Callejas, A.; Gomez, A.; Melchor, J.; Riveiro, M.; Massó, P.; Torres, J.; López-López, M.; Rus, G. Performance study of a torsional wave sensor and cervical tissue characterization. Sensors 2017, 17, 2078. [Google Scholar] [CrossRef] [PubMed]
  24. House, M.; Kaplan, D.L.; Socrate, S. Relationships between mechanical properties and extracellular matrix constituents of the cervical stroma during pregnancy. Semin. Perinatol. 2009, 33, 300–307. [Google Scholar] [CrossRef] [PubMed]
  25. Timmons, B.; Akins, M.; Mahendroo, M. Cervical remodeling during pregnancy and parturition. Trends Endocrinol. Metab. 2010, 21, 353–361. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  26. Melchor, J.; Rus, G. Torsional ultrasonic transducer computational design optimization. Ultrasonics 2014, 54, 1950–1962. [Google Scholar] [CrossRef] [PubMed]
  27. Melchor, J.; Muñoz, R.; Rus, G. Torsional ultrasound sensor optimization for soft tissue characterization. Sensors 2017, 17, 1402. [Google Scholar] [CrossRef] [PubMed]
  28. Rus, G.; Muñoz, R.; Melchor, J.; Molina, R.; Callejas, A.; Riveiro, M.; Massó, P.; Torres, J.; Moreu, G.; Molina, F.; et al. Torsion ultrasonic sensor for tissue mechanical characterization. In Proceedings of the 2016 IEEE International Ultrasonics Symposium (IUS), Tours, France, 18–21 September 2016; pp. 1–4. [Google Scholar]
  29. Food and Drug Administration. Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducer; U.S. Department Health and Human Services, Food and Drug Administration, Center for Devices and Radiological Health: Washington, DC, USA, 1997.
  30. Peralta, L.; Molina, F.; Melchor, J.; Gómez, L.; Massó, P.; Florido, J.; Rus, G. Transient elastography to assess the cervical ripening during pregnancy: A preliminary study. Ultraschall Med. Eur. J. Ultrasound 2015. [Google Scholar] [CrossRef] [PubMed]
  31. Badir, S.; Mazza, E.; Zimmermann, R.; Bajka, M. Cervical softening occurs early in pregnancy: Characterization of cervical stiffness in 100 healthy women using the aspiration technique. Prenat. Diagn. 2013, 33, 737–741. [Google Scholar] [CrossRef]
  32. Parra-Saavedra, M.; Gomez, L.; Barrero, A.; Parra, G.; Vergara, F.; Navarro, E. Prediction of preterm birth using the cervical consistency index. Ultrasound Obstet. Gynecol. 2011, 38, 44–51. [Google Scholar] [CrossRef]
  33. Gennisson, J.L.; Deffieux, T.; Fink, M.; M, T. Ultrasound elastography: Principles and techniques. Diagn. Interv. Imaging 2013, 94, 487–495. [Google Scholar] [CrossRef]
  34. Peralta, L.; Rus, G.; Bochud, N.; Molina, F.S. Mechanical assessment of cervical remodelling in pregnancy: Insight from a synthetic model. J. Biomech. 2015, 144, 1557–1565. [Google Scholar] [CrossRef]
  35. Peralta, L.; Rus, G.; Bochud, N.; Molina, F.S. Assessing viscoelasticity of shear wave propagation in cervical tissue by multiscale computational simulation. J. Biomech. 2015, 48, 1549–1556. [Google Scholar] [CrossRef] [PubMed]
  36. Yasar, T.K.; Royston, T.J.; Magin, R.L. Wideband mr elastography for viscoelasticity model identification. Magn. Reson. Med. 2013, 70, 479–489. [Google Scholar] [CrossRef] [PubMed]
  37. Rus, G.; Riveiro, M.; Molina, F. Effect of Contact Conditions of Torsional Wave Elastographic Probe on Human Cervix. Math. Probl. Eng. 2018, 2018, 6494758. [Google Scholar] [CrossRef]
Figure 1. Schematic diagram for the exploration with TW technique.
Figure 1. Schematic diagram for the exploration with TW technique.
Sensors 19 03249 g001
Figure 2. The prototyped TW probe.
Figure 2. The prototyped TW probe.
Sensors 19 03249 g002
Figure 3. Example of three emitted and received 1 kHz signals.
Figure 3. Example of three emitted and received 1 kHz signals.
Sensors 19 03249 g003
Figure 4. Normal quantile–quantile plots for the three velocity calculation procedures.
Figure 4. Normal quantile–quantile plots for the three velocity calculation procedures.
Sensors 19 03249 g004aSensors 19 03249 g004b
Figure 5. Relationship between cervical stiffness assessed by shear wave speed using 1 kHz waves and gestational age at time of examination.
Figure 5. Relationship between cervical stiffness assessed by shear wave speed using 1 kHz waves and gestational age at time of examination.
Sensors 19 03249 g005
Figure 6. Relationship between cervical stiffness assessed by shear wave speed using 1.5 kHz waves and gestational age at time of examination.
Figure 6. Relationship between cervical stiffness assessed by shear wave speed using 1.5 kHz waves and gestational age at time of examination.
Sensors 19 03249 g006
Figure 7. Relationship between cervical stiffness assessed by shear wave speed using 0.5 kHz waves and gestational age at time of examination.
Figure 7. Relationship between cervical stiffness assessed by shear wave speed using 0.5 kHz waves and gestational age at time of examination.
Sensors 19 03249 g007
Figure 8. Relationship between cervical length and gestational age at time of examination.
Figure 8. Relationship between cervical length and gestational age at time of examination.
Sensors 19 03249 g008
Figure 9. Relationship between cervical stiffness and cervical length.
Figure 9. Relationship between cervical stiffness and cervical length.
Sensors 19 03249 g009
Table 1. Features of the population in the study.
Table 1. Features of the population in the study.
CharacteristicsValue
Total population (N)18
Gestational age at test (weeks)26.4 (16 weeks to 35 weeks + 5 days)
Nulliparous (N)2 (11 %)
Cervical length (mm)33 (10–49)

Share and Cite

MDPI and ACS Style

Massó, P.; Callejas, A.; Melchor, J.; Molina, F.S.; Rus, G. In Vivo Measurement of Cervical Elasticity on Pregnant Women by Torsional Wave Technique: A Preliminary Study. Sensors 2019, 19, 3249. https://doi.org/10.3390/s19153249

AMA Style

Massó P, Callejas A, Melchor J, Molina FS, Rus G. In Vivo Measurement of Cervical Elasticity on Pregnant Women by Torsional Wave Technique: A Preliminary Study. Sensors. 2019; 19(15):3249. https://doi.org/10.3390/s19153249

Chicago/Turabian Style

Massó, Paloma, Antonio Callejas, Juan Melchor, Francisca S. Molina, and Guillermo Rus. 2019. "In Vivo Measurement of Cervical Elasticity on Pregnant Women by Torsional Wave Technique: A Preliminary Study" Sensors 19, no. 15: 3249. https://doi.org/10.3390/s19153249

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop