Iatrogenic coronary artery dissection induced during invasive absolute coronary blood flow measurement: optical coherence tomography findings

Marcos Garcia-Guimaraes, MD; Fernando Rivero, MD; Javier Cuesta, MD; Fernando Alfonso*, MD, PhD

* denotes wire artefact

A 65-year-old woman with a previous history of hypertension was referred to our institution due to effort angina with a positive exercise stress test. Coronary angiography showed normal coronary arteries (Panel A). A vasospasm provocation test with methyl-ergonovine was negative. With the aim of assessing the state of the microvasculature, absolute coronary blood flow measurement by thermodilution was performed as previously described1-3. First, a pressure wire (Certus™; St. Jude Medical, St. Paul, MN, USA) was progressed across the mid segment of the left anterior descending coronary artery (LAD) without any difficulties. Then, a 2.52 Fr multipurpose infusion catheter (RayFlow®; Hexacath, Paris, France) was advanced to the proximal-mid LAD (Panel B). Finally, continuous infusion of saline was initiated and absolute coronary blood flow was calculated. Unexpectedly, a clear angiographic haziness, with lumen narrowing, was detected at this site in the control angiogram (Panel C, white arrows). To clarify this image better, optical coherence tomography (OCT) was performed, revealing striking images, highly suggestive of localised but severe vasospasm. These images showed: 1) a thickened media and corrugated intima layer (yellow arrows) associated with lumen narrowing at the distal segment (Panel E), 2) a middle region depicting an intramural haematoma (IH) (Panel F), and 3) an entry door (ED) of a confined intimomedial dissection at the most proximal segment (Panel G). After intracoronary nitroglycerine administration, lumen narrowing readily resolved, with only mild lumen haziness persisting at the mid LAD. To elucidate a potential reason for this unexpected complication, the saline infusion pump was reviewed. This disclosed that, due to a technical mistake, the infusion flow rate was programmed at 120 ml/min (recommended rate ≤30 ml/min) with a maximum pressure limit of 600 pounds-force per square inch. Although the patient remained asymptomatic, she was eventually admitted for clinical observation. A control coronary angiography at 48 hours confirmed complete resolution of the angiographic abnormalities (Panel D). OCT at the most distal portion of the LAD showed an almost normal aspect (Panel H). However, at more proximal segments, OCT disclosed a very small residual IH with normal vessel lumen (Panel I, Panel J, white arrowheads). Cardiac biomarkers were negative and the patient was discharged uneventfully. At three-month follow-up she remains asymptomatic.

This case nicely illustrates that, when new invasive methods are introduced for the first time in the catheterisation laboratory, major attention to subtle details remains crucial in order to prevent potential complications. In our patient, high-pressure jets emerging from the side holes of the infusion catheter (as a result of the erroneous flow programming) probably induced an iatrogenic intimal tear and IH leading to severe associated vasospasm that, fortunately, resolved with conservative management. Furthermore, in this case OCT was instrumental in guiding the clinical decision-making process in our patient.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Volume 13 Number 3
Jun 20, 2017
Volume 13 Number 3
View full issue


Key metrics

On the same subject

State-of-the-Art

10.4244/EIJ-D-23-00448 Jan 15, 2024
Coronary spasm and vasomotor dysfunction as a cause of MINOCA
Yaker ZS et al
free

Clinical Research

10.4244/EIJ-D-21-00875 Aug 5, 2022
Features of atherosclerosis in patients with angina and no obstructive coronary artery disease
Pellegrini D et al
free

Clinical Research

10.4244/EIJ-D-21-00402 Feb 18, 2022
Principles and pitfalls in coronary vasomotor function testing
Feenstra R et al
free

Short report

10.4244/EIJ-D-20-00074 Jun 25, 2021
Safety of absolute coronary flow and microvascular resistance measurements by thermodilution
Keulards D et al
free
Trending articles
313.48

State-of-the-Art Review

10.4244/EIJ-D-21-00695 Nov 19, 2021
Transcatheter treatment for tricuspid valve disease
Praz F et al
free
42.95

Clinical research

10.4244/EIJ-D-23-00282 Jul 17, 2023
Early safety and feasibility of a first-in-class biomimetic transcatheter aortic valve - DurAVR
Kodali SK et al
30.35

EXPERT REVIEW

10.4244/EIJ-D-17-00437 Feb 20, 2018
Catheter-based therapies in acute pulmonary embolism
Schultz J et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved