Efficient Data Acquisition Technique for Clinical Application of Multileaf Collimator

다엽콜리메이터의 임상적용을 위한 효율적인 정보 취득 기술

  • 이재승 (한동대학교 선린병원 방사선종양학과) ;
  • 김정남 (경희의료원 방사선종양학과)
  • Published : 2008.11.28

Abstract

The MLC(multi leaf collimator) in charge of important role in radiation therapy field recently have been exchanging from shielding block into it rapidly, owing to being convenient. However, MLC can be occurred the leakage dose of inter_leaves and the error of algorithm in imput and output from digital signal. We compared the difference of imput method to MLC made by Varian Cop. with the error and effective field induced by MLC shaper and film scanner based on XimaVision value as using MLC layer of various shapes. According to comparing standard value with them to basic MLC layer (test1-5), MLC shaper was $0{\sim}0.29cm$, $0.23{\sim}3.59cm^2$ and film scanner was $0{\sim}0.78cm$, $0.24{\sim}3.89cm^2$. At the MLC layer to be applied in clinic, MLC shaper was $0{\sim}0.54cm$, $0.04{\sim}1.68cm^2$ and film scanner was $0{\sim}0.78cm$, $0.24{\sim}3.89cm^2$. The more distance and field from axis of central line increase, the more bigger the error value increases. There is a few mm error from standard point at the process which imput various information to apply MLC in clinic. and effective field did not have variation of monitor unit and dose owing to being a few cm2 error against real field. But there are some problem to shield critical organs because some part of target volume induced by the movement of organs can be not included, therefore we have to pay attention on the process to imput MLC layer

최근 방사선치료의 중추적 역할을 담당하고 있는 다엽콜리메이터는 사용의 편리성 때문에 기존의 납합 금 차폐물을 급속도로 대체하고 있다. 그러나 다엽콜리메이터는 엽간 누설선량이나 디지털 신호를 입출력 알고리즘에 따른 오류가 발생할 수 있다. 저자는 미국 Varian사에서 제공하는 다엽콜리메이터의 입력 방법에 따른 차이를 모의치료장치 프로그램(XimaVision)값을 기준으로 전자펜 정보 전송장치(MLC shaper)와 디지털 변환장치(film scanner)에서의 오차와 유효 조사야를 여러 형태의 차폐윤곽으로 비교하였다. 기준 값과 비교하여 기본적인 차폐 윤곽(test1-5)에서 전자펜 전송방식은 $0{\sim}0.29cm$, $0.23{\sim}3.59cm^2$, 디지털 변환장치 방식은 $0{\sim}0.20cm$, $0.72{\sim2.59cm^2$ 차이를 보였다. 임상에서 적용되는 차폐 윤곽에서 전자펜 전송방식은 $0{\sim}0.54cm$, $0.04{\sim}1.68cm^2$, 디지털 변환장치 방식은 $0{\sim}0.78cm$, $0.24{\sim}3.89cm^2$ 차이를 보였고, 대부분 선속 중심선에서 멀어질수록 조사야가 클수록 오차범위가 증가함을 알 수 있었다. 다엽콜리메이터의 임상적용을 위한 다양한 정보 입력과정에서 기준점으로부터 수 mm 이내에서 오차가 잇었다. 그리고 유효 조사야는 실제 면적에 대하여 수 $cm^2$이내에서 오차가 발생하여 기계적 선량(Monitor Unit)이나 조사야 내 선량변화에 관여하지 않았다. 그러나 임계장기를 차폐 할 경우 장기의 움직임으로 표적용적의 일부가 포함되지 않을 수 있기 때문에 문제의 여지가 있다. 그러므로 오류를 최소화하기 위한 다엽콜리메이터의 입력과정에 세심한 주의를 기울여야 한다.

