分光研究
Online ISSN : 1884-6785
Print ISSN : 0038-7002
ISSN-L : 0038-7002
X線分析の精度と有効温度因子
稻垣 道夫古橋 博中 重治
著者情報
ジャーナル フリー

1972 年 21 巻 3 号 p. 202-206

詳細
抄録

The importance of the effective temperature factor in X-ray analysis was shown by presenting the results of measurement of the effective Debye parameter Be f f on some crystalline samples.Calcium fluoride precipitated from Ca (OH) 2 by HF had three times larger value of Be f f than the commercial reagent grade one.This value corresponded to the value after grinding of reagent grade CaF2 for more than 20 hrs.Rutile formed by the phase transformation from anatase with very fine grain was found to have relatively large value of Be f f at the beginning of transformation.It was shown that very large error in the quantity of the crystalline phase might be introduced from the neglect of the effective temperature factor if intensity of the diffraction line at high angle side was used.It was also shown to obtain more reliable results on the ion distribution in spinels (GeCo2O4 and CoAl2O4) by finding the one gave the best linearity between log (Iobs/Icalc) and (sin 2θ/λ2).By this revised method, the effect of the difference in the effective temperature factor from specimen to specimen on the conclusion can be neglected.