INTERVENTIONAL FLASHLIGHT

DOI: 10.4244/EIJ-D-17-00192

Long-term outcome of a spontaneous coronary artery dissection treated with a bioresorbable scaffold

Nieves Gonzalo*, MD, PhD; Fernando Macaya, MD; José J. Gómez de Diego, MD; Javier Escaned, MD, PhD

A 34-year-old lady with no cardiovascular risk factors presented with an anterior non-ST-elevation myocardial infarction. Angiography showed a severe stenosis in the mid-LAD with TIMI 1 flow and no images of double lumen (Panel A, Moving image 1). To clarify the underlying pathology, optical coherence tomography (Panel B, Moving image 2) was performed demonstrating the presence of an intramural haematoma compressing the intima with critical luminal narrowing (Panel C, asterisk), and more proximally an apparently normal artery (Panel D, Panel E, Dx: diagonal branch). No “entry door” was visible by OCT. Given the compromised flow, the LAD was treated with a bioresorbable scaffold (BRS) (Absorb™; Abbott Vascular, Santa Clara, CA, USA) (3×18 mm). Coronary flow was restored with a good result in the scaffolded segment (Panel F-Panel H). A new moderate stenosis was generated (arrow in Panel F) by proximal displacement of the haematoma (* in Panel G, Panel I, Panel J, Moving image 3, Moving image 4). The flow in the LAD was TIMI 3 and the patient was asymptomatic so no additional interventions were performed in order to avoid further propagation of the dissection.

At follow-up, the patient remained symptom-free taking dual antiplatelet therapy for one year. Forty months post procedure she complained of atypical chest pain. A coronary CT scan was performed showing a normal LAD with no trace of the BRS except for the two radiopaque markers (Panel K-Panel O, arrowheads showing proximal and distal markers of the scaffold) and complete resolution of the proximal intramural haematoma that was initially left untreated (Panel K, Panel L, Panel P, Panel Q, arrows indicating diagonal and septal).

Spontaneous coronary artery dissection is usually managed conservatively given its natural history of spontaneous healing1. In cases where coronary flow is compromised, BRS are an interesting option to restore flow acutely without the disadvantages of a permanent metallic prosthesis2. However, no evidence about the long-term result of BRS in this setting has been reported until now . The present case illustrates a good anatomical result demonstrated with CT scan more than three years after BRS implantation in a young woman with spontaneous coronary artery dissection.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Supplementary data

Moving image 1. Baseline angiogram showing a severe stenosis in the mid-LAD with TIMI 1 flow.

Moving image 2. Baseline OCT showing intramural haematoma compressing the intima with critical luminal narrowing.

Moving image 3. Angiogram after scaffold implantation: good result in the scaffolded segment with a new moderate stenosis proximally.

Moving image 4. OCT after scaffold implantation: good result in the scaffolded segment and a new moderate stenosis by proximal propagation of the haematoma until the diagonal branch.

Supplementary data

To read the full content of this article, please download the PDF.

Moving image 1. Baseline angiogram showing a severe stenosis in the mid-LAD with TIMI 1 flow.

Moving image 2. Baseline OCT showing intramural haematoma compressing the intima with critical luminal narrowing.

Moving image 3. Angiogram after scaffold implantation: good result in the scaffolded segment with a new moderate stenosis proximally.

Moving image 4. OCT after scaffold implantation: good result in the scaffolded segment and a new moderate stenosis by proximal propagation of the haematoma until the diagonal branch.

Volume 13 Number 8
Oct 20, 2017
Volume 13 Number 8
View full issue


Key metrics

On the same subject

Image – Interventional flashlight

10.4244/EIJ-D-18-00709 Jan 18, 2019
The fate of spontaneous coronary artery dissection: insight from intravascular imaging at a late follow-up
Amabile N et al
free

SHORT REPORT

10.4244/EIJ-D-18-00519 Jan 18, 2019
Long-term follow-up of spontaneous coronary artery dissection treated with bioresorbable scaffolds
Macaya F et al
free

INTERVENTIONAL FLASHLIGHT

10.4244/EIJ-D-18-00035 Oct 12, 2018
Hybrid percutaneous treatment of iatrogenic coronary artery dissection complicating a spontaneous coronary artery dissection
García-Guimaraes M et al
free

Image – Interventional flashlight

10.4244/EIJ-D-22-01055 Jul 17, 2023
Aspiration technique in occlusive spontaneous coronary artery dissection
Spînu R et al
Trending articles
313.48

State-of-the-Art Review

10.4244/EIJ-D-21-00695 Nov 19, 2021
Transcatheter treatment for tricuspid valve disease
Praz F et al
free
42.95

Clinical research

10.4244/EIJ-D-23-00282 Jul 17, 2023
Early safety and feasibility of a first-in-class biomimetic transcatheter aortic valve - DurAVR
Kodali SK et al
30.35

EXPERT REVIEW

10.4244/EIJ-D-17-00437 Feb 20, 2018
Catheter-based therapies in acute pulmonary embolism
Schultz J et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 6.2
2022 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2023)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2024 Europa Group - All rights reserved