Keywords

References

  1. T. LoSasso, C. S. Chui, G. J. Kutcher, S. A. Leibel, Z. Fuks, and C. C. Ling, "The use of a multi-leaf collimator for conformal radiotherapy of carcinomas of the prostate and nasopharyn," Int. J. Radiation Oncology Biol, Phys. Vol.25, pp.161-170, 1993. https://doi.org/10.1016/0360-3016(93)90337-U
  2. W. D. Powlis, A. R. Smith, E. Cheng, "Innitiation of multileaf collimator conformal radiation theraphy," Int. J. Radiation Oncology Biol. Phys. Vol.25, pp.171-179, 1993. https://doi.org/10.1016/0360-3016(93)90338-V
  3. MLC system and maintenance guide, Varian associates inc. oncology system pub, 1995.
  4. Varian c-series multileaf collimator equipment specification, Varian associates inc. oncology system pub, 1993.
  5. MLC service manual, Varian associates inc. oncology system pub, 1993.
  6. L. Brewster, R. Mohan, G. Mageras, C. Burman, S. Leibel, and Z. Fuks, "Three dimensional conformal treatment planning with multileaf collimators." Int. J. Radiation Oncology Biol. Phys. Vol.33, pp.1081-1089, 1995. https://doi.org/10.1016/0360-3016(95)02061-6
  7. M. Partridge, P. M. Evans, A. Mosleh-Shirazi, and D. Convery, "Independent verification using portal imaging of intensity-modulated beam delivery by the dynamic MLC technique," Med. Phys. Vol.25, pp.1872-1879, 1998. https://doi.org/10.1118/1.598376
  8. Four years experience with the philips multileaf collimator, Medicamundi, Vol.39, No.3, p.117, 1994.
  9. Intensity modulated radiation therapy, "Movement of the multileaf collimator," Japanese journal of radiological technology, Vol.56, No.1, pp.17-21, 2000. https://doi.org/10.6009/jjrt.KJ00001356747
  10. S. C. Prasad, " Effects of collimator jaw setting on dose output for treatments with multileaf collimator," Medical dosimetry, Vol.23, No.4, pp.296-301, 1998. https://doi.org/10.1016/S0958-3947(98)00031-4
  11. G. Loi, E. Pignoli, M. Scorsetti, V. Cerreta, A. Somigliana, R. Marchesini, " Design and characterization of a dynamic multileaf collimator," Physics in Medicine & Biology, Vol.43, No.10, pp.3149-3156, 1998. https://doi.org/10.1088/0031-9155/43/10/033
  12. C. X. Yu, D. Yan, M. N. Du, and S. Zhou, "Optimization of leaf positions when shaping a radiation field with a multileaf collimator," Physics in Medicine & Biology, Vol.40, No.2, pp.305, 1995. https://doi.org/10.1088/0031-9155/40/2/008
  13. E. Grusell and A. Montelius, "The leakage of protons between plane collimator leaves in a multileaf collimator," National institute for radiological sciences japan-report-nirs M, Vol.103, pp.116-124, 1994.
  14. A. Mahadevan, A. Lee, E. Holupka, C. Sampson, and X. Liu, "An optimized multileaf feathering technique for matching single isocenter half beam 3 field head and neck radiation," International Journal of Radiation Oncology, Biology, Physics, Vol.63, p.499, 2005.
  15. S. C. Prasad, "Effects of collimator jaw setting on dose output for treatments with multileaf collimator," Medical dosimetry, Vol.23, No.4, pp.296-30, 1998. https://doi.org/10.1016/S0958-3947(98)00031-4
  16. S. V. Dalen, M. Keijzer, H. Huizenga, and P. R. M. Storchi, "Optimization of multileaf collimator settings for radiotherapy treatment planning," Physics in Medicine & Biology, Vol.45, No.12, pp.3615-3626, 2000. https://doi.org/10.1088/0031-9155/45/12/308
  17. P. C. Williams and P. Cooper, "High-resolution field shaping utilizing a masked multileaf collimator," Physics in Medicine & Biology, Vol.45, No.8, pp.2313-2330, 2000. https://doi.org/10.1088/0031-9155/45/8/317
  18. 이병용, 조병철, 장혜숙, "Multileaf collimator의 적정폭에 관한 연구," 의학물리, 제5권, pp.49-55, 1994.
  19. 양광모, 서현숙, "다엽 콜리메이터와 제작차폐물의 동시 사용시 표면선량 변화", 의학물리, 제13권, 제2호, pp.62-68, 2002.
  20. 이병용, 장혜숙, 조병철, "불균일 폭 분포를 갖는 Multileaf collimator 설계에 관한 연구", 의학물리, 제5권, 제2호, pp.45-48, 1994.
  21. 정갑수, "방사선치료에 있어서 Multileaf collimator의 원리와 그 응용", 대한방사선기술학회지, 제21권, 제1호, pp.24-28, 1998.
  22. 정갑수. "치료용 가속기에 대한 Multileaf collimator의 소프트웨어 개발 및 임상적용", 대한방사선기술학회지, 제22권, 제1호, pp.61-66, 1999.

Cited by

  1. Development of Manual Multi-Leaf Collimator for Proton Therapy in National Cancer Center vol.26, pp.4, 2015, https://doi.org/10.14316/pmp.2015.26.4.250